Skip to content

NaviNetics Skull Anchor Key and Drill NN1215Recalls & reported problems

Navinetics, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 26, 2024Class IIOpen, Classified

    Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.

    Z-1478-2024 · Root cause: Process control · Recall details · FDA record ↗

Other Neurological Stereotaxic Instrument devices

All Neurological Stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page