TURNPIKE IPN913576Recalls & reported problems
Teleflex Incorporated · Turnpike LP 135cm · Class 2
All models of TURNPIKE (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10841156100681), received between Mar 25, 2020 and May 9, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
- Separation Problem1
Patient problems
- Foreign Body In Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 9, 2024Material Separation; Material Separation
IT WAS REPORTED BY CLINICAL STAFF THAT: "WIGGLE WIRE STUCK BEHIND NEWLY PLACED STENT. INSERTED TURNPIKE LP TO FACILITATE RELEASE OF THE WIRE. TIP OF THE TURNPIKE APPEARS TO HAVE CLAMPED DOWN ON OR FUSED TO THE WIRE. UPON REMOVAL OF THE TURNPIKE LP FROM THE BODY IT WAS NOTED THAT PART OF THE TIP DETACHED FROM THE TURNPIKE LP AND REMAINED ON THE WIRE INSIDE THE BODY. THEN LASER WAS USED. VESSEL DISSECTED. PT CODED. PT PLACED ON ECMO. AFTER THE ECMO WAS IN PLACE THE WIRE AND THE TIP OF THE TURNPIKE THAT WAS STUCK, WAS RELEASED USING A GUIDE EXTENSION. THE PART OF THE TIP THAT HAD BROKEN OFF WAS STILL ON THE WIRE AND COMPLETELY REMOVED FROM THE PATIENT ALONG WITH THE WIRE. THE FRESH STENT THAT HAD BEEN PLACED WAS ALSO REMOVED ALONG WITH THE WIRE. NEW STENT PLACED. CASE COMPLETED. PT TRANSFERRED." ADDITIONAL INFORMATION: "THE USE OF THE TURNPIKE LP AND THE LASER WERE BOTH USED AS AN INTERVENTION DUE TO THE WIGGLE WIRE BEINGS PINNED AND UNABLE TO BE REMOVED FROM BEHIND A FRESHLY PLACED STENT. MULTIPLE DEVICES WERE USED TO REMOVE THE WIRE. DURING THIS PROCESS THE VESSEL DISSECTED, AND PATIENT CODED. THIS LED TO CPR AND EVENTUALLY ECMO. THE PATIENT WAS REPORTED AS DECEASED APPROXIMATELY 8-
Patient problems: Foreign Body In Patient
MAUDE report 19276354 ↗InjuryJan 17, 2024Material Separation; Material Separation
DR. WAS ATTEMPTING TO CROSS A TIGHT LAD LESION DURING ATHERECTOMY. HE ATTEMPTED CROSSING WITH A BALLOON AND TPK LP 135. WHEN TRYING TO REMOVE THE TURNPIKE, THE TIP BROKE OFF IN THE PATIENT. THEY TRIED USING PRONTO TO SUCK IT OUT- UNSUCCESSFULLY. AS OF NOW, THEY HAVE CONSULTED WITH THE SURGEONS AND HAVE DECIDED TO LEAVE IT IN THE PATIENT FOR NOW. PATIENT IS STILL IN ICU. THEY ARE OK AS PER CATH LAB MANAGER.
Patient problems: Foreign Body In Patient
MAUDE report 18524282 ↗MalfunctionMar 25, 2020Separation Problem; Separation Problem
IT WAS REPORTED THAT THE DEVICE FAILED TO PERFORM ITS FUNCTION. NO FURTHER COMPLICATIONS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED 26FEB2020: THE EVENT OCCURRED DURING CX CTO RE-CANALIZATION (IN-STENT RESTENOSIS) WITHOUT ANY ACTUAL FEEDBACK (NO RESISTANCE) DURING TORQUEING AND THE MICRO-CATHETER (MC) GOT STUCK INSIDE TURNPIKE LP. THERE WAS NO POSSIBILITY TO MOVE WIRE INSIDE MC AND FINALLY BOTH DEVICES (WIRE WITH MC) HAD TO BE REMOVED FROM THE VESSEL. AFTER VISUAL INSPECTION CATHETER WAS TWISTED AROUND WIRE. THE MAIN POINT WAS NO PERCEPTIBLE RESISTANCE WHEN MC GOT STUCK AND STARTED TO TWIST. NO FURTHER COMPLICATIONS WERE REPORTED. MANUFACTURER INVESTIGATION WAS PERFORMED ON 16MAR2020 AND IT WAS NOTED THAT THERE WAS TIP DAMAGE/SEPARATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9878533 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026