Skip to content

STAPLIZER 7320-03Recalls & reported problems

Conmed Corporation · STAPLIZER BONE STAPLE CARTRIDGE, 13 X 20 MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10845854015681), received between Apr 17, 2024 and Apr 17, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2024Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE 7320-03, 13 X 20MM STAPLIZER BONE STAPLE CARTRIDGE, QTY 3 WAS BEING USED ON (B)(6) 2024 DURING A RIGHT KNEE REVISION ACL WITH QUADS ALLOGRAFT AND THE ¿CONMED STAPILIZER BONE STAPLE CARTRIDGE 13MMX 20MM IS NOT FIRING FROM THE REPROCESSABLE 3M/HALL STAPLE GUN. THE CARTRIDGE IS BREAKING OFF PIECES OF THE 13MM STAPLE CARRIER AND DEBRIS IS GOING INTO THE PATIENT. THIS HAPPENED WITH TWO CARTDRIGES ON TWO DIFFERENT STAPLE GUNS. DEBRIS WAS REMOVED FROM PATIENT'S KNEE¿. THERE WAS NO REPORT OF IMPACT OR INJURY TO THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED WITH ARTHREX STAPLES AND A 10 MINUTE DELAY. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19126877 ↗

Other STAPLE, FIXATION, BONE devices

All STAPLE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page