Skip to content

Seal® SBC-580Recalls & reported problems

Endochoice, Inc. · Biopsy Valve · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10840253110067), received between Sep 17, 2018 and Dec 11, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash3
  • Detachment of Device or Device Component1
  • Device-Device Incompatibility1
  • Material Fragmentation1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 11, 2024Material Fragmentation; Device-Device Incompatibility; Material Fragmentation; Device-Device Incompatibility

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED WITH A SPEEDBAND SUPERVIEW SUPER 7 DURING AN ESOPHAGEAL VARICES - LIGATION BANDING PROCEDURE IN THE ESOPHAGUS PERFORMED FOR THE TREATMENT OF ESOPHAGEAL VARICES ON (B)(6) 2024. DURING THE PROCEDURE, THE CAP OF SEAL BIOPSY VALVE WAS DETACHED WHEN THE PHYSICIAN PLACED THE SPEEDBAND LIGATION DEVICE ONTO THE BIOPSY CAP. THE BIOPSY CAP FELL INTO THE WORKING CHANNEL OF THE SCOPE AND THEN INSIDE THE PATIENT'S BODY. THE FRAGMENT WAS RETRIEVED BY THE PHYSICIAN USING FORCEPS. IT WAS REPORTED THAT THE METAL ROD OF THE SPEEDBAND CAUSED THE DETACHMENT WHEN IT WAS SEATED ONTO THE BIOPSY CAP. THE PROCEDURE WAS COMPLETED WITH THE SAME ORIGINAL SEAL BIOPSY VALVE. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20907274 ↗
  • MalfunctionMay 3, 2019Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BIOPSY CAP POPPED OFF FROM THE BIOPSY PORT OF THE SCOPE CAUSING FLUIDS TO SPRAY EVERYWHERE. THEY WERE ABLE TO COMPLETE THE CASE WITH THIS DEVICE. THERE WAS NO SERIOUS INJURY NOR ANY ADVERSE PATIENT EFFECTS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8577746 ↗
  • MalfunctionJan 15, 2019Leak/Splash; Leak/Splash

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY/ESOPHAGOGASTRODUODENOSCOPY(EGD) PROCEDURE ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, THE BIOPSY CAP WAS LEAKING AND SPRAYING BODILY FLUID. IT WAS NOT REPORTED HOW THE PROCEDURE WAS COMPLETED. THERE WAS NO SERIOUS INJURY NOR ADVERSE PATIENT EFFECTS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8248943 ↗
  • MalfunctionSep 17, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SEAL BIOPSY VALVE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE LID OF THE SEAL BIOPSY VALVE CAME OPEN WHEN REMOVING A DEVICE CAUSING FLUID TO LEAK AND GOT ON TO A CAREGIVER'S FACE. AN IRRIGATION TO THE CAREGIVER'S EYES WAS DONE FOR 15 MINUTES. THERE WAS NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT AS A RESULT OF THIS EVENT. THE CAREGIVER WAS REPORTED TO BE FINE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7881413 ↗

Other Endoscopic Irrigation/Suction System devices

All Endoscopic Irrigation/Suction System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page