GUIDELINER IPN913565Recalls & reported problems
Teleflex Incorporated · GuideLiner V3 5F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10841156100575), received between Oct 15, 2020 and Oct 15, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2020Material Fragmentation; Material Fragmentation
AS REPORTED: DURING A PCI PLUS STENTING, A PIECE OF THE GUIDELINER V3 BROKE OFF, HOWEVER, FOUND AND SAVED. BROKEN PART IS IN A LITTLE CONTAINER. DEVICE PROFESSIONAL ALLEGES THE DEVICE DID NOT CONTRIBUTE TO THE EVENT. NO MEDICAL INTERVENTION REPORTED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO RECEIVE ADDITIONAL INFORMATION, HOWEVER, NO RESPONSE HAS BEEN RECEIVED AT THIS TIME. PORTED: DURING A PCI PLUS STENTING, A PIECE OF THE GUIDELINER V3 BROKE OFF, HOWEVER, FOUND AND SAVED. BROKEN PART IS IN A LITTLE CONTAINER. DEVICE PROFESSIONAL ALLEGES THE DEVICE DID NOT CONTRIBUTE TO THE EVENT. NO MEDICAL INTERVENTION REPORTED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO RECEIVE ADDITIONAL INFORMATION, HOWEVER, NO RESPONSE HAS BEEN RECEIVED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10684817 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page