Skinvive 98307Recalls & reported problems
Allergan, Inc. · Skinvive 2 x 1.0 mL · Class 3 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Allergan Skinvive dermal filler is an injectable product used for skin enhancement. All 183 reports are injuries and there are no deaths. Many describe lumps, swelling and redness after injection, and many note use outside the labeled procedure. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 180
- Deaths reported
- 0
- Injuries reported
- 180
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 180 adverse event reports that list this exact device (DI 10840228300066), received between Oct 12, 2023 and Sep 15, 2026 — classified as 180 injuries.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (70 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem80
- Improper or Incorrect Procedure or Method70
- Off-Label Use34
- Device Appears to Trigger Rejection1
- Use of Incorrect Control/Treatment Settings1
Patient problems
- Obstruction/Occlusion65
- Nodule49
- Subcutaneous Nodule24
- Abscess7
- Inflammation6
- Unspecified Infection6
- Hypersensitivity/Allergic reaction5
- Ischemia5
Recent adverse event reports
The 25 most recent of 180 reports, in the reporter's own words as published by the FDA.
InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem
Healthcare professional reported via a company representative that a patient was injected 2 ml in the cheeks with SKINVIVE¿ by JUVÉDERM®, \"Patient developed 30 red, swollen, and firm nodules. Every injection site was affected, although not every injection site exhibited all three symptoms\". The patient was treated with antihistamines and a methylprednisolone dose pack. The patient has begun to improve. The event is ongoing.
Patient problems: Nodule
MAUDE report 50766054 ↗InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem
Healthcare professional reported a patient was injected with SKINVIVE¿ by JUVÉDERM® into the right and left malar/zygomatic areas. Patient also concomitantly received nuera and back and cheek laser hair removal. The next day, patient reported presenting with left cheek erythema and good capillary refill. Swelling and mild tenderness also reported. Patient showed slight improvement of erythema each day. Patient was treating area with 2.5% hydrocortisone ointment and biopelle 40 serum multiple times a day. It was noted as of three days after onset the swelling remained and capillary refill was 1-2 seconds. The lower edge of the cheek area was a little darker in color and the redness of the upper edge had resolved by 75%. Patient was prescribed a medrol dose pack, starting with 4 a day then tapering. Patient also given Revivisomes, biopelle 40's 2.5% hydrocortisone, doxycycline 100mg. Patient also reportedly rook bacitracin ointment. Patient travelled overseas but returned two weeks later. It was noted the skin had cleared up, good capillary refill and very little erythema. There was a small amount of hyperpigmentation in the lower perinasal area remaining. Patient reportedly had pus
Patient problems: Abscess
MAUDE report 50762785 ↗InjuryAug 31, 2026Adverse Event Without Identified Device or Use Problem
Healthcare professional reported a patient was injected with 2mls of SKINVIVE¿ by JUVÉDERM® into the cheeks and about three months later patient experienced redness and little bumps on both sides, but more prevalent on the left. Patient was treated with hylenex that day and was prescribed antihistamines and antibiotics the following day. Symptoms initially dissipated but \"came back\".
Patient problems: Nodule
MAUDE report 50531971 ↗InjuryJul 9, 2026Improper or Incorrect Procedure or Method
HEALTHCARE PROFESSIONAL REPORTED VIA A COMPANY REPRESENTATIVE THAT A PATIENT WAS INJECTED 1 ML IN THE LIPS, UPPER LIP, SIDES OF CHIN WITH SKINVIVE¿ BY JUVÉDERM®, PATIENT PRESENTED "SENSITIVITY". LATER PHYSICIAN REPORTED "PATIENT EXPERIENCED REACTION 8 DAYS AFTER INJECTION, NODULE SWELLING TO RIGHT SIDE OF CHIN". MED DOSE PACK WAS GIVEN AS TREATMENT. THE EVENT IS ONGOING.
Patient problems: Inflammation; Nodule
MAUDE report 25774226 ↗InjuryJun 24, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED INJECTING A PATIENT WITH SKINVIVE¿ BY JUVÉDERM®. THE PATIENT EXPERIENCED ¿DIFFUSED SWELLING REDNESS.¿ THE PATIENT WAS PRESCRIBED ANTIBIOTICS AND STEROIDS. EVENT STATUS WAS UNKNOWN.
Patient problems: Inflammation
MAUDE report 25613218 ↗InjuryJun 24, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED INJECTING A PATIENT WITH SKINVIVE¿ BY JUVÉDERM®. THE PATIENT EXPERIENCED ¿DIFFUSED SWELLING REDNESS.¿ THE PATIENT WAS PRESCRIBED ANTIBIOTICS AND STEROIDS. EVENT STATUS WAS UNKNOWN.
Patient problems: Inflammation
MAUDE report 25613206 ↗InjuryJun 1, 2026Improper or Incorrect Procedure or Method
HEALTHCARE PROFESSIONAL REPORTED INJECTING A PATIENT WITH SKINVIVE¿ BY JUVÉDERM® IN THE LEFT MIDFACE WITH 0.7ML. IMMEDIATELY POST INJECTION, PATIENT EXPERIENCED "IRREGULAR, PATCHY BLANCHING WAS OBSERVED OVER THE LEFT MIDFACE. CAPILLARY REFILL TIME (CRT) WAS APPROXIMATELY 1 SECOND. THE PATIENT REPORTED NO PAIN, VISUAL DISTURBANCE, OR ABNORMAL SENSATION". BLANCHING RESOLVED SPONTANEOUSLY A FEW MINUTES POST-INJECTION AND EVOLVED INTO PURPLISH DISCOLORATION. INITIAL SYMPTOM MANAGEMENT INCLUDED 750 IU DOUBLE-DILUTED HYALURONIDASE. ONE HOUR POST-INJECTION, PURPLE COLOR REMAINED, AN ADDITIONAL 1200 IU DOUBLE-DILUTED HYALURONIDASE WAS ADMINISTERED. CRT <2 SECONDS. FIVE HOURS POST-INJECTION PATIENT NOTED MILD RED-PURPLE DISCOLORATION PERSISTED. SIXTEEN TO TWENTY-FOUR HOURS POST-INJECTION HCP NOTED THE DISCOLORATION PERSISTED BUT "REMAINED STABLE AND NON-PROGRESSIVE. THIRTY-SIX HOURS POST-INJECTION PATIENT WAS ASYMPTOMATIC AND FELT NORMAL. CRT <1 SECOND. PERSISTENT "ECCHYMOTIC DISCOLORATION CONSISTENT WITH BRUISING/VENOUS CONGESTION. NO BLISTERING, NECROSIS, OR NEUROLOGICAL SIGNS". PATIENT IS RECEIVING DAILY LOW-LEVEL RED LED LIGHT THERAPY. SYMPTOMS ARE ONGOING.
Patient problems: Obstruction/Occlusion
MAUDE report 25351629 ↗InjuryMay 26, 2026Off-Label Use
HEALTHCARE PROFESSIONAL REPORTED INJECTING A PATIENT WITH ONE SYRINGE OF SKINVIVE¿ BY JUVÉDERM® TO THE PERIORAL AREA. PATIENT EXPERIENCED "RAISED BUMPS". PATIENT WAS TREATED WITH HYLENEX. ONE WEEK LATER, PATIENT WAS TREATED WITH A SECOND ROUND OF HYLENEX AND DEXAMETHASONE 10 MG. SYMPTOMS ARE ONGOING.
Patient problems: Subcutaneous Nodule
MAUDE report 25279444 ↗InjuryMay 19, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED INJECTING A PATIENT IN THE CHEEKS AND PREAURICULAR WITH 2.7 ML OF JUVÉDERM® VOLUMA¿ XC AND IN THE JAWLINE WITH 1 ML OF JUVÉDERM® VOLUX¿ XC. TWO AND HALF MONTHS LATER, THE PATIENT WAS INJECTED IN THE JAWLINE WITH 1 ML OF JUVÉDERM® VOLUX¿ XC AND IN THE CHEEKS WITH 2 ML OF SKINVIVE¿ BY JUVÉDERM®. FOUR MONTHS LATER, THE PATIENT EXPERIENCED ¿DELAYED ONSET NODULES¿ AT THE INJECTION SITES. FOR THE NEXT 11 MONTHS, THE PATIENT HAD MULTIPLE TREATMENTS WITH HYLENEX, PREDNISONE, AND ANTIHISTAMINES. TWO MONTHS POST-ONSET, AN AEROBIC + ANAEROBIC WOUND CULTURE WAS PERFORMED, RESULTING AS ¿NEGATIVE.¿ THE EVENT PARTIALLY RESOLVED AT THE END OF THE TREATMENT ROUND BUT RESURFACED 15 MONTHS LATER. THAT MONTH, THE PATIENT WAS TREATED WITH 4 COURSES OF ANTIBIOTICS (DOXYCYCLINE, CLARITHROMYCIN, AND AUGMENTIN), PREDNISONE TAPER, ANTIHISTAMINES, AND 5FU. EVENT WAS ONGOING. THIS FILE CAPTURED THE SKINVIVE¿ BY JUVÉDERM®.
Patient problems: Subcutaneous Nodule
MAUDE report 25217494 ↗InjuryApr 14, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® AND EXPERIENCED BUMPS IN THE CHEEKS THAT HAVE REMAINED. ONE BUMP IS LARGER AND PALPABLE. ONE NODULE HAS SOME PAIN.
Patient problems: Nodule
MAUDE report 24875214 ↗InjuryApr 6, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL (HCP) REPORTED INJECTING A PATIENT WITH SKINVIVE¿ BY JUVÉDERM®. PATIENT LATER "STARTED GETTING GROWTH ON THE ZYGOMA" AND "HAD MULTIPLE GRANULOMAS REMOVED". HCP NOTES THE PATIENT IS STILL "ACTIVELY FORMING THEM" [GRANULOMAS]. BIOPSY WAS NOT PROVIDED. SYMPTOMS ARE ONGOING.
Patient problems: Granuloma
MAUDE report 24794500 ↗InjuryApr 3, 2026Off-Label Use
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH 2MLS OF SKINVIVE¿ BY JUVÉDERM® INTO THE LOWER FACE AND MARIONETTE AREAS AND OTHER PRODUCTS, INCLUDING SYLFIRM, AND SEVEN MONTHS LATER EXPERIENCED A BUMP UNDER THE SKIN IN THE RIGHT LOWER MANDIBLE. HEALTHCARE PROFESSIONAL WONDERS IF IT COULD BE A POSSIBLE DELAYED ONSET NODULE. A CT SCAN WAS DONE ABOUT A MONTH AND A HALF AFTER ONSET NOTING "DERMAL THICKENING". A KENALOG INJECTION WAS PERFORMED AND EVENT HAS NOT RESOLVED.
Patient problems: Subcutaneous Nodule
MAUDE report 24783225 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPARATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SKINVIVE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Bacterial Infection
MAUDE report 24747290 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPARATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SKINVIVE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Bacterial Infection
MAUDE report 24747281 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPERATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SKINVIVE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Bacterial Infection
MAUDE report 24747275 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPERATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SKINVIVE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Bacterial Infection
MAUDE report 24747269 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPERATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SKINVIVE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Subcutaneous Nodule
MAUDE report 24747004 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPARATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SKINVIVE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Subcutaneous Nodule
MAUDE report 24746999 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPERATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SEPARATE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Subcutaneous Nodule
MAUDE report 24746972 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® ON TWO SEPARATE OCCASIONS WITH NO INCIDENT. PATIENT LATER HAD A THIRD SKINVIVE¿ BY JUVÉDERM® TREATMENT AND EXPERIENCED SMALL BUMPS ONE MONTH POST INJECTION. HYALURONIDASE WAS INJECTED AND THE BUMPS RESOLVED. PATIENT LATER UNDERWENT A FOURTH TREATMENT WITH SKINVIVE¿ BY JUVÉDERM® AND THEN FIVE MONTHS LATER EXPERIENCED THE SIMILAR BUMPS AGAIN WITH INTERMITTENT RESOLUTION. ABOUT TWO MONTHS AFTER THE RECURRENCE OF THE BUMPS, PATIENT DEVELOPED MIGRAINES AND A SINUS INFECTION THAT REQUIRED A VISIT TO THE ER. THE BUMPS HAVE CONTINUED TO PERSIST. PATIENT WAS PRESCRIBED PREDNISONE, DOXYCYCLINE, AND IV ANTIBIOTICS AT THE ER. EVENT ONGOING.
Patient problems: Subcutaneous Nodule
MAUDE report 24746936 ↗InjuryMar 30, 2026Off-Label Use; Improper or Incorrect Procedure or Method
A HEALTHCARE PROFESSIONAL REPORTED THAT A PATIENT WAS INJECTED WITH 0.3 ML OF SKINVIVE¿ BY JUVÉDERM® IN THE FOREHEAD AND BETWEEN THE BROWS. POST INJECTION THE HCP NOTICED THAT THE INJECTED AREA LOOKED RED, AND THEY SAW A LITTLE WHITE SPOT IN THE GLABELLA. THE HCP STATED THAT THE SKIN WAS BLANCHED, AND THEY THOUGHT IT WAS AN OCCLUSION. THE HCP TREATED THE PATIENT WITH A BOX OF HYLENEX AND THE PATIENT WAS INJECTED 1 MONTH LATER WITH 0.4 ML OF HYLENEX AGAIN. ABOUT 5 MONTHS AFTER ONSET PATIENT ALSO UNDWERWENT AN IPL(INTENSE PULSE LIGHT) THERAPY FOR THE REDNESS AND A VI PEEL ABOUT TWO WEEKS LATER. EVENT ONGOING.
Patient problems: Obstruction/Occlusion
MAUDE report 24736652 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED INJECTING A PATIENT WITH 1CC OF JUVÉDERM® VOLUMA® XC IN THE CHEEKS, AND 2CC OF SKINVIVE¿ BY JUVÉDERM® IN THE CHEEKS. PATIENT CONCOMITANTLY RECEIVED 3 SYRINGES OF NON-ABBVIE VERSA IN THE CHIN AND CHEEKS, AS WELL AS 6U OF BOTOX IN THE CHIN. TWO YEARS LATER, PATIENT RECEIVED A SERIES OF NOUVADERM LASER TREATMENTS WITH EXOIE PLANT EXOSOMES. THE SECOND OCCURRED TWO MONTHS LATER USING THE NOUVADERM 1927 FRACTIONATED LASER TREATMENT. THE THIRD AND FINAL TREATMENT OCCURRED ABOUT TWO MONTHS LATER WHERE THE PATIENT RECEIVED A VI PURIFY WITH PRECISION PLUS AFTER THE NOUVADERM LASER. APPROXIMATELY TWO WEEKS LATER, PATIENT CONTRACTED COVID WHICH TRIGGERED SWELLING IN THE GLABELLA, UNDER EYES, AND LEFT CHIN/JAWBONE. PATIENT DID NOT ORIGINALLY BELIEVE THE SYMPTOMS WERE RELATED TO THE FILLER INJECTIONS 2 YEARS PRIOR. PATIENT HAD AN MRI PERFORMED. PATIENT NOTED A NODULE ON THE CHIN, LATER DESCRIBED BY THE HCP AS A GRANULOMA. HCP NOTED GRANULOMAS SEEM TO BE "IN [THE] GLABELLA ( [HCP] DIDN¿T PUT ANY FILLER THERE BUT DID TREAT [THE PATIENT] WITH BOTOX THERE" AS WELL AS IN THE UNDER EYE AREA, NEAR THE "MEDIAL CHEEK ALMOST IN THE TEAR TROUGH, AND ON [THE] RIGHT JAW". SYM
Patient problems: Granuloma
MAUDE report 24736200 ↗InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem
VIA SOCIAL MEDIA, A PATIENT REPORTED BEING INJECTED WITH SKINVIVE¿ BY JUVÉDERM®. THE PATIENT ADDITIONALLY REPORTED ¿JUST A LITTLE RED AND BUMPY FOR A FEW HOURS.¿ NO TREATMENT INFORMATION WAS PROVIDED. SYMPTOMS RESOLVED AFTER A FEW HOURS.
Patient problems: Nodule
MAUDE report 24722972 ↗InjuryMar 25, 2026Off-Label Use
A PATIENT REPORTED BEING INJECTED OFF LABEL IN THE LIPS WITH SKINVIVE¿ BY JUVÉDERM®. APPROXIMATELY TWO YEARS POST INJECTIONS, THE PATIENT ¿WAS PREGNANT AND LIPS SWELLED, NODULES FORMED AND PRODUCT MIGRATED.¿ NO TREATMENT INFORMATION OR SYMPTOMS RESOLUTION WAS PROVIDED.
Patient problems: Nodule
MAUDE report 24689968 ↗InjuryMar 19, 2026Off-Label Use
A HEALTHCARE PROFESSIONAL REPORTED A PATIENT WAS INJECTED WITH SKINVIVE¿ BY JUVÉDERM® BY ANOTHER HEALTHCARE PROFESSIONAL INTO THE CHEST REGION AND LATER EXPERIENCED DELAYED ONSET NODULES. THE INJECTING HEALTHCARE PROFESSIONAL REPORTEDLY ATTEMPTED TO DISSOLVE NODULES, BUT THEY HAVE REMAINED.
Patient problems: Subcutaneous Nodule
MAUDE report 24643942 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P110033 on Oct 22, 2013. Submitted by Allergan as “Juvederm Voluma Xc”. The FDA review took 785 days. The FDA has since approved 97 supplements (changes to the device or its labeling), most recently on Aug 28, 2026. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Common questions about the Skinvive 98307
Has the Skinvive 98307 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10840228300066), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Skinvive 98307?
The FDA's MAUDE database lists 180 adverse event reports for this device, received between Oct 12, 2023 and Sep 15, 2026. In those reports, the most common specific device problems are “Improper or Incorrect Procedure or Method” (70), “Off-Label Use” (34) and “Device Appears to Trigger Rejection” (1); the most common patient problems are “Obstruction/Occlusion” (65), “Nodule” (49) and “Subcutaneous Nodule” (24). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Skinvive 98307?
None of the reports are classified as deaths, 180 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Skinvive 98307 FDA approved?
It was approved through premarket approval (PMA) P110033 on Oct 22, 2013.
Is the Skinvive 98307 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Skinvive 98307, and is it still sold?
The FDA's device database lists Allergan, Inc. as the company responsible for it (brand Skinvive, model 98307). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Implant, Dermal, For Aesthetic Use devices
- JUVEDERM VOLUMA XC 94640Allergan, Inc. · Implant, Dermal, For Aesthetic Use673 reports
- JUVEDERM ULTRA XC 94154Allergan, Inc. · Implant, Dermal, For Aesthetic Use542 reports
- JUVEDERM VOLLURE XC 95661Allergan, Inc. · Implant, Dermal, For Aesthetic Use377 reports
- JUVEDERM ULTRA PLUS XC 94155Allergan, Inc. · Implant, Dermal, For Aesthetic Use319 reports
- JUVÉDERM VOLBELLA XC 95834Allergan, Inc. · Implant, Dermal, For Aesthetic Use187 reports
- Revanesse Lips+ with Lidocaine 1.2 mL 40149Prollenium Medical · Implant, Dermal, For Aesthetic Use71 reports
- JUVÉDERM VOLBELLA XC 96181Allergan, Inc. · Implant, Dermal, For Aesthetic Use62 reports
- Radiesse (+) Lidocaine 8063M0K2Merz North America, Inc. · Implant, Dermal, For Aesthetic Use61 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page