Diamondback Peripheral DBP-200CLASS145Recalls & reported problems
Cardiovascular Systems, Inc. · DIAMONDBACK PERIPHERAL, CLASSIC CROWN, 2.00MM, 145CM SHAFT, 140 KRPM HIGH SPEED, 90 KRPM MEDIUM SPEED, 60 KRPM LOW SPEED · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10850000491257), received between May 17, 2021 and Mar 23, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Difficult to Remove1
- Material Separation1
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 23, 2023Difficult to Remove; Difficult to Remove
THE DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED TO TREAT A LESION IN THE PROFUNDA ARTERY VIA THE BRACHIAL APPROACH. DURING TREATMENT, THE OAD MADE AN UNUSUAL NOISE. THE PHYSICIAN ATTEMPTED TO REMOVE THE OAD, BUT EXPERIENCED RESISTANCE. BECAUSE OF THE RESISTANCE, THE PHYSICIAN DECIDED TO WAIT UNTIL A PRE-SCHEDULED ENDARTERECTOMY TO REMOVE THE OAD. THE OAD WAS SUCESSFULLY REMOVED AND THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16600666 ↗InjuryMay 17, 2021Material Separation; Material Separation
WHEN TREATMENT WITH THE DIAMONDBACK PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS COMPLETED, THE DEVICE WOULD NOT MOVE DURING REMOVAL ATTEMPTS. THE PHYSICIAN PULLED ON THE OAD AND THE CROWN WAS OBSERVED TO DETACH IN THE ANTERIOR TIBIAL ARTERY. THE CROWN WAS REMOVED USING A SNARE AND THE PATIENT WAS IN STABLE CONDITION POST-PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 11835308 ↗
Other Catheter, Peripheral, Atherectomy devices
- HawkOne™ H1-MCovidien LP · Catheter, peripheral, atherectomy1 recall367 reports
- JETSTREAM™ XC 112266-001Boston Scientific Corporation · Catheter, peripheral, atherectomy288 reports
- JETSTREAM™ XC 112264-001Boston Scientific Corporation · Catheter, peripheral, atherectomy262 reports
- SET Rotarex®S 6F 135cm (Straub Medical®) Version 1.0Straub Medical AG · Catheter, Peripheral, Atherectomy2 deaths206 reports
- SET Rotarex™S 6 F x 135 cm 80237Straub Medical AG · Catheter, Peripheral, Atherectomy1 recall79 reports
- HawkOne™ H1-SCovidien LP · Catheter, peripheral, atherectomy1 recall70 reports
- Turbo-Elite Laser Atherectomy Catheter 420-006Spectranetics Corporation · Catheter, Peripheral, Atherectomy2 recalls19 reports
- Turbo-Elite Laser Atherectomy Catheter 417-152Spectranetics Corporation · Catheter, Peripheral, Atherectomy2 recalls16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page