EZ Steer ThermoCool BDI75TCFJRTRecalls & reported problems
Biosense Webster Inc · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10846835002898), received between Mar 24, 2016 and Apr 1, 2026 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Markings/Labelling Problem” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (P030031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Markings/Labelling Problem2
- Device Sensing Problem2
- No Display/Image1
- Patient Device Interaction Problem1
- Signal Artifact/Noise1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Cardiac Tamponade1
- No Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2026Device Sensing Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC ABLATION PROCEDURE WITH AN EZ STEER THERMOCOOL CATHETER FOR WHICH BIOSENSE WEBSTER¿S PRODUCT ANALYSIS LAB (PAL) IDENTIFIED THE PEEK HOUSING BROKEN. TEMPERATURE SENSOR ERROR. CHANGED CABLE AND CHANGED CATHETER. NO PATIENT CONSEQUENCE. THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION AND PER THE EVALUATION COMPLETION ON 09-MAR-2026, OBSERVED THE PEEK HOUSING BROKEN. THE EVENT WAS ORIGINALLY CONSIDERED NON-REPORTABLE, HOWEVER, BWI BECAME AWARE OF PEEK HOUSING BROKEN ON (B)(6) 2026 AND HAVE ASSESSED THIS RETURNED CONDITION AS REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24761323 ↗MalfunctionNov 7, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem
DURING AN ELECTROPHYSIOLOGY PROCEDURE AT THE HOSPITAL PROCOR, THE DOCTOR REQUESTED TO OPEN THE PRODUCT BIDIRECTIONAL IRRIGATED ABLATION CATHETER EZ STEER THERMOCOOL TC CURVE JF 7,5F, REFERENCE BDI75TCFJRT. WHEN OPENING THE INDICATED PRODUCT, WE FOUND THAT IT WAS NOT THE CORRECT ITEM. INSTEAD, WE FIND A PRODUCT SPECIFYING DF CURVE, INSIDE THE PACKAGE, BELONGING TO THE LOT 31081511M. IT IS IMPORTANT TO NOTE THAT THE DF CATHETER IS DESIGNED FOR SMALLER CURVE TECHNIQUES, WHILE THE JF IS USED FOR LARGER CURVE TECHNIQUES, WHICH CAN SIGNIFICANTLY IMPACT PROCEDURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20631649 ↗MalfunctionSep 12, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN AV NODE ABLATION PROCEDURE WITH A EZ STEER THERMOCOOL AND PACKAGING FOR THE EZ STEER THERMOCOOL CATHETER SHOWED IT WAS AN FJ CURVE BUT THE CATHETER HAD "DF" PRINTED ON THE ACTUAL CATHETER HANDLE. IT WAS REPORTED THAT THE PACKAGING FOR THE EZ STEER THERMOCOOL CATHETER SHOWED IT WAS AN FJ CURVE, BUT AFTER OPENING THE PACKAGING, IT WAS NOTICED THAT THE EZ STEER THERMOCOOL CATHETER HAD "DF" PRINTED ON THE ACTUAL CATHETER HANDLE, AND THE EZ STEER THERMOCOOL CATHETER WAS A DF CURVE, NOT AN FJ CURVE. THE PROCEDURE CONTINUED JUST USING THE EZ STEER THERMOCOOL DF CURVE CATHETER. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20208570 ↗InjuryJul 30, 2023Signal Artifact/Noise; Device Sensing Problem; Patient Device Interaction Problem; Signal Artifact/Noise; Device Sensing Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FLUTTER ABLATION PROCEDURE WITH A EZ STEER¿ THERMOCOOL® BI-DIRECTIONAL CATHETER. THE PATIENT SUFFERED CARDIAC TAMPONADE REQUIRING A PERICARDIOCENTESIS. A REPORT WAS BROUGHT TO BWI'S NOTICE: - DATE OF EVENT: (B)(6) 2023 - HOSPITAL NAME: (B)(6) HOSPITAL - BRAND NAME: BWI - PRODUCT NAME: THERMOCOOL F-J CURVE - PRODUCT CODE: 36Q5JM - LOT/BATCH/EXP: 30993121M - THE PRODUCT WAS NOT BEING USED IN A CLINICAL TRIAL. - THE EVENT HAPPENED DURING A PROCEDURE. - THEY WERE NOT IN THE PROCEDURE AT THE TIME OF THE EVENT. - SURGERY WAS PROLONGED DUE TO THE EVENT. IT WAS REPORTED THAT DURING AN ATRIAL FLUTTER ABLATION AND A STANDARD CTI (CAVOTRICUSPID ISHMUS) LINE THAT THERE WAS NOISE SEEN ON THE PROXIMAL ELECTRODES AND IMPEDANCE WAS DEEMED UNRELIABLE DURING ABLATION. THERE WAS A STEAM POP AND UNFORTUNATELY THE PATIENT EXPERIENCED A TAMPONADE, THAT REQUIRED DRAINING AND SURGICAL INTERVENTION. CUSTOMER PROVIDED OPERATOR¿S PROCEDURE NOTES BELOW: ELECTIVE ADMISSION FOR AF/AFL ABLATION ECGS WITH TYPICAL FLUTTER WAS SEEN BEFORE CAME IN AF. LOCAL ANESTHETIC AND SEDATION US GUIDED RFV ACCESS. 1X7 F SHEATH FOR CS (CORONARY SINUS). 1X8 F SHEATH THEN UPGRADED
Patient problems: Cardiac Tamponade
MAUDE report 17421830 ↗InjuryMar 24, 2016No Display/Image
DURING A CARDIAC ABLATION, PHYSICIAN WAS UNABLE TO SEE ELECTRICAL ACTIVITY ON EP MED SYSTEM. CABLES WERE REPLACED, STAFF CHECKED ALL THE CONNECTIONS AND EVENTUALLY CHANGED THE ABLATION CATHETER WITH ONLY SLIGHT IMPROVEMENT. ULTIMATELY, THEY HAD TO SWITCH ABLATION GENERATORS.
Patient problems: No Information
MAUDE report 5532313 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030031. One submission often covers many models, so check the product and lot details in each record.
- Z-0169-2023 · Sep 7, 2022 · Open, ClassifiedTHERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in con
- Z-0139-2015 · Oct 19, 2014 · TerminatedThermoCool SmartTouch Navigation Catheters, Catalog No. D132701, D132702, D132703, D132704, D132705, D133601, D133602, D133603.
- Z-0982-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-04-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0980-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-02-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0983-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-05-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0981-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-03-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0979-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-01-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0057-06 · Aug 8, 2005 · TerminatedThermocool Irrigated Catheters Celsius LABEL US Part Number# EU CATALOG # L-1189-01-S D17TCBLRT 35Q-13R PMA; L-1189-02-S D17TCDLRT 35Q-33R PMA; L-1189-03-S D17TCFLRT 35Q-53R P
Other catheter, percutaneous, cardiac ablation, for treatment of atrial flutter devices
- QDOT MICRO D139505Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter27 deaths915 reports
- IntellaNav MiFi™ Open-Irrigated M004PMR96200Boston Scientific Corporation · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter4 deaths269 reports
- QDOT MICRO D139504Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter7 deaths138 reports
- Sphere-9™ Catheter AFR-00001Medtronic, Inc. · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter1 recall5 deaths77 reports
- QDOT MICRO D139401Biosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter2 deaths114 reports
- NaviStar ThermoCool NI75TCDHBiosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter111 reports
- NaviStar RMT ThermoCool NR7TCSIYBiosense Webster Inc · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter9 deaths103 reports
- IntellaNav MiFi™ Open-Irrigated M004PMR9620K20Boston Scientific Corporation · catheter, percutaneous, cardiac ablation, for treatment of atrial flutter78 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page