Skip to content

Stealth RAD-SC30-175 (REF 7-10059-07)Recalls & reported problems

Cardiovascular Systems, Inc. · STEALTH Peripheral, Radial Access OAD, 1.75mm, 30mm, 180cm shaft, 120 krpm High Speed, 90 krpm Medium Speed, 60 krpm Low Speed · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10850000491325), received between Aug 9, 2023 and Nov 15, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Contaminated at the User Facility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contaminated at the User Facility1
  • Device Contamination with Body Fluid1
  • Entrapment of Device1
  • Material Separation1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 15, 2024Entrapment of Device; Material Separation; Device Contaminated at the User Facility; Entrapment of Device; Material Separation; Device Contaminated at the User Facility

    IT WAS REPORTED A STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE WAS USED TO TREAT A HEAVILY CALCIFIED, 90% STENOSED POPLITEAL ARTERY. DURING THE PROCEDURE, THE OAD CROWN BECAME STUCK IN THE SMALL BRANCH OF THE POPLITEAL ARTERY. IT WAS INDICATED THE VESSEL WAS TOO SMALL FOR THE OAD CROWN, WHICH LED TO THE OAD CROWN BECOMING STUCK. THE PHYSICIAN CHANGED ACCESS FROM THE LEFT RADIAL TO THE RIGHT FEMORAL. NITROGLYCERIN AND VERAPAMIL WERE ADMINISTERED, FOLLOWED BY BALLOON ANGIOPLASTY TO COMPLETE THE PROCEDURE. IT WAS INDICATED TISSUE WAS OBSERVED ON THE DEVICE EX-VIVO. THE PATIENT WAS IN STABLE CONDITION. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE OAD CROWN DETACHED. NO COMPONENTS WERE LEFT IN-VIVO.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 20696536 ↗
  • MalfunctionAug 9, 2023Device Contamination with Body Fluid; Device Contamination with Body Fluid

    A STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED FOR SIX TREATMENTS. AFTER THE TREATMENTS, IT WAS NOTICED THAT THE AMOUNT OF VIPERSLIDE IN THE BAG WAS NOT DECREASING DESPITE A LOT OF FLUSHING. BLOOD WAS OBSERVED IN THE SALINE TUBING. THE TREATMENT WAS COMPLETED WITH NO FURTHER ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17487856 ↗

Other Catheter, Peripheral, Atherectomy devices

All Catheter, Peripheral, Atherectomy devices →