EZ Steer NAV BN7TCJJ4LRecalls & reported problems
Biosense Webster Inc · Class 3
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10846835003055), received between Nov 9, 2020 and Aug 14, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (P990025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Patient Device Interaction Problem1
Patient problems
- Complete Heart Block1
- Heart Block1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 14, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN AVNRT (ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA) ABLATION WITH A EZ STEER¿ NAV BI-DIRECTIONAL ELECTROPHYSIOLOGY CATHETER AND THE PATIENT EXPERIENCED COMPLETE HEART BLOCK. IT WAS REPORTED THAT THE PHYSICIAN ACCIDENTALLY BURNED THE AV (ATRIOVENTRICULAR) NODE AND THE PATIENT LOST CONDUCTION AND WAS IN COMPLETE HEART BLOCK. THE NGEN FOOT PEDAL WAS USED TO APPLY ABLATION DURING THE PROCEDURE AND THAT THE PHYSICIAN ACCIDENTALLY HIT THE NGEN FOOT PEDAL AFTER THE PATIENT WAS IN HEART BLOCK. THE LOSS OF CONDUCTION WAS NOTICED AND CONFIRMED ON THE INTRACARDIAC SIGNALS DISPLAYED ON THE CARTO 3 SYSTEM. THE PROCEDURE WAS ABORTED. NO MEDICAL INTERVENTION HAD BEEN PERFORMED. THE PHYSICIAN HAD SET UP FOR A PACEMAKER AND THAT THE PATIENT WAS ADVISED THAT THE PATIENT WOULD RECEIVE A PACEMAKER HOWEVER IT WAS REPORTED THAT THE CONDUCTION WAS SLOWLY RECOVERING AND COMING BACK BUT NOT FULLY. THE PATIENT WAS MOVED TO THE POST-OP AREA TO BE MONITORED AND THAT IT WOULD BE DETERMINED AT A LATER TIME IF THE PATIENT WOULD NEED A PACEMAKER. HOWEVER IT WAS REPORTED THAT THE PATIENT HAD FULLY RECOVERED AND DID NOT NEED A DEVICE. THE PHYSICIAN'S OPINION ON THE CAUSE
Patient problems: Heart Block
MAUDE report 22806993 ↗InjuryNov 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A FEMALE PATIENT ((B)(6), 83KG) WITH 1ST DEGREE ATRIOVENTRICULAR (A/V) BLOCK PRIOR TO THE START OF THE PROCEDURE, UNDERWENT CARDIAC ABLATION PROCEDURE FOR ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) WITH EZ STEER¿ NAV BI-DIRECTIONAL ELECTROPHYSIOLOGY CATHETER AND SUFFERED A COMPLETE HEART BLOCK REQUIRING PACEMAKER IMPLANTATION. THE EVENT OCCURRED DURING USE OF BIOSENSE WEBSTER PRODUCTS. THE CATHETER POSITION MOVED POSTERIORLY WHILE ABLATION WAS BEING APPLIED. THE CATHETER WAS IN THIS POSITION FOR LESS THAN 2 SECONDS WHEN THE ENERGY APPLICATION WAS STOPPED. THE PHYSICIAN¿S OPINION IS THAT THE CAUSE OF THE EVENT IS PROCEDURE. MAX WATTAGE USED FOR THE ENTIRE PROCEDURE WAS 35 WATTS. A TEMPORARY PACING LEAD WAS PLACED, AND THE PATIENT WAS ADMITTED TO THE INTENSIVE CARE UNIT FOR OVERNIGHT OBSERVATION. THE PATIENT CONDITION IS UNCHANGED. EXTENDED HOSPITALIZATION WAS REQUIRED, THE PATIENT HAD A PERMANENT PACEMAKER IMPLANTED THE FOLLOWING DAY AND WAS DISCHARGED TO HOME.
Patient problems: Complete Heart Block; No Code Available
MAUDE report 10814595 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990025. One submission often covers many models, so check the product and lot details in each record.
- Z-0369-2010 · Oct 2, 2008 · TerminatedBiosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 8mm DS Catheters; Part Numbers: D-1259-01 & D-1259-02
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports