Skip to content

Split Leg Section 4-030-61Recalls & reported problems

Skytron LLC · Split Leg Section for 6700 SERIES & 3600B · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10841736112745), received between May 20, 2025 and May 20, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    SKYTRON TECHNICAL SUPPORT RECEIVED A PHONE CALL FROM FROEDTERT MEMORIAL LUTHERAN HOSPITAL DESCRIBING AN INCIDENT INVOLVING A SKYTRON SPLIT LEG SECTION FOR 6700 SERIES & 3600B TABLE ACCESSORY. THE FACILITY ALLEGES THAT THEY WERE SET UP FOR A CASE AND WHEN THE TABLE WAS ADJUSTED TO REVERSE TRENDELENBURG POSITION THE SPLIT LEG ACCESSORY DISENGAGED FROM THE SURGICAL TABLE AND THE PATIENT SLID TO FOOT GOING INTO STRADDLE POSITION ON THE GROUND. NO PATIENT OR STAFF INJURIES WERE REPORTED AS A RESULT OF THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22056256 ↗

Other Surgical Table Cushion devices

All Surgical Table Cushion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page