Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10846835009668), received between Oct 9, 2020 and Dec 16, 2021 — classified as 3 injuries.
The most frequently reported specific device problem is “Entrapment of Device” (1 report).
1 other recall cover devices cleared under the same FDA submission (K113213); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Entrapment of Device1
- Patient Device Interaction Problem1
Patient problems
- Cardiac Tamponade2
- Foreign Body In Patient1
- Unspecified Heart Problem1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 16, 2021Entrapment of Device; Patient Device Interaction Problem; Entrapment of Device; Patient Device Interaction Problem
IT WAS REPORTED THAT A (B)(6)-YEAR-OLD MALE (B)(6) PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) ABLATION PROCEDURE WITH A LASSO® NAV ECO CATHETER AND THE LASSO CATHETER GOT CAUGHT IN THE MITRAL VALVE WHICH REQUIRED SURGICAL INTERVENTION. AFTER THE CATHETER WAS INSERTED INTO THE PATIENT¿S BODY, DURING MAPPING WAS CONDUCTED AT THE LEFT ATRIUM (LA), THE LASSO GOT CAUGHT IN THE MITRAL VALVE (MV) AND COULD NOT BE REMOVED. THORACOTOMY WAS PERFORMED WITHOUT OMISSION. AN EMERGENCY SURGERY WAS PERFORMED TO OPEN THE CHEST. THE PHYSICIAN COMMENTED THAT ALTHOUGH IT WAS NOT AN ARTIFICIAL VALVE, THE TIP SHAPE OF THE LASSO SLID TOWARD THE LEFT VENTRICLE DURING MAPPING OF THE LA FLOOR, WHERE IT WAS CAUGHT BY A STRUCTURE. THE OPEN-CHEST SURGERY ENDED UNEVENTFULLY. THE OPERATION DID NOT REQUIRE VALVE REPLACEMENT, AND THE PATIENT UNDERWENT AFIB TREATMENT AND LEFT ATRIAL APPENDAGE (LAA) CLOSURE WITH THE MAZE OPERATION. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION AND HAS RETURNED TO THE GENERAL WARD. THE CHEST OPENING OPERATION WAS COMPLETED SUCCESSFULLY. THE PHYSICIAN¿S OPINION REGARDING THE CAUSE OF THIS ADVERSE EVENT IS THAT THIS IS PROCEDURE RELATED. THE PATIENT OUTCOME OF THE ADVERSE EVE
Patient problems: Foreign Body In Patient; Unspecified Heart Problem
MAUDE report 13015885 ↗InjuryOct 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED CARDIAC ABLATION PROCEDURE ON WHICH A THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER AND TWO LASSO® NAV ECO CATHETERS WERE USED FOR MAPPING, AND SUFFERED CARDIAC TAMPONADE REQUIRING PERICARDIOCENTESIS. DURING THE PROCEDURE, TWO LASSO® NAV CATHETERS AND A SOUNDSTAR ECO 8FG ULTRASOUND CATHETER WERE PLACED IN THE RIGHT ATRIUM (RA) FOR SEPTAL PUNCTURE. AFTERWARDS, THE THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER AND LASSO CATHETERS WERE PLACED IN THE LEFT ATRIUM (LA). LA GEOMETRY WAS CREATED WITH THE THREE CATHETERS. THE PHYSICIAN FELT ABNORMALITY WHILE USING THE CATHETERS. AFTER CREATING THE GEOMETRY, PERICARDIAL EFFUSION WAS CONFIRMED WITH THE SOUNDSTAR PLACED IN THE RA. REMAINDER OF THE PROCEDURE WAS ABORTED, AND PERICARDIOCENTESIS WAS PERFORMED TO DRAIN AN UNSPECIFIED AMOUNT OF FLUID FROM THE PERICARDIAL SPACE. EFFUSION DID NOT CONTINUE TO ACCUMULATE AFTER PERICARDIAL DRAINAGE. IT IS UNKNOWN IF EXTENDED HOSPITALIZATION WAS REQUIRED. PATIENT¿S OUTCOME IS UNKNOWN. THE PHYSICIAN COMMENTED IT IS UNKNOWN WHEN THE EVENT HAPPENED AND IF THERE WAS RELATIONSHIP WITH THE PRODUCTS. HE SAID HE WAS A LIT
Patient problems: Cardiac Tamponade
MAUDE report 10659399 ↗InjuryOct 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED CARDIAC ABLATION PROCEDURE ON WHICH A THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER AND TWO LASSO® NAV ECO CATHETERS WERE USED FOR MAPPING, AND SUFFERED CARDIAC TAMPONADE REQUIRING PERICARDIOCENTESIS. DURING THE PROCEDURE, TWO LASSO® NAV CATHETERS AND A SOUNDSTAR ECO 8FG ULTRASOUND CATHETER WERE PLACED IN THE RIGHT ATRIUM (RA) FOR SEPTAL PUNCTURE. AFTERWARDS, THE THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER AND LASSO CATHETERS WERE PLACED IN THE LEFT ATRIUM (LA). LA GEOMETRY WAS CREATED WITH THE THREE CATHETERS. THE PHYSICIAN FELT ABNORMALITY WHILE USING THE CATHETERS. AFTER CREATING THE GEOMETRY, PERICARDIAL EFFUSION WAS CONFIRMED WITH THE SOUNDSTAR PLACED IN THE RA. REMAINDER OF THE PROCEDURE WAS ABORTED, AND PERICARDIOCENTESIS WAS PERFORMED TO DRAIN AN UNSPECIFIED AMOUNT OF FLUID FROM THE PERICARDIAL SPACE. EFFUSION DID NOT CONTINUE TO ACCUMULATE AFTER PERICARDIAL DRAINAGE. IT IS UNKNOWN IF EXTENDED HOSPITALIZATION WAS REQUIRED. PATIENT¿S OUTCOME IS UNKNOWN. THE PHYSICIAN COMMENTED IT IS UNKNOWN WHEN THE EVENT HAPPENED AND IF THERE WAS RELATIONSHIP WITH THE PRODUCTS. HE SAID HE WAS A LIT
Patient problems: Cardiac Tamponade
MAUDE report 10659395 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113213. One submission often covers many models, so check the product and lot details in each record.
- Z-1617-2013 · Jun 19, 2013 · Terminated20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Designed to facilitate electrophysiological mapping of the atria of the heart.
Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices
- Advisor™ HD Grid, Sensor Enabled™ D-AVHD-DF16St. Jude Medical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths358 reports
- IntellaMap Orion™ M004RC64S0Boston Scientific Corporation · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- ACHIEVE™ 990063-020Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING7 deaths129 reports
- Inquiry™ IBI-81104Irvine Biomedical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 recall2 deaths64 reports
- DECANAV R7F282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 deaths108 reports
- LASSO NAV D134301Biosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 deaths82 reports
- DECANAV R7D282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death75 reports
- ACHIEVE ADVANCE™ 2ACH20Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING3 deaths69 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.