LifeBand® Load-distributing Band Model 200 (REF 8700-001013)Recalls & reported problems
Zoll Circulation, Inc. · Single-use chest compression band. (3 per package) · Class 2
All models of LifeBand Load-distributing Band (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10849111002988), received between Mar 13, 2025 and Mar 5, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem5
- Material Twisted/Bent4
- Retraction Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2026Material Twisted/Bent
THE CUSTOMER REPORTED THAT DURING A DEMO, THEY PLACED A TORSO MANIKIN ON THE AUTOPULSE NXT PLATFORM (B)(6). THE PLATFORM AND NXT BAND (LOT # UNKNOWN) KEPT PROVIDING COMPRESSIONS; HOWEVER, THE BAND TWISTED ON ITS OWN. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24527305 ↗MalfunctionApr 2, 2025Retraction Problem; Use of Device Problem; Material Twisted/Bent; Retraction Problem; Use of Device Problem; Material Twisted/Bent
DURING PATIENT RESUSCITATION WITH THE AUTOPULSE NXT PLATFORM, THE CREW OBSERVED A PROBLEM WITH AN NXT BAND (LOT #200888) THAT TWISTS INTERNALLY WITHIN THE DEVICE EVEN WHEN CORRECTLY ALIGNED ON THE PATIENT. THE FAILURE OCCURRED APPROXIMATELY 4 TO 5 MINUTES INTO OPERATION. A NOTICEABLE CHANGE IN THE SOUND OF THE AUTOPULSE MOTOR WAS NOTED, INDICATING THAT THE DEVICE WAS STRUGGLING AND BECOMING BOGGED DOWN. SHORTLY AFTER, IT STOPPED COMPRESSING. UPON ATTEMPTING A RESET, ONE SIDE OF THE BAND WAS DIFFICULT TO RETRACT, REVEALING THAT IT HAD BECOME BUNDLED WITHIN THE DEVICE. ADDITIONALLY, THE OUTER SLEEVE OF THE BAND APPEARED TO BE PULLED INTO ITSELF. AFTER RESETTING, THE BAND QUICKLY BECAME TANGLED AGAIN DUE TO THE CREASES AND FOLDS THAT HAD FORMED. ACCORDING TO THE CUSTOMER, THE PATIENT WAS OF AVERAGE SIZE. THE CREW REVERTED TO MANUAL CPR. NO PATIENT INJURIES WERE REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21755098 ↗MalfunctionApr 2, 2025Retraction Problem; Use of Device Problem; Material Twisted/Bent; Retraction Problem; Use of Device Problem; Material Twisted/Bent
DURING PATIENT RESUSCITATION WITH THE AUTOPULSE NXT PLATFORM, THE CREW OBSERVED A PROBLEM WITH AN NXT BAND (LOT #200888) THAT TWISTS INTERNALLY WITHIN THE DEVICE EVEN WHEN CORRECTLY ALIGNED ON THE PATIENT. THE FAILURE OCCURRED IMMEDIATELY UPON DEPLOYMENT. A NOTICEABLE CHANGE IN THE SOUND OF THE AUTOPULSE MOTOR WAS NOTED, INDICATING THAT THE DEVICE WAS STRUGGLING AND BECOMING BOGGED DOWN. UPON ATTEMPTING A RESET, ONE SIDE OF THE BAND WAS DIFFICULT TO RETRACT, REVEALING THAT IT HAD BECOME BUNDLED WITHIN THE DEVICE. ADDITIONALLY, THE OUTER SLEEVE OF THE BAND APPEARED TO BE PULLED INTO ITSELF. AFTER RESETTING, THE BAND QUICKLY BECAME TANGLED AGAIN DUE TO THE CREASES AND FOLDS THAT HAD FORMED. ACCORDING TO THE CUSTOMER, THE PATIENT WAS LARGER AND WELL WITHIN THE PARAMETERS OF THE AUTOPULSE. THE CREW REVERTED TO MANUAL CPR. NO PATIENT INJURIES WERE REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21755073 ↗MalfunctionApr 1, 2025Retraction Problem; Use of Device Problem; Material Twisted/Bent; Retraction Problem; Use of Device Problem; Material Twisted/Bent
DURING PATIENT RESUSCITATION WITH THE AUTOPULSE NXT PLATFORM, THE CREW OBSERVED A PROBLEM WITH AN NXT BAND (LOT #200888) THAT TWISTS INTERNALLY WITHIN THE DEVICE EVEN WHEN CORRECTLY ALIGNED ON THE PATIENT. THE FAILURE OCCURRED APPROXIMATELY 4 TO 5 MINUTES INTO OPERATION. A NOTICEABLE CHANGE IN THE SOUND OF THE AUTOPULSE MOTOR WAS NOTED, INDICATING THAT THE DEVICE WAS STRUGGLING AND BECOMING BOGGED DOWN. SHORTLY AFTER, IT STOPPED COMPRESSING. UPON ATTEMPTING A RESET, ONE SIDE OF THE BAND WAS DIFFICULT TO RETRACT, REVEALING THAT IT HAD BECOME BUNDLED WITHIN THE DEVICE. ADDITIONALLY, THE OUTER SLEEVE OF THE BAND APPEARED TO BE PULLED INTO ITSELF. AFTER RESETTING, THE BAND QUICKLY BECAME TANGLED AGAIN DUE TO THE CREASES AND FOLDS THAT HAD FORMED. ACCORDING TO THE CUSTOMER, THE PATIENT WAS OF AVERAGE SIZE. THE CREW REVERTED TO MANUAL CPR. NO PATIENT INJURIES WERE REPORTED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21748818 ↗MalfunctionMar 13, 2025Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTED THAT WHILE INSERTING THE NXT BAND PINS INTO THE AUTOPULSE NXT PLATFORM SLOT, IT FELT LIKE THE PINS WENT IN FURTHER THAN THEY SHOULD HAVE. THE CUSTOMER MENTIONED THAT THE PINS ON TWO NXT BANDS (LOT #UNKNOWN) WERE SLIGHTLY OFFSET COMPARED TO THOSE ON THE FUNCTIONING NXT BANDS. ADDITIONALLY, THE CUSTOMER NOTED THAT THE BANDS MAY HAVE BEEN INSERTED INCORRECTLY. THE CUSTOMER HAD NOT EXPERIENCED SIMILAR ISSUES WITH THE OTHER NXT BANDS, AS THOSE PINS APPEARED EVEN AND NOT OFFSET. THE REPORTED PROBLEM OCCURRED DURING THE TRAINING SESSION. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21599076 ↗MalfunctionMar 13, 2025Use of Device Problem; Use of Device Problem
THE CUSTOMER REPORTED THAT WHILE INSERTING THE NXT BAND PINS INTO THE AUTOPULSE NXT PLATFORM SLOT, IT FELT LIKE THE PINS WENT IN FURTHER THAN THEY SHOULD HAVE. THE CUSTOMER MENTIONED THAT THE PINS ON TWO NXT BANDS WERE SLIGHTLY OFFSET COMPARED TO THOSE ON THE FUNCTIONING NXT BANDS. ADDITIONALLY, THE CUSTOMER NOTED THAT THE BANDS MAY HAVE BEEN INSERTED INCORRECTLY. THE CUSTOMER HAD NOT EXPERIENCED SIMILAR ISSUES WITH THE OTHER NXT BANDS, AS THOSE PINS APPEARED EVEN AND NOT OFFSET. THE REPORTED PROBLEM OCCURRED DURING THE TRAINING SESSION. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21599075 ↗
Other Compressor, Cardiac, External devices
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External3,208 reports
- AutoPulse Li-ion Battery™ 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,151 reports
- AutoPulse® Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External1,136 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External317 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External276 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External250 reports
- AutoPulse®Resuscitation System 100Zoll Circulation, Inc. · Compressor, Cardiac, External241 reports
- LUCAS 99576Jolife AB · Compressor, cardiac, external2 recalls9 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page