Skip to content

Stealth RAD-SC30-150 (REF 7-10059-06)Recalls & reported problems

Cardiovascular Systems, Inc. · STEALTH Peripheral, Radial Access OAD, 1.50mm, 10mm, 200cm shaft, 120 krpm High Speed, 90 krpm Medium Speed, 60 krpm Low Speed · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10850000491318), received between Jun 10, 2021 and Jan 21, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert1
  • Difficult to Remove1
  • Fluid/Blood Leak1
  • Material Separation1
  • Mechanical Problem1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Extravasation1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2025Fluid/Blood Leak; Difficult to Insert; Fluid/Blood Leak; Difficult to Insert

    IT WAS REPORTED DURING THE PROCEDURE, THE STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS LEAKING FROM THE BASE AND THE VIPERWIRE GUIDE WIRE WAS UNABLE TO BE ADVANCED THROUGH THE OAD. A NEW OAD WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200613 ↗
  • MalfunctionSep 19, 2024Difficult to Remove; Material Separation; Unexpected Shutdown; Difficult to Remove; Material Separation; Unexpected Shutdown

    THE STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED FOR TREATMENT IN THE COMMON FEMORAL. THE OAD WAS SPUN ON LOW SPEED AND MEDIUM SPEED WITHOUT ISSUE. WHILE SPINNING THE OAD ON HIGH SPEED, THE OAD STALLED AND BECAME STUCK ON THE VIPERWIRE ADVANCE GUIDE WIRE. BOTH THE OAD AND VIPERWIRE WERE REMOVED TOGETHER. THE VESSEL WAS REWIRED WITH A NON-ABBOTT GUIDEWIRE AND ANGIOPLASTY WAS PERFORMED TO COMPLETE THE PROCEDURE. THE PATIENT WAS STABLE. ADDITIONAL INFORMATION WAS RECEIVED DURING DEVICE ANALYSIS THAT INDICATED THE DRIVESHAFT FRACTURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20256082 ↗
  • InjuryJun 10, 2021Mechanical Problem; Mechanical Problem

    A STEALTH PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED VIA LEFT RADIAL ACCESS TO SUCCESSFULLY TREAT LESIONS IN THE SUPERFICIAL FEMORAL, POPLITEAL, AND TIBIOPERONEAL TRUNK ARTERIES. THE OAD WAS ADVANCED TO THE ANTERIOR TIBIAL ARTERY AND TREATMENT WAS STARTED. THE OAD STOPPED SPINNING AND WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH BALLOON ANGIOPLASTY. DURING RECOVERY, THE PATIENT COMPLAINED OF FOOT PAIN UPON WAKING UP FROM SEDATION. ULTRASOUND WAS PERFORMED AND FLOW WAS CONFIRMED TO BE PRESENT. IT WAS DETERMINED THAT EXTRAVASATION HAD OCCURRED AND THE PATIENT WAS TRANSFERRED TO A HOSPITAL WHERE A FASCIOTOMY WAS PERFORMED. THE PATIENT WAS IN STABLE CONDITION FOLLOWING THIS PROCEDURE.

    Patient problems: Extravasation; Pain

    MAUDE report 11979025 ↗