Quick summary

This Simply Legacy 3 dental implant is a small screw-shaped root-form implant used to replace a tooth. The 34 reports mostly describe the implant failing to bond with the surrounding jawbone, along with pain, slow healing, infection, and the part not staying secure during placement. There were no deaths, and 34 injuries. No recalls name this device. Reports rose from 1 in the earlier 12 months to 6 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
34
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10841307107873), received between Sep 11, 2019 and Aug 27, 2026 — classified as 34 injuries.

The most frequently reported specific device problem is “Failure to Osseointegrate” (26 reports).

2 other recalls cover devices cleared under the same FDA submission (K192221); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate26
  • Loss of Osseointegration3
  • Positioning Failure3
  • Osseointegration Problem1
  • Patient-Device Incompatibility1
  • Premature Separation1

Patient problems

  • Failure of Implant26
  • No Clinical Signs, Symptoms or Conditions7
  • Inadequate Osseointegration6
  • Pain6
  • Unspecified Infection4
  • Impaired Healing3
  • No Consequences Or Impact To Patient3
  • Osteopenia/ Osteoporosis3

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 27, 2026Failure to Osseointegrate

    Failure to osseointegrate.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50484015 ↗
  • InjuryApr 13, 2026Positioning Failure

    LACK OF STABILITY DURING PLACEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24853369 ↗
  • InjuryDec 15, 2025Loss of Osseointegration; Loss of Osseointegration

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23810997 ↗
  • InjuryDec 10, 2025Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23772241 ↗
  • InjuryNov 5, 2025Premature Separation; Premature Separation

    PART NOT RETAINED ON MATING PART (E.G.).

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23474687 ↗
  • InjuryNov 5, 2025Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23474525 ↗
  • InjuryNov 15, 2024Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE OF INTEGRATION.

    Patient problems: Inadequate Osseointegration; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20699283 ↗
  • InjuryOct 1, 2024Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration; Unspecified Infection

    MAUDE report 20347019 ↗
  • InjurySep 11, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRODUCT COMPL: FAILURE TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 20193785 ↗
  • InjuryAug 3, 2024Failure to Osseointegrate; Failure to Osseointegrate

    PRODUCT COMPL: FAILURE TO OSSEOINTEGRATE

    Patient problems: Abscess; Failure of Implant

    MAUDE report 19901992 ↗
  • InjuryMay 23, 2024Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 19386061 ↗
  • InjuryMay 23, 2024Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 19386037 ↗
  • InjuryAug 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE. PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 17571756 ↗
  • InjuryAug 18, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 17571738 ↗
  • InjuryFeb 3, 2023Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Pain

    MAUDE report 16288959 ↗
  • InjuryJun 27, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Granuloma

    MAUDE report 14829930 ↗
  • InjuryMay 7, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT 81156, PATIENT EXPERIENCED FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant; Pain

    MAUDE report 14317327 ↗
  • InjuryApr 27, 2022Positioning Failure; Positioning Failure

    PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, PATIENT EXPERIENCED LACK OF PRIMARY STABILITY.

    Patient problems: Failure of Implant

    MAUDE report 14219639 ↗
  • InjuryMar 15, 2022Osseointegration Problem; Osseointegration Problem

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED LOSS OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Inflammation; Pain

    MAUDE report 13770970 ↗
  • InjuryFeb 24, 2022Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Wound Dehiscence

    MAUDE report 13589644 ↗
  • InjuryNov 30, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 12893055 ↗
  • InjuryOct 8, 2021Positioning Failure; Positioning Failure

    PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, PATIENT EXPERIENCED LACK OF PRIMARY STABILITY.

    Patient problems: Failure of Implant

    MAUDE report 12598046 ↗
  • InjuryJul 13, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 12158711 ↗
  • InjuryJul 6, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant; Granuloma; Unspecified Infection

    MAUDE report 12120740 ↗
  • InjuryMar 19, 2021Failure to Osseointegrate; Failure to Osseointegrate

    PER COMPLAINT (B)(4), POST OPERATION, PATIENT EXPERIENCED FAILURE OF IMPLANT TO INTEGRATE.

    Patient problems: Failure of Implant; Impaired Healing; Inflammation; Osteopenia/ Osteoporosis; Pain; Unspecified Infection

    MAUDE report 11529345 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K192221. One submission often covers many models, so check the product and lot details in each record.

  • Z-0198-2022 · Sep 23, 2021 · Open, Classified
    ImplantDirect, REF 834713, Legacy 2 Implant 4.7mmD x 13mmL, 4.5mmD Platform,
  • Z-0197-2022 · Sep 23, 2021 · Open, Classified
    ImplantDirect, REF 834711, Legacy 2 Implant 4.7mmD x 11.5mmL, 4.5mmD Platform,

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →