Skip to content

TURNPIKE IPN917513Recalls & reported problems

Teleflex Incorporated · Turnpike LP, 135cm, TCT · Class 2

All models of TURNPIKE (18)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10841156108731), received between Dec 6, 2022 and Aug 30, 2024 — classified as 1 death, 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1
  • Separation Problem1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT: "I AM REACHING OUT REGARDING A DAMAGED TELEFLEX PRODUCT (REF 5638). THIS PRODUCT WAS USED DURING A PATIENT PROCEDURE AND IT MALFUNCTIONED. " ADDITIONAL INFORMATION RECEIVED ON 19AUG2024: "I SPOKE WITH THE MANAGER SHE WAS IN THE CASE. SHE TOLD ME SHE DOES NOT HAVE THE "RESOURCES" TO RETRIEVE ALL THE DETAILS, BUT TOLD ME THAT IT WAS A VERY DIFFICULT CASE WITH A TIGHT LESION AND THE TURNPIKE SEEMED TO BE STUCK AND WHEN THEY TRIED TO MANEUVER THE CATHETER A PIECE OF THE TIP SEEMED TO DISLOCATE. THEY PLACED A WIRE NEXT TO THE MICROCATHETER AND WERE ABLE TO REMOVE EVERYTHING TOGETHER. NOTHING WAS LEFT BEHIND AND THEY WERE ABLE TO COMPLETE THE CASE USING ANOTHER TURNPIKE LP. THE PATIENT WAS REPORTED AS FINE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20105582 ↗
  • InjuryMar 25, 2024

    PATIENT PLACED ON TABLE AND PROCDURE STARTED SOON AFTER. BMW WIRE INSERTED THEN TURNPIKE WAS INSERTED. TURNPIKE WAS REMOVED SOON AFTER. TIP ON TURNPIKE WAS RETAINED IN VESSEL; NOTED ON FLUOROSCOPY.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 18967783 ↗
  • DeathDec 6, 2022Separation Problem; Separation Problem

    IT WAS REPORTED THAT: THE TIP OF TURNPIKE SEPARATED ONTO WIRE. WIRE WOULD NOT COME OUT OF GUIDE. GUIDE AND WIRE REMOVED TOGETHER. TIP OF TURNPIKE LP REMAINS ON WIRE. ADDITIONAL INFORMATION REPORTED ON 11NOV2022: THE ENTIRE TURNPIKE AND TIP WERE REMOVED FROM THE VESSEL AFTER TIP SEPARATION. THE PHYSICIAN THEN COMPLETED THE PROCEDURE. THE PATIENT CODED IN RECOVERY. DATE OF DEATH (B)(6) 2022. THERE WERE NO ISSUES IN THE LAB WITH THE PATIENT AS FAR AS HEMODYNAMICS, PAIN, ETC. THE PHYSICIAN REPORTED THAT HE DOES NOT BELIEVE THE TIP SEPARATION HAD ANYTHING TO DO WITH THE EXPIRATION OF THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 15924584 ↗

Other Catheter, percutaneous devices

All Catheter, percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page