Skip to content

Vmax 777405-101Recalls & reported problems

Carefusion Corporation · VMAX ENCORE 22D · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10846446021417), received between Jun 19, 2019 and Mar 31, 2021 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Mechanical Problem1
  • Patient Device Interaction Problem1
  • Smoking1

Patient problems

  • Injury1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2021Smoking; Smoking

    IT IS REPORTED TO VYAIRE MEDICAL THAT THE VMAX ENCORE 22 EXPERIENCED A BURNING SMELL UPON INSTALLATION OF SAMPLE PUMP. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11599994 ↗
  • MalfunctionOct 15, 2020Mechanical Problem; Mechanical Problem

    A CUSTOMER CONTACTED VYAIRE MEDICAL TO REPORT THAT THEY NOTED THE FAN INSIDE THE DELL PC WAS, "WHINING AND SMELLED HOT." UPON REPLACING THE PC, THEY STARTED GETTING GAS NOT CONNECTED WARNINGS EVEN THOUGH THEY WERE CONNECTED TO THE WALL O2. CUSTOMER WAS ADVISED TO PUT THE PC BACK TO THE ORIGINAL SYSTEM, AND REPLACE THE SOL CABLE FROM THE SYSTEM IN ORDER TO RESOLVE THE COMMUNICATION PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 10686886 ↗
  • InjuryJun 19, 2019Patient Device Interaction Problem; Patient Device Interaction Problem

    WHILE USING THE VMAX FOR PULMONARY FUNCTION TESTING, THE PATIENT'S LIPS WERE STICKING TO THE MOUTHPIECE AND CAUSING THEIR LIPS TO BLEED.

    Patient problems: Injury

    MAUDE report 8713841 ↗

Other Calculator, Predicted Values, Pulmonary Function devices

All Calculator, Predicted Values, Pulmonary Function devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page