Hex Tool HT1.25Recalls & reported problems
Implant Direct Sybron Manufacturing LLC · Class 1
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10841307119609), received between Jul 5, 2019 and Sep 7, 2023 — classified as 12 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Fracture” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture9
- Material Deformation2
- Break1
- Defective Device1
- Loss of Osseointegration1
- Material Integrity Problem1
- Osseointegration Problem1
- Positioning Problem1
Patient problems
- Failure of Implant15
- No Consequences Or Impact To Patient3
- No Information2
- Granuloma1
- Insufficient Information1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
InjurySep 7, 2023Fracture; Fracture
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, BROKEN OR FRACTURED COMPONENT WAS OBSERVED
Patient problems: Failure of Implant
MAUDE report 17697631 ↗InjuryFeb 17, 2023Fracture; Fracture
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, BROKEN OR FRACTURED COMPONENT WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 16399252 ↗InjuryFeb 17, 2023Fracture; Fracture
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, BROKEN OR FRACTURED COMPONENT WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 16397403 ↗InjuryJan 10, 2023Fracture; Fracture
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, BROKEN OR FRACTURED COMPONENT WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 16128791 ↗InjurySep 5, 2022Material Deformation; Material Deformation
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, DEFORMATION WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 15355350 ↗InjuryMay 11, 2022Material Deformation; Material Deformation
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, PRODUCT FRACTURED.
Patient problems: Failure of Implant
MAUDE report 14366608 ↗InjuryMar 24, 2022Fracture; Fracture
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, PRODUCT FRACTURE WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 13891381 ↗InjuryMar 7, 2022Loss of Osseointegration; Loss of Osseointegration
PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PRODUCT FRACTURE WAS OBSERVED.
Patient problems: Failure of Implant; Granuloma
MAUDE report 13685244 ↗InjuryOct 12, 2021Fracture; Fracture
PER COMPLAINT (B)(4), PRODUCT FRACTURE WAS OBSERVED.
Patient problems: Failure of Implant
MAUDE report 12615130 ↗InjuryMay 26, 2021Separation Failure; Osseointegration Problem; Separation Failure; Osseointegration Problem
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, PRODUCT FRACTURE WAS OBSERVED.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 11884379 ↗InjuryOct 22, 2020Fracture; Fracture
PER COMPLAINT (B)(4), PRODUCT FRACTURED IS EXPERIENCED.
Patient problems: Failure of Implant
MAUDE report 10716872 ↗InjuryOct 22, 2020Fracture; Fracture
PER COMPLAINT (B)(4), PRODUCT FRACTURED IS EXPERIENCED.
Patient problems: Failure of Implant
MAUDE report 10716870 ↗MalfunctionDec 11, 2019Positioning Problem; Positioning Problem
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE, THE DRIVER GOT FRACTURED.
Patient problems: Failure of Implant; No Information
MAUDE report 9449797 ↗MalfunctionNov 28, 2019Defective Device; Defective Device
PER COMPLAINT (B)(4), HT1.25S AND HT1.25 HEX TOOL BENT WHILE REMOVING HEALING CAP.
Patient problems: Failure of Implant; No Consequences Or Impact To Patient
MAUDE report 9389854 ↗MalfunctionOct 11, 2019Fracture; Fracture
PER COMPLAINT (B)(4), DURING CLINICAL PROCEDURE IT WAS OBSERVED THAT THE DRIVER HAD BROKEN WHILE TORQUING IMPLANT.
Patient problems: Failure of Implant; No Consequences Or Impact To Patient
MAUDE report 9178184 ↗MalfunctionJul 26, 2019Break; Break
PER COMPLAINT (B)(4), A DENTAL IMPLANT DRIVER SNAPPED IN HALF WHILE THE DENTIST WAS WORKING IN THE PATIENT'S MOUTH. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED .
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8833901 ↗MalfunctionJul 5, 2019Material Integrity Problem; Material Integrity Problem
PER COMPLAINT (B)(4), DURING THE CLINICAL PROCEDURE, THE DOCTOR COULD NOT REMOVE THE IMPRESSION POST AND STRIPPED 3 DRIVERS. PER THE COMPLAINT, THE DOCTOR STATES THAT HE DID NOT USE A TORQUE WRENCH TO TIGHTEN SCREW BUT WHEN HE TRIED TO REMOVE THE SCREW FROM THE IMPLANT IT WAS STUCK. HE HAS STRIPPED OR BROKEN MULTIPLE TIPS AND TRIED THE ULTRA SONIC TO TRY TO BREAK UP ANYTHING WEDGING THE ABUTMENT IN PLACE, NOTHING WORKED. ADDITIONALLY, THE DOCTOR HAS APPLIED TO UP TO 60 NCM OF COUNTERCLOCKWISE TORQUE AND THE SCREW WOULDN'T MOVE. THE IMPLANT WAS EXTRACTED.
Patient problems: No Information
MAUDE report 8765249 ↗
Other Accessories, Implant, Dental, Endosseous devices
- N/A 68015189Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous388 reports
- N/A 68015191Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous319 reports
- N/A 68015193Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous283 reports
- N/A 68015187Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous266 reports
- N/A 68015188Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous185 reports
- N/A 68015192Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous159 reports
- N/A 68015190Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous147 reports
- NobelReplace® 37045Nobel Biocare AB · Accessories, implant, dental, endosseous2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.