Xprezzon Xprezzon Monitor (REF 91393)Recalls & reported problems
Spacelabs Healthcare (Washington), Inc · XPREZZON MONITOR,91393 · Class 2
All models of Xprezzon (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10841522100079), received between Jul 10, 2025 and Dec 16, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K112962); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
- Power Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 16, 2025Power Problem; Power Problem
THE CUSTOMER REPORTED THAT WHILE MONITORING A PATIENT AN XPREZZON BEDSIDE MONITOR SHUT DOWN, WOULD NOT BOOT BACK UP, AND WOULD EMIT A CONSTANT BEEPING TONE. THE CUSTOMER SHIPPED THE UNIT TO SPACELABS HEALTHCARE FOR DEPOT REPAIR. THE UNIT WAS EVALUATED BY AN EQUIPMENT SERVICE CENTER TECHNICIAN WHO VERIFIED THE REPORTED FAILURE. THE ISSUE WAS IDENTIFIED TO BE A RESULT OF A FAULTY VIDEO PCBA. THE UNIT WAS REPAIRED BY THE TECHNICIAN AND RETURNED TO THE CUSTOMER. THE ISSUE WAS DUE TO A FAULTY VIDEO PCBA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23824267 ↗MalfunctionJul 10, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED THAT THEIR XPREZZON BEDSIDE MONITOR HAD CRASHED WHILE IN USE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE BIOMEDICAL ENGINEER HAD CONFIRMED THAT THE DEVICE WAS FAILING THE POWER ON SELF TEST FOR A DRAM ISSUE PRIOR TO PUTTING THE UNIT INTO USE, BUT THEY HAD PUT IT INTO SERVICE REGARDLESS. THE DEVICE HAS SINCE BEEN REMOVED FROM SERVICE AND WILL BE RETURNED TO SPACELABS HEALTHCARE FOR ADDITIONAL INVESTIGATION. AT THIS TIME THE DEVICE HAS NOT BEEN RECEIVED. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22476009 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112962. One submission often covers many models, so check the product and lot details in each record.
- Z-0169-2026 · Sep 5, 2025 · Open, ClassifiedBrand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0893-01 Software Version: N/A Product Description: 16 inch DVI-I to
- Z-0167-2026 · Sep 5, 2025 · Open, ClassifiedBrand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to D
- Z-0168-2026 · Sep 5, 2025 · Open, ClassifiedBrand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-12 Software Version: N/A Product Description: 12 foot DVI-I to
- Z-1144-2015 · Jan 28, 2015 · TerminatedSpacelabs Healthcare XPREZZON Bedside Monitor, Model 91393. The Spacelabs Healthcare XPREZZON Bedside Monitor passively displays data generated by Spacelabs parameter modules, Fle
- Z-0196-2013 · Oct 5, 2012 · TerminatedThe touchscreen display (model 94267) designed for use with the XPREZZON bedside monitor (model 91393) is width-452 mm (17.2 inches) and height- 368 mm (14.5 inches). The depth wit
- Z-2019-2012 · Jun 12, 2012 · TerminatedSpacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzo
- Z-1705-2012 · May 22, 2012 · TerminatedProduct is the Spacelabs Medical Xprezzon Bedside Monitor, Model 91393. The Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacela
Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page