TRAPLINER IPN913564Recalls & reported problems

Teleflex Incorporated · TrapLiner 8F · Class 2

All models of TRAPLINER (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10841156100568), received between Aug 23, 2023 and Nov 7, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Separation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT: "THE TRAPLINER WAS ADVANCED TOO FAR EXPOSING THE COLLAR OUTSIDE OF THE GUIDE CATHETER. WHEN REALIZED PHYSICIAN ATTEMPTED TO CORRECT AND PULL BACK INTO THE GUIDE CATHETER CAUSING THE TRAPLINER TO BREAK INTO PIECES AT THE FLEX POINT WHILE CROSSING THE VD-OM. CATHETER WAS RETRIEVED SUCCESSFULLY WITH A GUIDELINER AND BALLOON AND PROCEDURE WAS COMPLETED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23494576 ↗
  • MalfunctionMay 9, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT: "UPON REMOVAL FROM THE HOOP, THE TRAPLINER WAS FOUND TO BE IN TWO SEPARATE PIECES. ANOTHER TRAPLINER WAS USED SUCCESSFULLY." THE ISSUE WAS IDENTIFIED PRIOR TO USE; THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21992540 ↗
  • MalfunctionApr 30, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: DURING A PROCEDURE, THE RT PREPPING THE TRAPLINER FLUSHED THE DEVICE. THE RT NOTED THAT THE TIP WAS SEPARATED. THE DEVICE WAS NEVER INSERTED INTO THE PATIENT. IT WAS SET ASIDE AND ANOTHER 8F TRAPLINER WAS PREPPED AND USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19211327 ↗
  • MalfunctionAug 23, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT: A PIECE OF TRAPLINER BROKE OFF WHEN TECH WAS PREPPING IT. NOT USED ON A PATIENT. NO ADDITIONAL INFORMATION RECIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17601453 ↗

Other Catheter, percutaneous devices