LifeBand® Load-distributing Band 100 (REF 8700-0706-01)Recalls & reported problems
Zoll Circulation, Inc. · Single-use chest compression band. (3 per package) · Class 2
All models of LifeBand Load-distributing Band (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10849111001691), received between Oct 11, 2018 and Nov 20, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem5
- Material Split, Cut or Torn5
- Use of Device Problem5
- Break1
- Connection Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information3
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Connection Problem; Connection Problem
THE CUSTOMER REPORTED THAT ONE OF THE AUTOPULSE LIFEBANDS (LOT # UNKNOWN) FROM THE 3-PACK WILL NOT CLIP INTO THE AUTOPULSE PLATFORM. NO FURTHER INFORMATION WAS PROVIDED. IT IS UNKNOWN WHEN THE PROBLEM OCCURRED. HOWEVER, PATIENT USE INFORMATION WAS REQUESTED, BUT NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Insufficient Information
MAUDE report 23610370 ↗MalfunctionApr 18, 2023Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
THE CUSTOMER REPORTED THAT DURING THE STRENUOUS EFFORT TO POSITION THE PATIENT ON THE X-RAY TABLE, THE LIFEBAND (LOT # UNKNOWN) BROKE APART. THE CUSTOMER SENT PICTURES SHOWING THAT ONE OF THE HINGED GUARDS OF THE LIFEBAND COVER PLATE WAS BROKEN. THE TYVEK LINER WAS RIPPED, AND THE BAND WAS JAMMED AND TWISTED. WHILE INSTALLING THE SECOND LIFEBAND (LOT # UNKNOWN), THE BAND LOCKING CLIP BROKE. THE CUSTOMER CONTINUED THE RESUSCITATION ATTEMPT WITH MANUAL CPR. NO CONSEQUENCES OR IMPACTS ON THE PATIENT. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: MFR # 3010617000-2023-00397 FOR THE FIRST LIFEBAND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16766092 ↗MalfunctionApr 18, 2023Break; Material Integrity Problem; Material Split, Cut or Torn; Break; Material Integrity Problem; Material Split, Cut or Torn
THE CUSTOMER REPORTED THAT DURING THE STRENUOUS EFFORT TO POSITION THE PATIENT ON THE X-RAY TABLE, THE LIFEBAND (LOT # UNKNOWN) BROKE APART. THE CUSTOMER SENT PICTURES SHOWING THAT ONE OF THE HINGED GUARDS OF THE LIFEBAND COVER PLATE WAS BROKEN. THE TYVEK LINER WAS RIPPED, AND THE BAND WAS JAMMED AND TWISTED. WHILE INSTALLING THE SECOND LIFEBAND (LOT # UNKNOWN), THE BAND LOCKING CLIP BROKE. THE CUSTOMER CONTINUED THE RESUSCITATION ATTEMPT WITH MANUAL CPR. NO CONSEQUENCES OR IMPACTS ON THE PATIENT. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: MFR # 3010617000-2023-00411 FOR THE SECOND LIFEBAND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16766089 ↗MalfunctionJun 29, 2022Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTED THAT THE LIFEBAND (LOT #UNKNOWN) WILL NOT CLIP ON THE AUTOPULSE PLATFORM (SN (B)(4) AND CLAIMED TO HAVE BROKEN OFF THE LIFEBAND'S BELT CLIP WHEN TRYING TO INSTALL ON THE AUTOPULSE PLATFORM. IT IS UNKNOWN WHEN THE PROBLEM OCCURRED. HOWEVER, PATIENT USE INFORMATION WAS REQUESTED, BUT NO ADDITIONAL INFORMATION WAS PROVIDED. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: MFR (B)(4) FOR THE AUTOPULSE PLATFORM (SN (B)(4).
Patient problems: Insufficient Information
MAUDE report 14857294 ↗MalfunctionOct 21, 2021Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTED THAT THE AUTOPULSE LIFEBAND (LOT #158870) DOES NOT CLIP TO THE AUTOPULSE PLATFORM. OUT OF THE THREE LIFEBAND RECEIVED, TWO ARE FUNCTIONAL. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12670738 ↗MalfunctionAug 19, 2021Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
THE LIFEBAND (LOT # 158335) WAS USED TO RESUSCITATE A PATIENT IN CARDIAC ARREST. BEFORE INSTALLING THE LIFEBAND TO THE AUTOPULSE PLATFORM, THE USER CHECKED THE PROPER ALIGNMENT OF THE LIFEBAND, AND NO TWIST WAS OBSERVED. AFTER AUTOPULSE PLATFORM WAS POWERED ON, NO COMPRESSIONS WERE PERFORMED BECAUSE THE LIFEBAND GOT TWISTED. PER THE CUSTOMER, THE AUTOPULSE PLATFORM DID NOT DISPLAY ANY ERROR MESSAGES. THE USER RE-STARTED THE AUTOPULSE 3 OR 4 TIMES AND CONTINUED TO USE THE AUTOPULSE, BUT THE ISSUE PERSISTED. IN ADDITION, THE USER STATED THAT SOMETIMES THE LIFEBAND RAGGED. NO DEATH OR INJURY WAS REPORTED. NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12346371 ↗MalfunctionJun 12, 2021Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
DURING DEVICE CHECK, THE LIFEBAND (LOT # UNKNOWN) SHEARED AT THE PLACE WHERE THE LIFEBAND ENTERS THE PLATFORM. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11988567 ↗MalfunctionDec 31, 2020Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
THE AUTOPULSE PLATFORM SYSTEM WAS USED TO RESUSCITATE A (B)(6) MALE PATIENT IN CARDIAC ARREST. THE CAUSE OF CARDIAC ARREST IS UNKNOWN, AND IT WAS WITNESSED BY THE PATIENT'S SPOUSE. THE PATIENT WAS IN CARDIAC ARREST FOR ABOUT 4 MINUTES BEFORE RECEIVING MANUAL CPR, WHICH WAS PERFORMED BY THE POLICE, AND SUBSEQUENTLY, BY THE FIRE DEPARTMENT FOR APPROXIMATELY 16 MINUTES PRIOR TO THE USE OF THE AUTOPULSE PLATFORM. AFTER THE AUTOMATED CPR WAS PERFORMED FOR ABOUT 29 MINUTES, EN ROUTE TO THE HOSPITAL, THE LIFEBAND (LOT # UNKNOWN) WAS TORN. AFTER THE TEARING OF THE LIFEBAND, THE EMS CREW STOPPED USING THE AUTOPULSE AND REVERTED TO MANUAL CPR, WHICH WAS PERFORMED FOR ABOUT 9 MINUTES. RETURN OF SPONTANEOUS CIRCULATION (ROSC) WAS NOT ACHIEVED, AND THE PATIENT WAS PRONOUNCED DEAD AT THE HOSPITAL. AS PER THE CUSTOMER, THE PATIENT'S DEATH WAS NOT RELATED TO THE AUTOPULSE SYSTEM. AFTER THE CALL, WHILE THE CUSTOMER WAS TRYING TO REMOVE THE LIFEBAND FROM THE PLATFORM, IT WAS NOTED THAT THE LIFEBAND WAS TORN AT THE PLATFORM'S DRIVESHAFT, NEXT TO THE LIFEBAND CLIP. WHEN ATTEMPTING TO REMOVE THE LIFEBAND CLIP, THE CLOTH PORTION OF THE LIFEBAND BECAME COMPLETELY DETACHED FROM ITS CLIP, AND THE LIFEBAND
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11101231 ↗MalfunctionDec 29, 2020Use of Device Problem; Use of Device Problem
PER A REPORTER, THE LIFEBAND 3 PACK HAD AN UNKNOWN ISSUE DURING THE CALL. THE ISSUE WAS NOTED WITHIN A FEW MINUTES AFTER THE PLATFORM STARTED COMPRESSIONS. ANOTHER PLATFORM WAS UTILIZED, HOWEVER, A SIMILAR ISSUE OCCURRED TWO MORE TIMES. NO ADDITIONAL INFORMATION WAS PROVIDED. THE PATIENT'S STATUS INFORMATION WAS REQUESTED BUT THE CUSTOMER DID NOT PROVIDE A RESPONSE, THEREFORE THE PATIENT'S STATUS IS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 11086054 ↗MalfunctionAug 21, 2020Use of Device Problem; Use of Device Problem
DURING USE ON A FEMALE PATIENT WHO WAS APPROXIMATELY (B)(6) LBS. WITH LARGE BREAST/CHEST AREA, CUSTOMER REPORTED THAT THE LIFEBAND FAILED AFTER 18 MINUTES ON THE AUTOPULSE PLATFORM. PER REPORTER, THE AUTOPULSE PLATFORM WAS USED ON A FLAT TILE FLOOR WITH NO COMPLICATIONS. PATIENT'S STATUS INFORMATION WAS REQUESTED BUT THE CUSTOMER DID NOT PROVIDE A RESPONSE, THEREFORE PATIENT'S STATUS IS UNKNOWN. THE REPORT FOR AUTOPULSE PLATFORM IS SUBMITTED UNDER MFR 3010617000-2020-00878.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10440222 ↗MalfunctionOct 11, 2018Use of Device Problem; Use of Device Problem
AS REPORTED, AFTER THE PATIENT WAS RESUSCITATED BY THE AUTOPULSE, THE LIFEBAND'S YELLOW ALIGNMENT TAB BROKE OFF AND THE LIFEBAND IMMEDIATELY CAME OFF FROM THE AUTOPULSE PLATFORM. THE ISSUE OCCURRED UPON ARRIVAL AT THE HOSPITAL. NO PATIENT CONSEQUENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7958569 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.
- Z-0118-2023 · Sep 22, 2022 · Open, ClassifiedAutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR
Other Compressor, Cardiac, External devices
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External3,208 reports
- AutoPulse Li-ion Battery™ 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,151 reports
- AutoPulse® Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,136 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External317 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External276 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External250 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External241 reports
- LUCAS 99576Jolife AB · Compressor, cardiac, external2 recalls9 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page