WEBSTER F5QA252RTRecalls & reported problems
Biosense Webster Inc · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10846835007213), received between Aug 13, 2021 and Sep 22, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Patient Device Interaction Problem2
- Break1
- Material Twisted/Bent1
Patient problems
- Cardiac Tamponade1
- Perforation of Vessels1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 22, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) ABLATION PROCEDURE WITH A WEBSTER¿ ELECTROPHYSIOLOGY CATHETER AND SUFFERED CARDIAC TAMPONADE (CT) REQUIRING PERICARDIOCENTESIS. IT WAS REPORTED THAT A PERICARDIAL EFFUSION WAS NOTED POST USE OF BIOSENSE WEBSTER PRODUCTS. THE CAUSE IS UNKNOWN; HOWEVER, THE PHYSICIAN'S OPINION ON THE CAUSE IS THAT IT IS PROCEDURE RELATED AND NOT CATHETER RELATED. A PERICARDIAL PUNCTURE WAS PERFORMED TO DRAIN BLOOD. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT OUTCOME WAS REPORTED AS ¿IMPROVED, ON THAT DAY, NO FURTHER INFORMATION AT THE MOMENT¿. THE PATIENT REQUIRED EXTENDED HOSPITALIZATION FOR MONITORING. TRANSSEPTAL PUNCTURE WAS PERFORMED WITH AN ABBOTT BRK NEEDLE. PRIOR TO NOTING THE CT, ABLATION WAS PERFORMED. THERE WAS NO EVIDENCE OF A STEAM POP. IT IS UNKNOWN AT WHAT PHASE THE EVENT OCCURRED; HOWEVER, IT WAS DETECTED POST PROCEDURE. AN IRRIGATED CATHETER WAS USED IN THE EVENT AND THE FLOW SETTINGS WERE: 8ML/MIN UP TO 30W, 15ML/MIN ABOVE 30W. THE CORRECT CATHETER SETTINGS WERE SELECTED ON THE GENERATOR. THE PUMP WAS SWITCHING FROM LOW TO HIGH FLOW DURING ABLATION. NO ERROR MESSAGES WERE OBSER
Patient problems: Cardiac Tamponade
MAUDE report 12513575 ↗InjuryAug 13, 2021Break; Material Twisted/Bent; Patient Device Interaction Problem; Break; Material Twisted/Bent; Patient Device Interaction Problem
IT WAS REPORTED THAT A (B)(6) MALE PATIENT UNDERWENT AN ELECTROPHYSIOLOGICAL EXAMINATION PROCEDURE WITH A WEBSTER CATHETER. THE CATHETER TIP BROKE AND BECAME KNOTTED AND ULTIMATELY HAD TO BE SURGICALLY REMOVED. THE WEBSTER CATHETER WAS INSERTED INTO THE VENA FEMORALIS FOR A DIAGNOSTIC ELECTROPHYSIOLOGY PROCEDURE. DURING THE PROCEDURE, THE CATHETER TIP BROKE AFTER HAVING VASCULAR WALL CONTACT. THE BROKEN POSITION WAS VISIBLE UNDER FLUOROSCOPY. THE PHYSICIAN TRIED TO RETRIEVE THE CATHETER, BUT DUE TO THE BROKEN TIP IT WAS NOT POSSIBLE TO REMOVE THE CATHETER FROM THE VEIN. TWISTING AND PULLING THE CATHETER RESULTED IN BECOMING KNOTTED AND ULTIMATELY HAVING TO BE SURGICALLY REMOVED. THE VENA FEMORALIS HAD TO BE OPENED COMPLETELY IN THE AREA OF THE GROIN., THIS ADVERSE EVENT WAS DISCOVERED DURING USE OF BIOSENSE WEBSTER, INC. (BWI) PRODUCTS. THE PHYSICIAN¿S OPINION ON THE CAUSE OF THIS ADVERSE EVENT WAS A BWI PRODUCT MALFUNCTION. THE PATIENT REQUIRED EXTENDED HOSPITALIZATION BECAUSE OF THE NECESSARY ADDITIONAL SURGICAL INTERVENTION TO REMOVE THE CATHETER FROM THE VEIN. HEALING TIME AND LENGTH OF STAY IN HOSPITAL WAS EXTENDED. A THERMOCOOL® SMARTTOUCH® SF CATHETER WAS NOT USED. NO ABLATI
Patient problems: Perforation of Vessels
MAUDE report 12320939 ↗
Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices
- Advisor™ HD Grid, Sensor Enabled™ D-AVHD-DF16St. Jude Medical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths358 reports
- IntellaMap Orion™ M004RC64S0Boston Scientific Corporation · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- ACHIEVE™ 990063-020Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING7 deaths129 reports
- Inquiry™ IBI-81104Irvine Biomedical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 recall2 deaths64 reports
- DECANAV R7F282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 deaths108 reports
- LASSO NAV D134301Biosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 deaths82 reports
- DECANAV R7D282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death75 reports
- ACHIEVE ADVANCE™ 2ACH20Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING3 deaths69 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.