LifeBand® Load-distributing Band 100 (REF 8700-0706-14)Recalls & reported problems
Zoll Circulation, Inc. · Single-use chest compression band. (3 per package) · Class 2
All models of LifeBand Load-distributing Band (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10849111002247), received between May 6, 2020 and Mar 13, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K112998); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Material Integrity Problem2
- Use of Device Problem2
- Fitting Problem1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2024Break; Material Integrity Problem; Break; Material Integrity Problem
DURING CHECKING LIFEBANDS BEFORE TREATMENT, THE CUSTOMER FOUND THE BAND CLIP AND HINGED BELT GUARD OF THE TWO NEWLY RECEIVED LIFEBANDS (LOT#190702) WERE BROKEN. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18896190 ↗MalfunctionMar 13, 2024Break; Material Integrity Problem; Break; Material Integrity Problem
DURING CHECKING LIFEBANDS BEFORE TREATMENT, THE CUSTOMER FOUND THE BAND CLIP AND HINGED BELT GUARD OF THE TWO NEWLY RECEIVED LIFEBANDS (LOT#190702) WERE BROKEN. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18896164 ↗MalfunctionOct 16, 2023Fitting Problem; Fitting Problem
DURING SHIFT CHECK, ONE OF THE HINGED BELT GUARDS OF THE LIFEBAND (LOT 173316) HAD SOME ISSUES AND WAS UNABLE TO ENGAGE TO THE ROLLER. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17941177 ↗MalfunctionMay 8, 2020Use of Device Problem; Use of Device Problem
DURING THE PATIENT USE, THE PLATFORM DISPLAYED USER ADVISORY (UA) 02 (COMPRESSION TRACKING ERROR) ERROR. THE LIFEBAND WAS NOTED TWISTED INTERMITTENTLY DURING THE OPERATION. THE USER REINSTALLED THE LIFEBAND TO MAKE SURE IT IS PROPERLY INSTALLED, HOWEVER, THE ISSUE PERSISTS. THE CREW IMMEDIATELY REVERTED TO MANUAL CPR. NO CONSEQUENCES OR IMPACT TO PATIENT. THE PLATFORM WAS TESTED BACK AT THE STATION AND WAS CONFIRMED TO PERFORM AS INTENDED WITH THE NEW LIFEBAND. PER A REPORTER, THE LIFEBAND WITH THE ISSUE WAS INSTALLED ON THE PLATFORM FOR ABOUT A MONTH. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: MFR 3010617000-2020-00484 FOR AUTOPULSE PLATFORM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10041026 ↗MalfunctionMay 7, 2020Material Twisted/Bent; Material Twisted/Bent
DURING PATIENT USE, CUSTOMER REPORTED THAT LIFEBAND WAS TWISTED INTERMITTENTLY ON THE AUTOPULSE PLATFORM (SERIAL #(B)(4)) WHILE OPERATING AND IT DISPLAYED ADVISORY (UA) 02 (COMPRESSION TRACKING ERROR) ERROR MESSAGE. PATIENT'S STATUS INFORMATION WAS REQUESTED BUT THE CUSTOMER DID NOT PROVIDE A RESPONSE, THEREFORE PATIENT'S STATUS IS UNKNOWN. THE AUTOPULSE PLATFORM ASSOCIATED WITH THIS COMPLAINT IS SUBMITTED UNDER MFR 3010617000-2020-00487.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10037040 ↗MalfunctionMay 6, 2020Use of Device Problem; Use of Device Problem
THE AUTOPULSE PLATFORM (SN: (B)(4)) WAS USED TO TREAT A PATIENT IN CARDIAC ARREST. THE PATIENT WAS PLACED ON THE PLATFORM, AND THE COMPRESSIONS WERE PERFORMED FOR ALMOST 30 MINUTES. THEN, THE PLATFORM STOPPED COMPRESSIONS AND DISPLAYED USER ADVISORY (UA) 02 (COMPRESSION TRACKING ERROR) ERROR MESSAGE. AS REPORTED, THE LIFEBAND (LOT # 143423) KEPT GETTING TWISTED INTERMITTENTLY, AND THE ISSUE WAS OBSERVED 2 OR 3 TIMES DURING THE USE OF THE PLATFORM. THE USER TRIED TO CLEAR THE ERROR MULTIPLE TIMES, BUT THE ISSUE WAS NOT RESOLVED. NO CONSEQUENCES OR IMPACT TO PATIENT. PLEASE SEE THE FOLLOWING RELATED MFR REPORT: MFR # 3010617000-2020-00482 FOR THE AUTOPULSE PLATFORM (SN: (B)(4)).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10030438 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112998. One submission often covers many models, so check the product and lot details in each record.
- Z-0118-2023 · Sep 22, 2022 · Open, ClassifiedAutoPulse Li-ion Battery used on the AutoPulse Resuscitation System Model 100. used as an adjunct to manual CPR
Other Compressor, Cardiac, External devices
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External3,208 reports
- AutoPulse Li-ion Battery™ 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,151 reports
- AutoPulse® Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,136 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External317 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External276 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External250 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External241 reports
- LUCAS 99576Jolife AB · Compressor, cardiac, external2 recalls9 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page