Skip to content

Sterile Interface Cable D133701Recalls & reported problems

Biosense Webster Inc · Cath to Generator CBL Sterile · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10846835010121), received between May 29, 2026 and May 29, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (P240006); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem1
  • Electrical Shorting1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2026Communication or Transmission Problem; Electrical Shorting

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A STERILE INTERFACE CABLE AND A SIGNAL INTERFERENCE (NOISE/LOSS) ISSUE OBSERVED ON THE IC, BS, OR ALL ECG (BS + IC) CHANNELS ON ALL SYSTEMS AND THE EKG BECAME UNAVAILABLE FOR THE PHYSICIAN TO MONITOR THE PATIENT¿S HEART RHYTHM. INITIALLY IT WAS REPORTED THAT THERE WAS A "CURRENT LEAK" ALERT BEING DISPLAYED ON THE TRUPULSE¿ MONITOR. THEY REPLACED THE VARIPULSE¿ CATHETER CABLE (STERILE INTERFACE CABLE) AND THE ISSUE WAS RESOLVED. THE PROCEDURE CONTINUED. ADDITIONAL INFORMATION WAS RECEIVED ON 20-MAY-2026. A SIGNAL INTERFERENCE (NOISE/LOSS) ISSUE WAS OBSERVED ON THE INTRACARDIAC (IC), BODY SURFACE (BS), OR ALL ECG (BS + IC) CHANNELS ON ALL SYSTEMS. THE EKG BECAME UNAVAILABLE FOR THE PHYSICIAN TO MONITOR THE PATIENT¿S HEART RHYTHM. THE CATHETER WAS IN THE PATIENT¿S BODY. THE "CURRENT LEAK" ALERT EVENT WAS ORIGINALLY CONSIDERED NON-REPORTABLE, HOWEVER, BWI BECAME AWARE ON 20-MAY-2026 OF A SIGNAL INTERFERENCE (NOISE/LOSS) ISSUE OBSERVED ON THE IC, BS, OR ALL ECG (BS + IC) CHANNELS ON ALL SYSTEMS AND THE EKG BECAME UNAVAILABLE FOR THE PHYSICIAN TO MONITOR THE PATIENT¿S HEART RHYTHM AND HAVE ASSESSED THIS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25337203 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P240006. One submission often covers many models, so check the product and lot details in each record.

  • Z-1205-2025 · Jan 5, 2025 · Completed
    VARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use in catheter based cardiac electrophysiological mapping (stimulating and recordin

Other Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation devices

All Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation devices →

Sources and updates