AriaTele 96281Recalls & reported problems

Spacelabs Healthcare (Washington), Inc · ARIATELE TELEMETRY TRANSMITTER · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10841522100352), received between Jan 9, 2026 and Feb 20, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K121480); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Alarm System2
  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hypoxia1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026Device Alarm System

    THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TO REQUEST REPAIR FOR AN ARIATELE THAT FAILED TO ALARM. NO PATIENT OR USER HARM WAS REPORTED. SPACELABS HEALTHCARE TECHNICAL SUPPORT ATTEMPTED TO CONTACT THE CUSTOMER FOR ADDITIONAL INFORMATION REGARDING REPORTED FAILURE TO ALARM, HOWEVER THE CUSTOMER HAS NOT RESPONDED TO ATTEMPTS TO GET ADDITIONAL INFORMATION. THE DEVICE HAS BEEN RECEIVED, HOWEVER HAS NOT YET BEEN EVALUATED. A FOLLOW-UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24408320 ↗
  • InjuryFeb 2, 2026Off-Label Use

    THE CUSTOMER REPORTED TO SPACELABS HEALTHCARE THAT A PATIENT INCIDENT OCCURRED WITH AN ARIATELE DEVICE. DURING THE EVENT, THE PATIENT APPEARED TO BE HYPOXIC DESPITE NO INDICATIONS OF LOW OXYGEN SATURATION BY THE ARIATELE. THIS WAS NOTICED BY CLINICAL STAFF AND A SECONDARY HANDHELD PULSE OXIMETER WAS USED TO CONFIRM THE PATIENT WAS HYPOXIC. TREATMENT WAS ADMINISTERED AND THE PATIENT SURVIVED. DURING TROUBLESHOOTING, THE TECHNICAL SUPPORT REPRESENTATIVE CONFIRMED THAT THE CUSTOMER WAS USING AN SP02 SENSOR WHICH IS NOT COMPATIBLE WITH THE ARIATELE TRANSMITTER. THE CUSTOMER WAS PROVIDED DOCUMENTATION THAT STATES THE SENSOR WAS INCOMPATIBLE AND TO ONLY USE APPROVED SENSORS GOING FORWARD. THE CAUSE OF THE REPORTED ISSUE WAS DUE TO THE USE OF INCOMPATIBLE SP02 SENSORS WHICH PROVIDED INACCURATE RESULTS.

    Patient problems: Hypoxia

    MAUDE report 24229104 ↗
  • MalfunctionJan 9, 2026Device Alarm System

    THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TO REPORT THAT THEY HAD AN ARIATELE TRANSMITTER THAT HAD AN ALARM THAT WAS NOT WORKING PROPERLY. SPACELABS HEALTHCARE TECHNICAL SUPPORT MADE SEVERAL ATTEMPTS TO GET ADDITIONAL INFORMATION REGARDING THE ALARM ISSUE; HOWEVER, NO FURTHER DETAILS WERE AVAILABLE. THE CUSTOMER SHIPPED THE DEVICE TO SPACELABS HEALTHCARE FOR DEPOT REPAIR. THE UNIT WAS TESTED BY AN EQUIPMENT SERVCE CENTER TECHNICIAN. THE TECHNICIAN COULD NOT VERIFY THE PROBLEM OF 'WILL NOT ALARM' BECAUSE THE DEVICE IS NOT CAPABLE OF PRODUCING ALARMS. HOWEVER, DID VERIFY A DISPLAY RELATED FAILURE. THE UNIT HAD A FAULTY DISPLAY PCB CAUSING A BLANK SCREEN AND DAMAGE TO THE CASE. THE UNIT WAS REPAIRED AND AFTER PASSING ALL FINAL FUNCTIONAL TESTING, THE UNIT WAS RETURNED TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24031212 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121480. One submission often covers many models, so check the product and lot details in each record.

  • Z-2586-2014 · Jul 2, 2014 · Terminated
    ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Physiological patient monitor. Used in a hospital environment.
  • Z-0871-2014 · Dec 19, 2013 · Terminated
    ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with either adult or neonatal patient populations in a hospital environment.

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →