simplyLegacy™ DRIVER HT2.5SGRecalls & reported problems
Implant Direct Sybron Manufacturing LLC · Class 1
All models of simplyLegacy DRIVER (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10841307119869), received between Mar 15, 2019 and Jan 24, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Failure to Osseointegrate1
- Material Deformation1
- Separation Failure1
Patient problems
- Failure of Implant2
- No Information2
- Unspecified Infection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 24, 2022Failure to Osseointegrate; Failure to Osseointegrate
PER COMPLAINT (B)(4), AFTER CLINICAL PROCEDURE, PATIENT EXPERIENCED FAILURE OF IMPLANT TO INTEGRATE.
Patient problems: Failure of Implant; Unspecified Infection
MAUDE report 13336801 ↗MalfunctionJun 7, 2019Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
PER COMPLAINT (B)(4), THE IMPLANT DRIVER IS TOO RETENTIVE. THE CLINICIAN HAS TO HOLD THE CARRIER WITH RONGEURS TO REMOVE THE MOUNT. WHEN PLACING THE DRIVER DIRECTLY IN THE IMPLANT, THE IMPLANT COMES OUT WHEN TOOL REMOVAL IS ATTEMPTED. IN SOFT BONE, THE IMPLANT PULLS OUT. ADDITIONALLY, PER THE COMPLAINT, THE COLOR CODING ON THE TOOLS HAS WORN OUT.
Patient problems: No Information
MAUDE report 8679971 ↗MalfunctionMar 15, 2019Separation Failure; Material Deformation; Separation Failure; Material Deformation
PER COMPLAINT (B)(4), A DRIVER COULD NOT BE REMOVED FROM THE DENTAL IMPLANT TRANSFER DURING INITIAL PLACEMENT. THE IMPLANT WAS REMOVED AND REPLACED WITH A NEW ONE WITHIN THE SAME VISIT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Failure of Implant; No Information
MAUDE report 8425228 ↗
Other Accessories, Implant, Dental, Endosseous devices
- N/A 68015189Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous388 reports
- N/A 68015191Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous319 reports
- N/A 68015193Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous283 reports
- N/A 68015187Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous266 reports
- N/A 68015188Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous185 reports
- N/A 68015192Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous159 reports
- N/A 68015190Dentsply Implants Manufacturing GmbH · Accessories, implant, dental, endosseous147 reports
- NobelReplace® 37045Nobel Biocare AB · Accessories, implant, dental, endosseous2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page