Skip to content

Legacy™3 Implant 864213Recalls & reported problems

Implant Direct Sybron Manufacturing LLC · 4.2mmD x 13mmL, 3.5mmD Platform · Class 2 · Implant

All models of Legacy3 Implant (75)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
17
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 10841307102533), received between Nov 19, 2018 and Sep 1, 2022 — classified as 17 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of Osseointegration” (9 reports).

11 other recalls cover devices cleared under the same FDA submission (K090234); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loss of Osseointegration9
  • Failure to Osseointegrate4
  • Osseointegration Problem2
  • Device Appears to Trigger Rejection1
  • Inaccurate Information1
  • Patient Device Interaction Problem1

Patient problems

  • Failure of Implant17
  • Unspecified Infection8
  • Inflammation6
  • Osteopenia/ Osteoporosis5
  • Pain4
  • No Consequences Or Impact To Patient3
  • Fibrosis1
  • Granuloma1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • InjurySep 1, 2022Loss of Osseointegration; Loss of Osseointegration

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Osteopenia/ Osteoporosis; Unspecified Infection

    MAUDE report 15336539 ↗
  • InjuryJul 14, 2022Loss of Osseointegration; Loss of Osseointegration

    PER COMPLAINT (B)(4), PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 15014220 ↗
  • InjuryJun 13, 2022Loss of Osseointegration; Loss of Osseointegration

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Granuloma; Inflammation; Osteopenia/ Osteoporosis; Pain

    MAUDE report 14675238 ↗
  • InjuryMay 7, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4). AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Pain

    MAUDE report 14317354 ↗
  • InjuryNov 22, 2021Loss of Osseointegration; Loss of Osseointegration

    PER COMPLAINT (B)(4), PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 12852516 ↗
  • InjuryJul 13, 2021Loss of Osseointegration; Loss of Osseointegration

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED LOSS OR FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 12158715 ↗
  • InjuryJan 29, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PER (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 11250619 ↗
  • InjuryDec 15, 2020Loss of Osseointegration; Loss of Osseointegration

    PER (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 11012260 ↗
  • InjuryJul 21, 2020Failure to Osseointegrate; Failure to Osseointegrate

    AS PER (B)(4) AFTER A CLINICAL PROCEDURE A DENTAL IMPLANT DISPLAYED A FAILURE OF OSSEOINTEGRATION AND THE IMPLANT WAS EXPLANTED.

    Patient problems: Failure of Implant; Inflammation; Osteopenia/ Osteoporosis; Unspecified Infection

    MAUDE report 10305160 ↗
  • InjuryFeb 21, 2020Loss of Osseointegration; Loss of Osseointegration

    PER (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Fibrosis; Inflammation; Organ Dehiscence; Osteopenia/ Osteoporosis; Unspecified Infection

    MAUDE report 9736862 ↗
  • InjuryFeb 5, 2020Osseointegration Problem; Osseointegration Problem

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Inflammation; Pain; Swelling; Unspecified Infection

    MAUDE report 9669637 ↗
  • InjuryNov 13, 2019Osseointegration Problem; Osseointegration Problem

    PER COMPLAINT (B)(4), PATIENT EXPERIENCED LACK OF PRIMARY STABILITY.

    Patient problems: Failure of Implant; No Consequences Or Impact To Patient

    MAUDE report 9313693 ↗
  • InjuryNov 13, 2019Loss of Osseointegration; Loss of Osseointegration

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant; Inflammation; Unspecified Infection

    MAUDE report 9313685 ↗
  • InjuryOct 18, 2019Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), DURING POST OPERATIVE CHECK PATIENT EXPERIENCED FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant; No Consequences Or Impact To Patient

    MAUDE report 9207897 ↗
  • MalfunctionOct 7, 2019Inaccurate Information; Inaccurate Information

    PER COMPLAINT (B)(4), DURING A LOT CHECK, 4 IMPLANTS WERE DISCOVERED TO BE MISPACKAGED.

    Patient problems: No Patient Involvement

    MAUDE report 9164415 ↗
  • InjurySep 10, 2019Loss of Osseointegration; Loss of Osseointegration

    PER COMPLAINT (B)(4), DURING POST OPERATIVE CHECK, PATIENT EXPERIENCED LOSS OR FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant; No Consequences Or Impact To Patient

    MAUDE report 8979685 ↗
  • InjuryNov 29, 2018Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PER COMPLAINT (B)(4), THE PATIENT EXPERIENCE INTERMITTENT PAINFUL SENSATION APICALLY AND FACIALLY. THE PATIENT WANTED THE IMPLANT EXPLANTED AFTER SURGICAL RE-ENTRY DID NOT RESOLVE ISSUE.

    Patient problems: Failure of Implant; Inflammation; Osteopenia/ Osteoporosis; Pain; Unspecified Infection

    MAUDE report 8116956 ↗
  • InjuryNov 19, 2018Patient Device Interaction Problem; Patient Device Interaction Problem

    PER COMPLAINT (B)(4), THE PATIENT STATED THAT THEY FELT NUMB. THE DOCTOR NOTED POSSIBLE NERVE IMPINGEMENT WITH IMPLANT APPROXIMATELY TO THE INFERIOR ALVEOLAR. THE IMPLANT WAS EXTRACTED.

    Patient problems: Failure of Implant; Numbness

    MAUDE report 8085550 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K090234. One submission often covers many models, so check the product and lot details in each record.

  • Z-1157-2020 · Oct 11, 2019 · Terminated
    Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmD Platform, RxOnly, Sterile R,
  • Z-1041-2020 · Oct 7, 2019 · Terminated
    Legacy 3 Implant, Size: 4.2mmD x 13mmL, 3.5mmD Platform, REF: 864213
  • Z-0364-2020 · Aug 9, 2019 · Terminated
    Implant Direct, REF: 833708, Legacy 2 Implant, Surface: HA, 3.7mmD x 8mmL, 3.5mmD Platform, RxOnly, Sterile R, Made in USA Dental Implants
  • Z-0018-2020 · Aug 8, 2018 · Terminated
    ImplantDirect Legacy 3 Implant, REF 855216, SBM implant assembly, 2mmD x 16mmL, 4.5mmD platform, Rx, Sterile. The mislabeled model/part number on the vial cap is 835216. The re
  • Z-1167-2018 · Jan 19, 2018 · Terminated
    SwishTapered Implant Intended for use in support for fixed bridgework.
  • Z-0185-2018 · Oct 10, 2017 · Terminated
    ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for
  • Z-0007-2018 · Jul 21, 2017 · Terminated
    Implant Direct Swish Duo-Grip Healing Collar, Length: 3mm, Platform: 4.8 mm, Rx only, Sterile Dental Implant component.
  • Z-1112-2017 · Dec 12, 2016 · Terminated
    SwishTapered Implant
  • Z-0961-2017 · Nov 28, 2016 · Terminated
    GoDirect Screw-Retaining Top Part Numbers: 4500-61
  • Z-1920-2018 · Nov 15, 2016 · Terminated
    ImplantDirect Legacy 4 Implant, 5.7mmD x 5.7mmD Platform, Part 885708, Rx, Sterile. The incorrect cap label shows part 855708, 5.7mmD Platform, 8mmL.
  • Z-1267-2016 · Mar 4, 2016 · Terminated
    The Spectra-System Dental Implant 2008 system is comprised of dental implant fixtures and prosthetic devices that compose a two-piece implant system. GoDirect Implants Part Numbe

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates