MC3 Crescent RA Dual Lumen Catheter 15FR 70415Recalls & reported problems
Michigan Critical Care Consultants, Inc. · The Jugular Dual Lumen Catheter is a single use dual lumen catheter, which provides both venous drainage and reinfusion of blood via the jugular vein, that is indicated for use in adult and pediatric patients with acute respiratory failure requiring VenoVenous Extracorporeal Membrane Oxygenation where other available treatment options have failed and continued clinical deterioration is expected or the risk of death is imminent. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10850007732216), received between May 7, 2022 and Jul 30, 2025 — classified as 1 death, 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Material Deformation1
- Mechanical Problem1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Respiratory Insufficiency1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2025Mechanical Problem; Material Deformation; Mechanical Problem; Material Deformation
IT WAS REPORTED THAT DURING USE THE MC3 CRESCENT 15FR RIGHT ATRIAL CANNULA WAS FOUND TO HAVE A BEND AT THE TIP OF THE CANNULA, WHICH PREVENTED THE DEVICE FROM BEING ADVANCED. THE SURGEON OPENED A NEW CANNULA AND COMPLETED THE INSERTION WITHOUT FURTHER ISSUE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22662303 ↗InjuryJan 12, 2023Fracture; Fracture
ON (B)(6) 2022, AN 8 MONTH OLD PATIENT WAS CANNULATED WITH A 15 FR CRESCENT RA DUAL LUMEN CATHETER. THE PATIENT WAS DESCRIBED AS BEING VERY ACTIVE AND NOT SEDATED. ON (B)(6) 2022, THE PATIENT WAS OFF ECMO FOR A BRIEF PERIOD OF TIME DURING EXCHANGE PROCEDURE DUE TO REPORTED CATHETER FRACTURE. THE PATIENT DID NOT EXPERIENCE ANY INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16148086 ↗InjuryDec 22, 2022Fracture; Fracture
ADDITIONAL INFORMATION RECEIVED INDICATED 3011468686-2022-00020 AND 3011468686-2022-00022 ARE THE SAME PATIENT. THE PATIENT, CONFIRMED TO BE 8 MONTHS OLD, WITH A HISTORY OF BRONCHOPULMONARY DYSPLASIA WAS TREATED WITH ECMO. ON (B)(6) 2022, THE PATIENT WAS INITIALLY CANNULATED. ON (B)(6) 2022 AND (B)(6) 2022, THE PATIENT WAS RE-CANNULATED THROUGH THE RIGHT IJ. THE PATIENT WAS NOT SEDATED. ON (B)(6) 2022, THE PATIENT WAS DECANNULATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16036277 ↗DeathMay 7, 2022Obstruction of Flow; Obstruction of Flow
MC3 RECEIVED INFORMATION ON (B)(6) 2022 THAT DURING USE OF A 15 FR MC3 CRESCENT CATHETER AND A 15FR BIO-MEDICUS CANNULA AN ADVERSE EVENT OCCURRED ON (B)(6) 2022 DURING AN UNSUCCESSFUL ATTEMPT TO INITIATE ADEQUATE CARDIOPULMONARY SUPPORT. THE CUSTOMER REPORTED THAT THE 15 FR CRESCENT CATHETER WAS PLACED IN THE RIGHT INTERNAL JUGULAR VEIN (RIJV) PERCUTANEOUSLY. NO FLUOROSCOPY WAS USED FOR CANNULATION, INSTEAD AN ECHOCARDIOGRAM WAS PERFORMED AT THE BEDSIDE. THE CIRCUIT WAS ATTACHED TO THE CATHETERS AND RPMS INCREASED TO 2000, AND LPMS COULD NOT INCREASE HIGHER THAN 350(ML/MIN). THE CUSTOMER REPORTED THAT THE VENOUS RETURNS WERE POSITIVE, AND THE MEMBRANE PRESSURES WERE IN THE 90'S. THE SURGEON MANIPULATED THE CATHETER WITH NO CHANGE, ECHOCARDIOGRAM WAS PERFORMED AT BEDSIDE. SURGEON ADDED A 15FR BIO-MEDICUS CANNULA AND Y'D IN WITH THE MC3 CRESCENT CATHETER, THEY WERE STILL NOT ABLE TO GET FLOW. THE SVO2 REMAINED IN 90'S, ALERTING THE CUSTOMER THAT THEY WERE JUST RECIRCULATING. THE CUSTOMER STATED THAT BOTH CATHETERS WHEN UNCLAMPED HAD GREAT BLOOD RETURN AND WERE ABLE TO FLUSH. THE PATIENT BEGAN TO CODE ONCE THEY WERE UNABLE TO FLOW ADEQUATELY ON EXTRACORPOREAL MEMBRANE OXYGENATION (ECM
Patient problems: Respiratory Insufficiency
MAUDE report 14331960 ↗
Other Dual Lumen Ecmo Cannula devices
- MC3 Crescent Jugular Dual Lumen Catheter 70132Michigan Critical Care Consultants, Inc. · Dual Lumen Ecmo Cannula9 reports
- MC3 Crescent Jugular Dual Lumen Catheter 70128Michigan Critical Care Consultants, Inc. · Dual Lumen Ecmo Cannula5 reports
- MC3 Crescent RA Dual Lumen Catheter 13FR 70413Michigan Critical Care Consultants, Inc. · Dual Lumen Ecmo Cannula1 death1 report
- MC3 Crescent Jugular Dual Lumen Catheter 70130Michigan Critical Care Consultants, Inc. · Dual Lumen Ecmo Cannula1 report
- MC3 Crescent(tm) Jugular Dual Lumen Catheter 70128Michigan Critical Care Consultants, Inc. · Dual Lumen Ecmo Cannula1 report
- MC3 Crescent Jugular Dual Lumen Catheter 70126Michigan Critical Care Consultants, Inc. · Dual Lumen Ecmo Cannula1 report
- MC3 Crescent Jugular Dual Lumen Catheter 70124Michigan Critical Care Consultants, Inc. · Dual Lumen Ecmo Cannula1 report
- PROTEK DUO+ 5150-5131Sorin Group Italia Srl · Dual lumen ECMO cannula1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page