Zenith Renu G55409Recalls & reported problems
Cook Incorporated · Zenith Renu, AAA Ancillary Graft Main Body Extension · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10827002554099), received between Oct 13, 2017 and Oct 13, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Markings/Labelling Problem1
- Material Separation1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 13, 2017Material Separation; Device Markings/Labelling Problem; Material Separation; Device Markings/Labelling Problem
THE AREA REPRESENTATIVE REPORTED DURING A TYPE 1A ENDOLEAK REPAIR, WHILE UNSHEATHING THE ZENITH RENU AAA ANCILLARY GRAFT MAIN BODY EXTENSION, ONE OF THE GOLD MARKERS ON THE DEVICE CAME OFF AND EMBOLIZED IN THE LUMBAR ARTERY. THE GRAFT WAS IMPLANTED. AS REPORTED, THE DOCTOR IS NOT SURE WHETHER INTERVENTION WILL BE PERFORMED. CURRENTLY THE DOCTOR IS DEALING WITH THE PRIMARY PROBLEM OF THE ENDOVASCULAR ANEURYSM REPAIR. NO ADVERSE EFFECTS HAVE BEEN REPORTED TO THE PATIENT DUE TO THIS OCCURRENCE
Patient problems: Foreign Body In Patient
MAUDE report 6949126 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page