VitalFlow Set 58112Recalls & reported problems

Michigan Critical Care Consultants, Inc. · The VitalFlow set contains components used to complete an extracorporeal circuit for ECMO procedures, including tubing assemblies, the Nautilus VF ECMO Oxygenator, and the VitalFlow Centrifugal Pump. The VitalFlow set contains a preassembled drainage and return loop “ECMO Circuit”, tubing assemblies, and other components used to prepare a basic extracorporeal circuit. The “Priming Circuit” contains components to supplement the basic extracorporeal circuit, as needed per hospital protocols for setting up the ECMO circuit. A venous reservoir and a one-way valve are included in the priming circuit to facilitate ease of priming the ECMO circuit. The Nautilus VF ECMO oxygenator with Balance Biosurface is used to oxygenate blood, remove carbon dioxide and regulate blood temperature. Blood enters the device and passes through both the heat exchange membrane, where temperature is adjusted, and the gas transfer membrane, where oxygen is added and carbon dioxide is removed. The VitalFlow Centrifugal Pump is designed to move blood by centrifugal force generated by a combination of a smooth rotating cone and low-profile impeller fins. Energy is transferred from the pump in the form of pressure and velocity as the blood is driven toward the outlet port of the pump. To limit friction and heat generation, the VitalFlow centrifugal pump utilizes a pivot bearing design on a dual ceramic pivot. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10850007732339), received between Jan 8, 2025 and Aug 7, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak3
  • Obstruction of Flow3
  • Noise, Audible2
  • Crack1
  • Device Alarm System1
  • Increase in Pressure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 7, 2026Obstruction of Flow; Noise, Audible

    Medtronic received information that during use of a Vital flow tubing Set, it was reported that the patient was on the circuit, status post (s/p) left ventricular assist device (LVAD) insertion with a 29FR, right ventricular assist device (RVAD) placed via the right internal jugular (RIJ) vein. The patient was awake and alert the whole time. And doing fine compared to prior. There was High Flow and Pressures since insertion. The Extracorporeal Membrane Oxygenation (ECMO) specialist heard a decoupling noise and began trouble shooting with a hand crank for an hour until exchange was complete. Noise continued but the customer was not sure if noise completely stopped, but the flow reduced on the data. The device was replaced to complete the procedure. There was no adverse patient effect associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50175440 ↗
  • InjuryAug 7, 2026Obstruction of Flow

    Medtronic received information that during use of a Vital flow tubing set, it was reported that the device had a flow/ clot issues. The biomed believes the device sucked a clot from the left ventricle of the patient into the pump head. Flow stopped and a hand crank was used for roughly 10 mins, and this provided some flow. Customer placed the cone back to the motor drive and was able to get some flow The customer believed the clot in the left ventricle of the heart became dislodged and migrated into the motor drive. The patient was previously on heparin, the issue did not worsen over time. The RPM at time of event was 3800, the flow at time of event- 0.33LPM, and the only blood loss was the blood in the clotted off circuit. Anticoagulation used Heparin 100 units/hr, not ACTs were collected, aPTT 40-50. Plasmalyte was used to prime the circuit. The location of this issue was at the cone and pre ox side of vital flow membrane oxygenator. The device was replaced to complete the procedure. There was no adverse patient effect associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50171911 ↗
  • MalfunctionMar 23, 2026Obstruction of Flow; Noise, Audible

    IT WAS REPORTED THAT DURING USE, THE VITALFLOW TUBE SET PUMP EXHIBITED A GRINDING NOISE, PUMP FAILURE, AND THE PUMP STOPPED. THE DEVICE WAS REPLACED TO COMPLETE THE CASE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24660556 ↗
  • MalfunctionFeb 10, 2026Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A VITALFLOW SET, THE CUSTOMER REPORTED THAT THE OXYGENATOR HAD A LEAKING TUBING CONNECTION. THE DEVICE WAS REPLACED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24301923 ↗
  • MalfunctionJan 8, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT A LEAK WAS IDENTIFIED ON THE OUTLET SIDE AT THE PINK PORT OF THE MC3 VITALFLOW NAUTILUS EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) OXYGENATOR PRIOR TO USE. THERE WAS NO NOTED DAMAGE TO THE PACKAGING DURING SETUP, AND NO OTHER COMPONENTS WERE AFFECTED. THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE SERIAL NUMBER OF THE OXYGENATOR IS (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24001992 ↗
  • MalfunctionNov 27, 2025Device Alarm System; Crack; Fluid/Blood Leak; Device Alarm System; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT THE MC3 VITAL FLOW NAUTILUS EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) OXYGENATOR EXPERIENCED AN UNRESOLVABLE OXYGENATOR FAILURE, WITH ERRORS E47 (OXY COMMUNICATION LOST) AND E49 (OXY ERROR - OUTLET BOARD) OCCURRING WHILE THE DEVICE WAS RUNNING IN PRIME MODE PRIOR TO PATIENT USE. THE E47 ERROR WAS OBSERVED ON A NEW CIRCUIT AFTER THE DEVICE INITIALLY FUNCTIONED APPROPRIATELY, OCCURRING APPROXIMATELY 12 HOURS AFTER PRIMING, AND WAS PRECEDED BY THE E49 ERROR. THEY LEAVE THEIR PRIMED BACKUP CONSOLES RUNNING IN PRIME MODE IN ORDER FOR A SPECIALIST TO DO CIRCUIT CHECKS DURING SHIFT. THEY QUICKLY RULED OUT BOTH CABLE AND CONSOLE AS THE SOURCE AND FOCUSED ON THE OXYGENATOR ITSELF. THEY WERE STORED IN A LOCKED PUMP ROOM. MOISTURE DROPLETS WERE OBSERVED IN THE OUTLET SENSOR HOUSING, AND FLUID WAS OBSERVED LEAKING FROM THE BOTTOM OF THE DEVICE WHEN IT WAS REMOVED FROM THE MOUNTING BRACKET. NO HEATER C OOLER WATER HOSES HAD BEEN CONNECTED, INDICATING THE FLUID ORIGINATED FROM PRIMING. ALTHOUGH NO VISIBLE EXTERNAL DAMAGE OR CRACKS WERE SEEN, MOISTURE WAS EVIDENT IN THE ARTERIAL SENSOR HOUSING, SUGGESTING A POSSIBLE CRACK, BREAK, OR HOLE AT THAT LOCATION. A COMPLETE CRACK/BR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23666010 ↗
  • MalfunctionJan 8, 2025Increase in Pressure; Increase in Pressure

    IT WAS REPORTED THAT PRIOR TO USE OF A MC3 NAUTILUS VITALFLOW (VF) EXTRA CORPOREAL MEMBRANE OXYGENATION (ECMO) OXYGENATOR,THE INLET PRESSURES WERE NOT REGISTERING . THE ISSUE WAS ADDRESSED BY UNPLUGGING THE NAUTILUS AND PLUGGING IT BACK IN, AFTER WHICH IT WAS CORRECTED AND WORKED PROPERLY. THE USE OF THE DEVICE WAS CONTINUED TO COMPLETE THE CASE.THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21099607 ↗

FDA clearance and classification

Cleared through 510(k) K240880 on Sep 12, 2024 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “VitalFlow Set with Balance Biosurface”. The FDA review took 164 days. FDA record ↗

The FDA regulates product code QJZ (extracorporeal system for long-term respiratory / cardiopulmonary failure) under 21 CFR 870.4100 as a Class II device. 21 CFR 870.4100 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Common questions about the VitalFlow Set 58112

Has the VitalFlow Set 58112 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10850007732339), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VitalFlow Set 58112?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Jan 8, 2025 and Aug 7, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (3), “Obstruction of Flow” (3) and “Noise, Audible” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VitalFlow Set 58112?

None of the reports are classified as deaths, 2 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VitalFlow Set 58112 FDA approved?

It was cleared through 510(k) K240880 on Sep 12, 2024. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the VitalFlow Set 58112 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the VitalFlow Set 58112, and is it still sold?

The FDA's device database lists Michigan Critical Care Consultants, Inc. as the company responsible for it (brand VitalFlow Set, model 58112). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure devices

All Extracorporeal System For Long-Term Respiratory / Cardiopulmonary Failure devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page