ACIST CVi® Angiographic Injection System Injector Head 014657Recalls & reported problems
Acist Medical Systems, Inc. · CVi Injector, p/n 700706-009 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 3
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10841716101233), received between Feb 1, 2018 and Jan 6, 2026 — classified as 3 deaths, 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Device Displays Incorrect Message1
Patient problems
- Air Embolism6
- Death2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
DeathJan 6, 2026Adverse Event Without Identified Device or Use Problem
DURING A CORONARY ANGIOGRAPHY, AFTER THE USER COMPLETED A FULL FLUSH OF THE CONSUMABLE TUBING AND CATHETER, AND CONFIRMED THAT THE SYSTEM WAS FUNCTIONING PROPERLY, THE FIRST CONTRAST MEDIA INJECTION WAS COMPLETED WITHOUT ANY ISSUES. DURING THE SECOND INJECTION OF CONTRAST MEDIA, APPROXIMATELY 6 CUBIC CENTIMETERS OF AIR WAS INJECTED INTO THE PATIENT. AIR BUBBLES WERE OBSERVED IN THE PATIENT'S LEFT CORONARY SIDE DURING THE INJECTION. THE PATIENT EXPIRED ON (B)(6) 2025.
Patient problems: Air Embolism
MAUDE report 23979826 ↗MalfunctionSep 30, 2024
PERSONNEL IN CARDIAC CATHETERIZATION LAB WERE CLEANING ROOM BETWEEN CASES, NOTED A SMELL AND SOME SMOKE AND A SMALL LINE OF "SPARK" AND BURN WITHIN THE WELL OF THE SYRINGE SPACE WAS NOTED. THE DEVICE WAS IMMEDIATELY UNPLUGGED. THE DEVICE WAS NOT BEING USED ON A PATIENT AT THE TIME OF THE INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20332644 ↗InjuryApr 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
USER FACILITY REPORTED THAT A (B)(6) MALE WITH NON-ST-ELEVATION MYOCARDIAL INFARCTION (NSTEMI) WAS TAKEN TO THE CATH LAB FOR LEFT HEART CATHETERIZATION (LHC) WITH PERCUTANEOUS CORONARY INTERVENTION (PCI). ACIST ANGIOGRAPHIC INJECTION SYSTEM, MODEL CVI WAS USED FOR CONTRAST INJECTIONS. UPON THE SIXTH INJECTION OF CONTRAST INTO THE PATIENT, AIR WAS VISIBLE WITHIN THE LEFT ANTERIOR DESCENDING ARTERY (LAD). THE PATIENT EXPERIENCED ACUTE CHRONIC CHEST PAIN FOR 1-2 MINUTES POST INJECTION AND BLOOD PRESSURE < 90 MMHG SYSTOLIC. THE PATIENT TOLERATED THE AIR INJECTION WITH NO DETERIORATION IN CLINICAL CONDITION.
Patient problems: Air Embolism
MAUDE report 14218923 ↗InjurySep 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AT THE START OF A CORONARY ANGIOGRAPHY PROCEDURE OF A (B)(6)YEAR OLD MALE, IT WAS NOTICED THAT THERE WAS APPROXIMATELY 10 ML OF AIR IN THE RIGHT CORONARY ARTERY (RCA) AND LEFT ANTERIOR DESCENDING ARTERY (LAD). THE PATIENT EXPERIENCED CHEST PAIN FOR A DURATION OF TEN MINUTES, ST SEGMENT ELEVATION, BRADYCARDIA, AND BLOOD PRESSURE < 90 MMHG SYSTOLIC. THE PATIENT RECOVERED THE SAME DAY.
Patient problems: Air Embolism
MAUDE report 12467162 ↗InjuryAug 31, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING AN ANGIOPLASTY PROCEDURE AT THE SECOND INJECTION DURING FINAL CONTROL, SOME LARGE BUBBLES OF AIR WERE INJECTED INTO A PATIENT. THE PATIENT EXPERIENCED CHEST PAIN AND ST SEGMENT ELEVATION FOR A DURATION OF A FEW MINUTES. THE PATIENT RECOVERED THE SAME DAY.
Patient problems: Air Embolism
MAUDE report 12394138 ↗InjuryJul 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A PERCUTANEOUS CORONARY INTERVENTION FOR LEFT ANTERIOR DESCENDING STENOSIS AND USING THE ACIST ANGIOGRAPHIC INJECTION SYSTEM, MODEL CVI, THE USER REPORTED THAT JUST AFTER THE CONTRAST INJECTION INTO PATIENT'S LEFT CIRCUMFLEX ARTERY, AIR INFUSION WAS OBSERVED IN THE LEFT ANTERIOR DESCENDING ARTERY (LAD). THE PATIENT'S ECG SHOWED ST SEGMENT DEPRESSION OF UNKNOWN DURATION. THE PATIENT SUBSEQUENTLY DIED ON (B)(6) 2021.
Patient problems: Air Embolism
MAUDE report 12245791 ↗InjuryApr 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A LEFT HEART CATHETERIZATION FOR CHEST PAIN, SLOW FLOW WAS NOTED IN THE LEFT CORONARY SYSTEM AFTER THE FIRST INJECTION. AIR EMBOLUS WAS SUSPECTED BUT NOT VERIFIED. THE PATIENT EXPERIENCED CHEST PAIN FOR A DURATION OF 15 MINUTES, HYPOTENSION AND BRADYCARDIA WITH RESOLUTION BY THE END OF THE PROCEDURE. THE PROCEDURE WAS COMPLETED AND THE PATIENT RECOVERED THE SAME DAY.
Patient problems: Air Embolism
MAUDE report 11737670 ↗DeathFeb 6, 2018Adverse Event Without Identified Device or Use Problem
USER FACILITY REPORTED: A (B)(6) FEMALE PATIENT WAS EXPERIENCING UNSTABLE ANGINA AND WAS UNDERGOING A CORONARY ANGIO. A CORONARY DISSECTION OCCURRED AFTER INJECTION INTO THE RIGHT CORONARY ARTERY (RCA). THE PATIENT EXPERIENCED SHOCK, CHEST PAIN, ST-SEGMENT ELEVATION, VENTRICULAR FIBRILLATION, HYPOTENSION, AND MYOCARDIAL INFARCTION. THE PATIENT EXPIRED (B)(6) 2017.
Patient problems: Death
MAUDE report 7245334 ↗DeathFeb 1, 2018Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A STEMI PROCEDURE WAS PERFORMED ON A (B)(6) YEAR OLD MALE PATIENT TRANSFERRED FROM THE EMERGENCY DEPARTMENT. AT THE VERY BEGINNING OF THE PROCEDURE, ERROR MESSAGES APPEARED, "AIR COLUMN DETECTED" AND "TURN OFF STOPCOCK TO THE PATIENT TO PURGE." ONE ENTIRE BOTTLE OF CONTRAST WAS PURGED. A NEW SET UP WAS DRAWN, AND THE SAME ERROR MESSAGES OCCURRED. AFTER THIS, MANUAL INJECTIONS WERE DONE. THE PATIENT EXPERIENCED CHEST PAIN, ABNORMAL HEART RHYTHM (VTACH, VFIB, ASYSTOLE). BLOOD PRESSURE <90 MMHG SYSTOLIC, MYOCARDIAL INFARCTION, AND CARDIAC BIOMARKER ELEVATION, TROP 11.9. THE OUTCOME OF THE EVENT WAS THAT THE PATIENT EXPIRED.
Patient problems: Death
MAUDE report 7234496 ↗
Other Injector And Syringe, Angiographic devices
- Monarch® 00884450298022Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls37 reports
- Blue Diamond™ 00884450498743Merit Medical Systems, Inc. · Injector and syringe, angiographic2 recalls0 reports
- Monarch® 00884450298053Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- Monarch® 00884450298060Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC2 recalls0 reports
- NAMIC 80085007Medline Industries, Inc. · Injector and syringe, angiographic1 recall36 reports
- NA 622510B. Braun Medical Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC85 reports
- NAMIC 80087007Medline Industries, Inc. · Injector and syringe, angiographic1 recall20 reports
- Blue Diamond™ 00884450381168Merit Medical Systems, Inc. · INJECTOR AND SYRINGE, ANGIOGRAPHIC1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page