NaviStar RMT NR7TCS4YRecalls & reported problems
Biosense Webster Inc · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10846835008463), received between Jun 14, 2019 and Oct 29, 2020 — classified as 2 injuries.
1 other recall cover devices cleared under the same FDA submission (P990025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Cardiac Tamponade2
- Cardiac Arrest1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INITIALLY THIS EVENT WAS ASSESSED AS A SERIOUS INJURY UNDER THE THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER. ADDITIONAL INFORMATION WAS RECEIVED ON 10/1/2020 STATING THAT THE 4MM NON-IRRIGATED CATHETER WAS BEING USED FOR MAPPING. ABLATION WAS PERFORMED WITH THIS CATHETER FOR A FEW SECONDS ONLY. SINCE IT WAS CONFIRMED THAT ABLATION WAS PERFORMED FOR A FEW SECONDS WITH THE 4 MM CATHETER, ITS RELATIONSHIP WITH THE ADVERSE EVENT OCCURRENCE CANNOT BE EXCLUDED; THEREFORE, THIS EVENT IS ALSO NOW ASSESSED AS A SERIOUS INJURY UNDER THE 4MM NON-IRRIGATED CATHETER WITH THE AWARENESS DATE OF (B)(6) 2020. IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT AN ATRIOVENTRICULAR REENTRANT TACHYCARDIA/WOLFF-PARKINSON-WHITE SYNDROME (AVRT/WPW) ABLATION PROCEDURE WITH A THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER AND A NAVI-STAR¿ RMT ELECTROPHYSIOLOGY CATHETER AND SUFFERED CARDIAC TAMPONADE (REQUIRING PERICARDIOCENTESIS) AND CARDIAC ARREST (REQUIRING CARDIOPULMONARY RESUSCITATION (CPR)). DURING THE PROCEDURE, A 4MM NON-IRRIGATED CATHETER WAS BEING USED FOR MAPPING THE LEFT ATRIUM. ABLATION WAS DONE WITH THIS CATHETER FOR A FEW SECONDS ONLY; HOWEVER, THE PHYSICIAN NOT
Patient problems: Cardiac Arrest; Cardiac Tamponade
MAUDE report 10753301 ↗InjuryJun 14, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A MALE PATIENT UNDERWENT AN ATRIOVENTRICULAR NODAL REENTRY TACHYCARDIA (AVNRT) ABLATION PROCEDURE WITH A NAVI-STAR¿ RMT ELECTROPHYSIOLOGY CATHETER AND SUFFERED CARDIAC TAMPONADE REQUIRING NO MEDICAL OR SURGICAL INTERVENTION. DURING THE ABLATION PHASE, THE PATIENT SUDDENLY BECAME HYPOTENSIVE. CARDIAC TAMPONADE WAS CONFIRMED BY ECHOCARDIOGRAM. NO MEDICAL OR SURGICAL INTERVENTION WAS PERFORMED. THE PATIENT WAS REPORTED IN STABLE CONDITION AND REQUIRED FURTHER MONITORING. EXTENDED HOSPITALIZATION WAS NOT REQUIRED. THERE¿S NO INFORMATION REGARDING PHYSICIAN CAUSALITY OPINION. NO BIOSENSE WEBSTER INC.(BWI) PRODUCT MALFUNCTIONS WERE REPORTED. TRANSSEPTAL PUNCTURE WAS NOT PERFORMED. AN UNKNOWN 8 FR SHORT SHEATH WAS USED DURING THE CASE. THE OVERALL TIME FOR ABLATION AT THE SITE OF THE INJURY WAS AROUND 2 MINUTES IN THE SLOW PATHWAY OF THE AVNRT. IT IS NOT CLEAR IF THE TAMPONADE HAPPENED DURING THE ABLATION OR DURING THE CATHETER MANIPULATION IN THE RIGHT ATRIUM. AN UNKNOWN GENERATOR WAS USED IN TEMPERATURE CONTROL MODE WITH POWER AT 30 WATTS AND A TEMPERATURE CUT-OFF AT 50 DEGREES. THE POWER DID NOT TITRATE DURING THE ABLATION. NO IMPEDANCE OR TEMPERATURE INCREASES WER
Patient problems: Cardiac Tamponade
MAUDE report 8701928 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990025. One submission often covers many models, so check the product and lot details in each record.
- Z-0369-2010 · Oct 2, 2008 · TerminatedBiosense Webster NaviStar RMT Diagnostic/Ablation Steerable Tip 8mm DS Catheters; Part Numbers: D-1259-01 & D-1259-02
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.