ThermoCool SF NAV D131501Recalls & reported problems
Biosense Webster Inc · Class 3
All models of ThermoCool SF NAV (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10846835009231), received between Feb 21, 2017 and Nov 27, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
8 other recalls cover devices cleared under the same FDA submission (P030031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Patient Device Interaction Problem1
Patient problems
- Cardiac Tamponade1
- Heart Failure/Congestive Heart Failure1
- Pericardial Effusion1
- Pericarditis1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 27, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
AN ARTICLE WAS RECEIVED ENTITLED "50W VS 40W DURING HIGH-POWER SHORT-DURATION ABLATION FOR PAROXYSMAL ATRIAL FIBRILLATION" THIS COMPLAINT IS FROM A LITERATURE SOURCE. THE FOLLOWING LITERATURE CITE HAS BEEN REVIEWED: COSTEA A, DIAZ JC, OSORIO J, MATOS CD, HOYOS C, GOYAL S, TE C, D'SOUZA B, RASTOGI M, LOPEZ-CABANILLAS N, IBANEZ LC, THORNE C, VARLEY AL, ZEI PC, SAUER WH, ROMERO JE. 50W VS 40W DURING HIGH-POWER SHORT-DURATION ABLATION FOR PAROXYSMAL ATRIAL FIBRILLATION: A MULTICENTER PROSPECTIVE STUDY. JACC CLIN ELECTROPHYSIOL. 2023 SEP 5:S2405-500X(23)00618-7. DOI: 10.1016/J.JACEP.2023.08.005. EPUB AHEAD OF PRINT. PMID: 37804258. OBJECTIVE/METHODS/STUDY DATA: BACKGROUND HIGH-POWER SHORT-DURATION (HPSD) RADIOFREQUENCY ABLATION OF ATRIAL FIBRILLATION (AF) INCREASES FIRST-PASS PULMONARY VEIN ISOLATION (PVI) AND FREEDOM FROM ATRIAL ARRHYTHMIAS WHILE DECREASING PROCEDURAL TIME. HOWEVER, THE OPTIMAL POWER SETTING IN TERMS OF SAFETY AND EFFICACY HAS NOT BEEN DETERMINED. OBJECTIVES THIS STUDY COMPARED THE PROCEDURAL CHARACTERISTICS AND CLINICAL OUTCOMES OF 50W VS 40W DURING HPSD ABLATION OF PAROXYSMAL AF. METHODS PATIENTS FROM THE REAL-AF PROSPECTIVE MULTICENTER REGISTRY (REAL-WORLD EXPERIENC
Patient problems: Heart Failure/Congestive Heart Failure; Pericardial Effusion; Pericarditis
MAUDE report 18211843 ↗InjuryFeb 21, 2017Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A (B)(6) MALE PATIENT UNDERWENT A BIPOLAR ABLATION PROCEDURE (WITH 2 CATHETERS) FOR VENTRICULAR TACHYCARDIA WITH A THERMOCOOL SF NAV CATHETER AND A THERMOCOOL SMARTTOUCH UNI-DIRECTIONAL SF CATHETER AND SUFFERED A CARDIAC TAMPONADE REQUIRING PERICARDIOCENTESIS. TWO HOURS POST-PROCEDURE, A TAMPONADE WAS DETECTED. PERICARDIOCENTESIS YIELDED AN UNKNOWN AMOUNT OF FLUID. PATIENT DID NOT REQUIRE EXTENDED HOSPITALIZATION AS A RESULT OF THIS ADVERSE EVENT. PATIENT FULLY RECOVERED. THERE WERE NO FACTORS CITED THAT MAY HAVE CONTRIBUTED TO THE ADVERSE EVENT. PHYSICIAN¿S OPINION REGARDING THE CAUSE OF THE ADVERSE EVENT IS THAT IT WAS PROCEDURE-RELATED. TRANSSEPTAL PUNCTURE WAS PERFORMED WITH AN UNSPECIFIED NEEDLE AND A ST. JUDE MEDICAL-ABBOTT SL0 SHEATH. SHEATH IN USE WAS A ST. JUDE MEDICAL-ABBOTT AGILIS L CURVE. GENERATOR PARAMETERS INCLUDED POWER CONTROL MODE WITH POWER UP TO 30 WATTS (NOT TITRATED) AND IMPEDANCE UP TO 150-160 OHMS. OVERALL ABLATION TIME WAS 7.5 MINUTES. LAST ABLATION CYCLE TIME WAS 10 SECONDS. IRRIGATED CATHETER FLOW WAS SET AT 15 ML/MIN FOR BOTH CATHETERS. PATIENT RECEIVED ANTICOAGULANT (HEPARIN) DURING THE PROCEDURE WITH ACTIVATED CLOTTING TIME MAINTAI
Patient problems: Cardiac Tamponade
MAUDE report 6347540 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030031. One submission often covers many models, so check the product and lot details in each record.
- Z-0169-2023 · Sep 7, 2022 · Open, ClassifiedTHERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in con
- Z-0139-2015 · Oct 19, 2014 · TerminatedThermoCool SmartTouch Navigation Catheters, Catalog No. D132701, D132702, D132703, D132704, D132705, D133601, D133602, D133603.
- Z-0981-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-03-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0983-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-05-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0980-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-02-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0982-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-04-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0979-2012 · Nov 22, 2010 · TerminatedEZ Steer ThermoCool Nav Part Number D-1292-01-S Diagnostic/Ablation Catheter instructions for use. The Biosense EZ STEER~ THERMOCOOL~ NAV Diagnostic/Ablation Deflectable Tip Cat
- Z-0057-06 · Aug 8, 2005 · TerminatedThermocool Irrigated Catheters Celsius LABEL US Part Number# EU CATALOG # L-1189-01-S D17TCBLRT 35Q-13R PMA; L-1189-02-S D17TCDLRT 35Q-33R PMA; L-1189-03-S D17TCFLRT 35Q-53R P
Other Cardiac ablation percutaneous catheter devices
- THERMOCOOL SMARTTOUCH SF D134805Biosense Webster Inc · Cardiac ablation percutaneous catheter1 recall176 deaths4,106 reports
- THERMOCOOL SMARTTOUCH SF D134804Biosense Webster Inc · Cardiac ablation percutaneous catheter44 deaths784 reports
- THERMOCOOL SMARTTOUCH SF D134801Biosense Webster Inc · Cardiac ablation percutaneous catheter15 deaths608 reports
- ThermoCool SmartTouch D132705Biosense Webster Inc · Cardiac ablation percutaneous catheter25 deaths574 reports
- THERMOCOOL SMARTTOUCH SF D134702Biosense Webster Inc · Cardiac ablation percutaneous catheter19 deaths366 reports
- THERMOCOOL SMARTTOUCH SF D134701Biosense Webster Inc · Cardiac ablation percutaneous catheter11 deaths278 reports
- ThermoCool SmartTouch D133602Biosense Webster Inc · Cardiac ablation percutaneous catheter8 deaths225 reports
- SMARTABLATE SAT001Biosense Webster Inc · Cardiac ablation percutaneous catheter185 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page