Skip to content

3100A HFOV 11518-850LRecalls & reported problems

Carefusion Corporation · BOXED PATIENT CIRCUIT ASSY, 31A, FILTERED, 850L · Class 3

All models of 3100A HFOV (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10846446003185), received between Jul 4, 2020 and Jul 4, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Defective Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Defective Component1

Patient problems

  • Respiratory Failure1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 4, 2020Defective Component; Defective Component

    THE CUSTOMER REPORTED THAT A PATIENT WAS BEING VENTILATED WITH AN MAP OF 14 CM/H20. HOWEVER, WITHIN 12 HOURS, THE MAP SPIKED TO 20 CM/H2O AND COULD NOT BE REDUCED BY THE ADJUSTMENT DIAL. UPON CHECKING, THE BOXED PATIENT CIRCUIT ASSY, 31A, FILTERED, 850L WAS SATURATED. ONCE IT WAS REMOVED, THE MAP RETURNED TO NORMAL. IT WAS LATER INDICATED THAT THE PATIENT DECOMPENSATED AND HAD TO BE DISCONNECTED FROM THE HFOV AND MANUALLY VENTILATED. CHEST X-RAY REVEALED RIGHT MIDDLE LOBE COLLAPSE AND LEFT-SIDED DIAPHRAGMATIC HERNIA.

    Patient problems: Respiratory Failure

    MAUDE report 10230993 ↗

Other Ventilator, High Frequency devices

All Ventilator, High Frequency devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page