Spectra-System®/ Legacy™ Guided Surgery Handle Kit G-HKRecalls & reported problems
Implant Direct Sybron Manufacturing LLC · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10841307119210), received between Jun 9, 2017 and Dec 1, 2021 — classified as 4 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Separation Failure3
- Device-Device Incompatibility1
- Failure to Advance1
- Failure to Osseointegrate1
- Fracture1
- Material Separation1
- Nonstandard Device1
- Osseointegration Problem1
Patient problems
- Failure of Implant5
- No Consequences Or Impact To Patient3
- No Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryDec 1, 2021Osseointegration Problem; Osseointegration Problem
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, DEFORMATION WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 12909083 ↗MalfunctionMar 11, 2021Separation Failure; Separation Failure
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, GUIDED SURGICAL KITS AND INSERTS ARE BINDING.
Patient problems: Failure of Implant
MAUDE report 11458466 ↗InjuryDec 29, 2020Fracture; Fracture
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, BROKEN OR FRACTURED COMPONENT WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 11087759 ↗MalfunctionFeb 19, 2020Separation Failure; Separation Failure
PER COMPLAINT (B)(4), SURGICAL GUIDE HANDLE IS APPARENTLY WARPED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9725527 ↗InjuryNov 20, 2019Failure to Osseointegrate; Failure to Osseointegrate
PER COMPLAINT (B)(4), COMPONENT WAS FRACTURED OR BROKEN AND COULD NOT BE SEPARATED.
Patient problems: Failure of Implant; No Information
MAUDE report 9346769 ↗MalfunctionNov 6, 2019Separation Failure; Separation Failure
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE IT WAS OBSERVED THAT THE COMPONENTS COULD NOT BE SEPARATED.
Patient problems: Failure of Implant; No Consequences Or Impact To Patient
MAUDE report 9283253 ↗MalfunctionJun 13, 2019Material Separation; Material Separation
PER COMPLAINT (B)(4), A G-SLEEVE INSTRUMENT RING POPPED OUT WHILE DRILLING A PATIENT'S BONE FOR DENTAL IMPLANT PLACEMENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. THE PROCEDURE WAS COMPLETED WITHIN THE SAME VISIT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8696587 ↗InjuryNov 1, 2017Device-Device Incompatibility; Device-Device Incompatibility
DRILLS BECAME STUCK IN THE GUIDED HANDLES DURING SURGICAL PROCEDURE. THIS CAUSED THE OSTEOTOMY TO BE INACCURATELY DRILLS AND THE IMPLANT SPUN DUE TO OVER "PREPERATION" CAUSE BY THE GUIDED HANDLE MALFUNCTION.
Patient problems: No Information
MAUDE report 6995239 ↗MalfunctionJun 9, 2017Nonstandard Device; Failure to Advance
THE 2.3 GUIDE INSERT WILL NOT ALLOW THE 2.3 DRILL TO PASS THROUGH, THE DIAMETER IS TOO SMALL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6632595 ↗
Other Accessories, Implant, Dental, Endosseous devices
- N/A 68015189Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous388 reports
- N/A 68015191Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous319 reports
- N/A 68015193Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous283 reports
- N/A 68015187Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous266 reports
- N/A 68015188Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous185 reports
- N/A 68015192Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous159 reports
- N/A 68015190Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous147 reports
- NobelReplace® 37045Nobel Biocare AB · Accessories, implant, dental, endosseous2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page