Skip to content

ReActive® Implant 755710Recalls & reported problems

Implant Direct Sybron Manufacturing LLC · Ø3.7 x 8 Platform 3.5mmD · Class 2 · Implant

All models of ReActive Implant (22)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This ReActive® Implant 755710 is a small, root-form dental implant used as an anchor for a replacement tooth. The 60 reports mostly describe the implant failing to fuse with the surrounding jawbone or losing that fusion, along with difficulty getting a stable position during placement;59 reports involved an injury and there were no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
60
Deaths reported
0
Injuries reported
59
Recalls naming it
0

Summary

The FDA's MAUDE database contains 60 adverse event reports that list this exact device (DI 10841307103745), received between Sep 5, 2019 and Apr 13, 2026 — classified as 59 injuries, 1 malfunction.

The most frequently reported specific device problem is “Failure to Osseointegrate” (28 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 4 reportsMar 2024: 3 reportsApr 2024: 6 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate28
  • Positioning Failure9
  • Loss of Osseointegration6
  • Osseointegration Problem6
  • Premature Separation6
  • Positioning Problem3
  • Component Incompatible2
  • Fracture1

Patient problems

  • Failure of Implant43
  • No Clinical Signs, Symptoms or Conditions13
  • Osteopenia/ Osteoporosis12
  • Inflammation11
  • Inadequate Osseointegration10
  • No Consequences Or Impact To Patient6
  • Fibrosis4
  • Unspecified Infection4

Recent adverse event reports

The 25 most recent of 60 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 13, 2026Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24861228 ↗
  • InjuryFeb 18, 2026Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24383987 ↗
  • InjuryNov 4, 2025Positioning Failure; Osseointegration Problem; Positioning Failure; Osseointegration Problem

    LACK OF STABILITY DURING PLACEMENT.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23461408 ↗
  • InjuryMar 10, 2025Osseointegration Problem; Osseointegration Problem

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21567895 ↗
  • InjuryFeb 7, 2025Component Incompatible; Component Incompatible

    FAILURE TO OSSEOINTEGRATE

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338323 ↗
  • InjuryFeb 7, 2025Component Incompatible; Component Incompatible

    FAILURE TO OSSEOINTEGRATE

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338321 ↗
  • InjurySep 13, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRIMARY STABILITY COULDN¿T BE ACHIEVED.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20217951 ↗
  • InjurySep 13, 2024Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20214765 ↗
  • InjuryApr 23, 2024Osseointegration Problem; Osseointegration Problem

    PART/INTERFACE DOES NOT FIT/ALIGN.

    Patient problems: Failure of Implant

    MAUDE report 19167213 ↗
  • InjuryApr 23, 2024Osseointegration Problem; Osseointegration Problem

    PART/INTERFACE DOES NOT FIT/ALIGN.

    Patient problems: Failure of Implant

    MAUDE report 19167030 ↗
  • InjuryApr 16, 2024Positioning Failure; Positioning Failure

    PRIMARY STABILITY COULDN¿T BE ACHIEVED.

    Patient problems: Inflammation

    MAUDE report 19119545 ↗
  • InjuryApr 15, 2024Positioning Failure; Positioning Failure

    TRAUMA / CLINICAL COMPLICATION ONLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19110244 ↗
  • InjuryApr 3, 2024Premature Separation; Premature Separation

    PART NOT RETAINED ON MATING PART (E.G., DROPPED FROM DRIVER).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19035392 ↗
  • InjuryApr 2, 2024Premature Separation; Premature Separation

    DROPPED FROM THE IMPLANT DRIVER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19027790 ↗
  • InjuryMar 26, 2024Premature Separation; Premature Separation

    PART NOT RETAINED ON MATING PART (E.G.,

    Patient problems: Failure of Implant

    MAUDE report 18982069 ↗
  • InjuryMar 26, 2024Premature Separation; Premature Separation

    PART NOT RETAINED ON MATING PART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18982054 ↗
  • InjuryMar 2, 2024Premature Separation; Premature Separation

    PART NOT RETAINED ON MATING PART.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18823332 ↗
  • InjuryFeb 13, 2024Premature Separation; Premature Separation

    COMPL:DROPPED FROM THE IMPLANT DRIVER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18698503 ↗
  • InjuryFeb 13, 2024Positioning Failure; Positioning Failure

    PRIMARY STABILITY COULDN¿T BE ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18698494 ↗
  • InjuryFeb 13, 2024Positioning Failure; Positioning Failure

    PRIMARY STABILITY COULDN¿T BE ACHIEVED.

    Patient problems: Failure of Implant

    MAUDE report 18698493 ↗
  • InjuryFeb 13, 2024Positioning Failure; Positioning Failure

    PRIMARY STABILITY COULDN¿T BE ACHIEVED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18698492 ↗
  • InjuryJan 12, 2024Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE OF OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 18506148 ↗
  • InjuryOct 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT 203372452, AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Inflammation; Osteopenia/ Osteoporosis

    MAUDE report 17958048 ↗
  • InjuryOct 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Inflammation; Osteopenia/ Osteoporosis

    MAUDE report 17958046 ↗
  • InjuryOct 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Inflammation; Osteopenia/ Osteoporosis

    MAUDE report 17958038 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page