TURNPIKE IPN917517Recalls & reported problems
Teleflex Incorporated · Turnpike Smooth, 135cm, TCT · Class 2
All models of TURNPIKE (18)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10841156108779), received between Aug 18, 2022 and May 8, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Fracture1
- Insufficient Device Problem Information1
Patient problems
- Foreign Body In Patient1
- Obstruction/Occlusion1
- Reduced Blood Flow1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 8, 2023Fracture; Fracture
IT WAS REPORTED THAT: WE HAD AN ISSUE YESTERDAY WITH A TURNPIKE TIP FRACTURING IN THE PATIENTS DIAG ARTERY. WHEN IT WAS FINALLY RETRIEVED , IT WAS BOUND TO THE WIRE. ADDITIONAL INFORMATION: THE ABOVE ISSUE OCCURRED DURING A PCI PROCEDURE ON THE (B)(6) 2023, THE PATIENT SUFFERED A DISRUPTION OF THE LAD & IMPAIRED FLOW TO THE LAD PER MD'S NOTE. TO FIX THIS, MD NEEDED TO STENT LAD AND DIAGONAL. THE TURNPIKE WAS REMOVED IN ITS ENTIRETY. THE PATIENT'S VESSELS WERE DESCRIBED AS CALCIFIED. POST PROCEDURE THE PATIENT'S CONDITION WAS REPORTED AS STABLE, THEY WERE SENT TO ICU, THEN DISCHARGED HOME FROM THERE.
Patient problems: Obstruction/Occlusion; Reduced Blood Flow
MAUDE report 16884404 ↗MalfunctionAug 18, 2022Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
AS REPORTED: COMPLETE TIP SEPARATION. WHEN THE TURNPIKE SMOOTH WAS ADVANCED, AND THE PHYSICIAN STARTED ADVANCING IN A CLOCKWISE MOTION THE ENTIRE TIP IMMEDIATELY SEPARATED WITH VERY LITTLE RESISTANCE. HE SAID IT DIDN'T FEEL RIGHT, AND UPON REMOVAL FOR INSPECTION THE TIP WAS ALREADY SEPARATED AND DETACHED. THE TIP WAS LEFT IN, AND ANOTHER GUIDEWIRE WAS UNABLE TO CROSS. THE TIP WAS NOT ABLE TO BE RETRIEVED. ADDITIONAL INFORMATION ON 05AUG2022: THE ENTIRE TIP OF THE TURNPIKE SMOOTH CAME OFF AND IT WAS NOT RETRIEVABLE AND REMAINS IN THE VESSEL. IT WAS REPORTED THAT THERE WAS A CLEAR CHANNEL AND THE PHYSICIAN EXPERIENCED LITTLE RESISTANCE ADVANCING THE CATHETER. AN ANGIOGRAM WAS SUBMITTED FOR INVESTIGATION. PATIENT'S CURRENT CONDITION IS REPORTED AS FINE.
Patient problems: Foreign Body In Patient
MAUDE report 15254499 ↗
Other Catheter, percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page