Skip to content

WEBSTER D610DRP10RTRecalls & reported problems

Biosense Webster Inc · Class 2

All models of WEBSTER (29)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10846835000238), received between Feb 9, 2018 and Sep 11, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break3
  • Adverse Event Without Identified Device or Use Problem1
  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2023Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A SUPRAVENTRICULAR TACHYCARDIA (SVT) CARDIAC ABLATION PROCEDURE WITH A WEBSTER¿ ELECTROPHYSIOLOGY CATHETER AND DURING THE OPERATION, THE TIP OR SPLINES OF THE CATHETER WAS FOUND BENT (NO EXPOSED WIRES). A SECOND DEVICE WAS USED TO COMPLETE THE OPERATION. THERE WAS NO ADVERSE EVENT REPORTED ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17726512 ↗
  • MalfunctionJan 18, 2022Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A WEBSTER¿ ELECTROPHYSIOLOGY CATHETER AND A BROKEN TIP ISSUE OCCURRED. IT WAS REPORTED THAT DURING THE PROCEDURE, THE TIP OF THE CATHETER WAS FOUND BENT, AND WIRE WAS EXPOSED. A SECOND CATHETER WAS USED TO COMPLETE THE OPERATION. THERE WAS NO PATIENT ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13285960 ↗
  • MalfunctionDec 30, 2021Contamination /Decontamination Problem; Adverse Event Without Identified Device or Use Problem; Contamination /Decontamination Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH A WEBSTER¿ ELECTROPHYSIOLOGY CATHETER AND AN ISSUE OF FOREIGN MATERIAL WAS ENCOUNTERED. AN ISSUE OF FOREIGN MATTER WAS REPORTED ALONG WITH WHAT WAS DESCRIBED AS A ¿PROTRUSION¿ ON THE DISTAL TIP OF THE CATHETER. HOWEVER, REVIEW OF A PHOTO PROVIDED, IT WAS DETERMINED THAT THE PROTRUSION WAS ACTUALLY THE DEVICE¿S ANCHOR WINDOW AND NOT A PROTRUSION. THE ANCHOR WINDOW IS PART OF THE CATHETER¿S DESIGN. A SECOND CATHETER WAS USED TO COMPLETE THE OPERATION. THERE WAS NO PATIENT CONSEQUENCE SINCE THE DEVICE WAS NOT USED IN THE PATIENT. THIS EVENT IS CONSIDERED MDR REPORTABLE FOR THE FOREIGN MATTER ISSUE REPORTED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN CLARIFICATION TO THIS COMPLAINT. HOWEVER, NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13121586 ↗
  • MalfunctionFeb 9, 2018Break; Break

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PROCEDURE WITH A WEBSTER CATHETER AND A BROKEN TIP WITH INTERNAL COMPONENT EXPOSURE OCCURRED DURING THE PROCEDURE. THE CATHETER WAS CHANGED AND THE PROCEDURE WAS CONTINUED. THE PROCEDURE WAS COMPLETED WITH NO PATIENT CONSEQUENCE. THIS EVENT IS MDR REPORTABLE BECAUSE IF THE CATHETER INTEGRITY IS NOT MAINTAINED THERE IS A POTENTIAL RISK TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7259927 ↗

FDA clearance and classification

The device record lists FDA submission K892265; its details are not in the openFDA clearance files we use.

The FDA regulates product code DRF (catheter, electrode recording, or probe, electrode recording) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the WEBSTER D610DRP10RT

Has the WEBSTER D610DRP10RT been recalled?

No FDA recall record in our data names this exact device identifier (DI 10846835000238), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the WEBSTER D610DRP10RT?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Feb 9, 2018 and Sep 11, 2023. In those reports, the most common specific device problems are “Break” (3) and “Contamination /Decontamination Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the WEBSTER D610DRP10RT?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the WEBSTER D610DRP10RT, and is it still sold?

The FDA's device database lists Biosense Webster Inc as the company responsible for it (brand WEBSTER, model D610DRP10RT). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices

All CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page