C-WIRE 00505004200Recalls & reported problems
Conmed Corporation · C-WIRE PAK DOUBLE ENDED ORTHOPAEDIC WIRE, SPADE, .035" (0.89 MM) · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10845854029091), received between Aug 23, 2024 and Aug 23, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 23, 2024Material Fragmentation; Material Fragmentation
THE DISTRIBUTOR REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, 00505004200, C-WIRE PAK DOUBLE ENDED ORTHOPAEDIC WIRE, WAS BEING USED ON (B)(6) 2024 DURING A FRACTURE OF THE DISTAL PHALANX OF THE MIDDLE FINGER AND THE ¿C-WIRE BROKE DURING INSERTION.¿. PER FURTHER ASSESSMENT IT WAS FOUND THAT THE FRAGMENTS DID FALL INTO THE SURGICAL SITE AND WERE RETRIEVED. THE PROCEDURE WAS COMPLETED WITH THE USE OF AN ALTERNATE BACKUP DEVICE. THERE WAS NO IMPACT OR INJURY FOR THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20047546 ↗
Other PIN, FIXATION, SMOOTH devices
- RIGIDFIX 210133Medos International Sàrl · PIN, FIXATION, SMOOTH91 reports
- SMART TOE ST0A-16PStryker GmbH · PIN, FIXATION, SMOOTH1 recall5 reports
- SMART TOE ST0-19PStryker GmbH · PIN, FIXATION, SMOOTH1 recall2 reports
- HammerTube™ System P40-002-0275-SParagon 28, Inc. · PIN, FIXATION, SMOOTH1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
- ARROW-LOK 1Arrowhead De LLC · Pin, Fixation, Smooth1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.