Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10846835009682), received between Dec 21, 2021 and Dec 21, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K113213); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Patient Device Interaction Problem1
Patient problems
- Cardiac Tamponade1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 21, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT AN (B)(6) FEMALE ((B)(6)) PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) ABLATION PROCEDURE WITH A THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER AND A LASSO® NAV ECO CATHETER AND SUFFERED CARDIAC TAMPONADE REQUIRING PERICARDIOCENTESIS. IT WAS REPORTED THAT THE PHYSICIAN PERFORMED A SEPTAL PUNCTURE WITH AN ACUNAV AND PERFORMED PULMONARY VEIN ISOLATION (PVI) WITH A LASSO CATHETER IN THE PULMONARY VEIN (PV), AND ABOUT 2 HOURS AFTER THE PROCEDURE WAS STARTED, THE EVENT OCCURRED. CAVO-TRICUSPID ISTHMUS (CTI) ABLATION WAS PERFORMED AFTER PVI WAS COMPLETED. BECAUSE THE BLOCK OF CTI WAS DIFFICULT, SL1 SHEATH WAS REPLACED WITH VIZIGO, AND ADDITIONAL ABLATION WAS PERFORMED. AFTER COMPLETION OF THE CTI BLOCK, BLOOD PRESSURE WAS CONFIRMED TO HAVE DECREASED WHILE CHECKING PVI. BECAUSE THE MOVEMENT OF THE CARDIAC SHADOW WAS POOR, ECHOCARDIOGRAPHY WAS PERFORMED, AND EFFUSION WAS CONFIRMED, SO PERICARDIOCENTESIS WAS PERFORMED FOR DRAINAGE. THE SYMPTOMS IMPROVED WITH DRAINAGE. THE PHYSICIAN COMMENTED THAT AS ARTERIAL BLOOD WAS DRAWN BY DRAINAGE, IT WAS THOUGHT TO HAVE OCCURRED DURING PVI, BUT IT WAS NOT CLEAR EXACTLY WHERE AND HOW THE TAMPONADE OCCURRED. ADDI
Patient problems: Cardiac Tamponade
MAUDE report 13050540 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113213. One submission often covers many models, so check the product and lot details in each record.
- Z-1617-2013 · Jun 19, 2013 · Terminated20-Pole LASSO Nav Catheter Eco, Catalog # D-1349-02S. Designed to facilitate electrophysiological mapping of the atria of the heart.
Other CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING devices
- Advisor™ HD Grid, Sensor Enabled™ D-AVHD-DF16St. Jude Medical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths358 reports
- IntellaMap Orion™ M004RC64S0Boston Scientific Corporation · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING6 deaths356 reports
- ACHIEVE™ 990063-020Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING7 deaths129 reports
- Inquiry™ IBI-81104Irvine Biomedical, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 recall2 deaths64 reports
- DECANAV R7F282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING4 deaths108 reports
- LASSO NAV D134301Biosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING2 deaths82 reports
- DECANAV R7D282CTBiosense Webster Inc · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING1 death75 reports
- ACHIEVE ADVANCE™ 2ACH20Medtronic, Inc. · CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING3 deaths69 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.