AriaTele 96281Recalls & reported problems
Spacelabs Healthcare (Washington), Inc · ARIATELE TELEMETRY TRANSMITTER · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10841522100352), received between Jan 9, 2026 and Feb 20, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K121480); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Alarm System2
- Off-Label Use1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Hypoxia1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2026Device Alarm System
THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TO REQUEST REPAIR FOR AN ARIATELE THAT FAILED TO ALARM. NO PATIENT OR USER HARM WAS REPORTED. SPACELABS HEALTHCARE TECHNICAL SUPPORT ATTEMPTED TO CONTACT THE CUSTOMER FOR ADDITIONAL INFORMATION REGARDING REPORTED FAILURE TO ALARM, HOWEVER THE CUSTOMER HAS NOT RESPONDED TO ATTEMPTS TO GET ADDITIONAL INFORMATION. THE DEVICE HAS BEEN RECEIVED, HOWEVER HAS NOT YET BEEN EVALUATED. A FOLLOW-UP REPORT WILL BE SUBMITTED IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24408320 ↗InjuryFeb 2, 2026Off-Label Use
THE CUSTOMER REPORTED TO SPACELABS HEALTHCARE THAT A PATIENT INCIDENT OCCURRED WITH AN ARIATELE DEVICE. DURING THE EVENT, THE PATIENT APPEARED TO BE HYPOXIC DESPITE NO INDICATIONS OF LOW OXYGEN SATURATION BY THE ARIATELE. THIS WAS NOTICED BY CLINICAL STAFF AND A SECONDARY HANDHELD PULSE OXIMETER WAS USED TO CONFIRM THE PATIENT WAS HYPOXIC. TREATMENT WAS ADMINISTERED AND THE PATIENT SURVIVED. DURING TROUBLESHOOTING, THE TECHNICAL SUPPORT REPRESENTATIVE CONFIRMED THAT THE CUSTOMER WAS USING AN SP02 SENSOR WHICH IS NOT COMPATIBLE WITH THE ARIATELE TRANSMITTER. THE CUSTOMER WAS PROVIDED DOCUMENTATION THAT STATES THE SENSOR WAS INCOMPATIBLE AND TO ONLY USE APPROVED SENSORS GOING FORWARD. THE CAUSE OF THE REPORTED ISSUE WAS DUE TO THE USE OF INCOMPATIBLE SP02 SENSORS WHICH PROVIDED INACCURATE RESULTS.
Patient problems: Hypoxia
MAUDE report 24229104 ↗MalfunctionJan 9, 2026Device Alarm System
THE CUSTOMER CONTACTED SPACELABS HEALTHCARE TO REPORT THAT THEY HAD AN ARIATELE TRANSMITTER THAT HAD AN ALARM THAT WAS NOT WORKING PROPERLY. SPACELABS HEALTHCARE TECHNICAL SUPPORT MADE SEVERAL ATTEMPTS TO GET ADDITIONAL INFORMATION REGARDING THE ALARM ISSUE; HOWEVER, NO FURTHER DETAILS WERE AVAILABLE. THE CUSTOMER SHIPPED THE DEVICE TO SPACELABS HEALTHCARE FOR DEPOT REPAIR. THE UNIT WAS TESTED BY AN EQUIPMENT SERVCE CENTER TECHNICIAN. THE TECHNICIAN COULD NOT VERIFY THE PROBLEM OF 'WILL NOT ALARM' BECAUSE THE DEVICE IS NOT CAPABLE OF PRODUCING ALARMS. HOWEVER, DID VERIFY A DISPLAY RELATED FAILURE. THE UNIT HAD A FAULTY DISPLAY PCB CAUSING A BLANK SCREEN AND DAMAGE TO THE CASE. THE UNIT WAS REPAIRED AND AFTER PASSING ALL FINAL FUNCTIONAL TESTING, THE UNIT WAS RETURNED TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24031212 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121480. One submission often covers many models, so check the product and lot details in each record.
- Z-2586-2014 · Jul 2, 2014 · TerminatedARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Physiological patient monitor. Used in a hospital environment.
- Z-0871-2014 · Dec 19, 2013 · TerminatedARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with either adult or neonatal patient populations in a hospital environment.
Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices
- IntelliVue MX40 1.4GHz MX40 1.4 GHz Smart HoppingPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall47 deaths2,541 reports
- NA PU-621RNihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)17 deaths1,666 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- NA PU-681RANihon Kohden Corporation · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)22 deaths1,493 reports
- IntelliVue Multi-Measurement Module X3 IntelliVue X3Philips Medizin Systeme Böblingen GmbH · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall19 deaths1,172 reports
- Patient Information Center IXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)82 deaths941 reports
- Patient Information Center iXPhilips Medical Systems Hsg · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)83 deaths814 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.