STEALTH PRD-MAX-200Recalls & reported problems
Cardiovascular Systems, Inc. · STEALTH, 2.00 MAX CROWN, 145CM SHAFT · Class 2
All models of Stealth (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10850026568711), received between Sep 2, 2022 and Aug 12, 2024 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Difficult to Remove1
- Material Separation1
Patient problems
- Vascular Dissection3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Discomfort1
- Embolism/Embolus1
- Intraoperative Pain1
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Kidney or Urinary Problem1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 12, 2024Difficult to Remove; Difficult to Remove
A STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS SUCCESSFULLY ADVANCED THROUGH THE 6 X 45 SHEATH WITH NO RESISTANCE MET. SIX LOW SPEED TREATMENTS WERE PERFORMED DOWN THE ENTIRETY OF A HEAVILY CALCIFIED, 30-40% STENOSED, 5MM LEFT SUPERFICIAL FEMORAL ARTERY (SFA). THE PATIENT WAS UNCOMFORTABLE DURING THE TREATMENT. POST-TREATMENT, THE OAD WAS ATTEMPTED TO BE REMOVED FROM THE PATIENT WHEN RESISTANCE WAS MET AT THE DISTAL PART OF THE SHEATH. THE OAD CONTROL KNOB WAS IN LOCKED POSITION AND THE CROWN WAS NOT MOVING IN EITHER DIRECTION. THE SHEATH WAS FLUSHED AND NO MOVEMENT CONTINUED. A BUDDY WIRE WAS USED IN AN ATTEMPT TO DISLODGE THE OAD CROWN. THE OAD WAS ABLE TO BE REMOVED AFTER THE BUDDY WIRE WAS ABLE TO DISLODGE THE OAD CROWN. THE PHYSICIAN DID NOT OBSERVE ANY ISSUES WITH THE WIRE AND THE OAD.
Patient problems: Discomfort
MAUDE report 19960773 ↗MalfunctionMay 22, 2023Material Separation; Material Separation
A STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS BEING USED TO TREAT A SEVERELY CALCIFIED LESION IN THE RIGHT COMMON FEMORAL ARTERY. AS THE OAD WAS ADVANCED, A HIGH PITCH NOISE WAS HEARD AND THE OAD STOPPED ADVANCING THROUGH THE LESION. THE OAD WAS ATTEMPTED TO BE ADVANCED THREE TIMES BEFORE THE PHYSICIAN STOPPED THE PROCEDURE TO EVALUATE THE OAD. THE SALINE SHEATH APPEARED TO BE SEPARATED, BUT NOT DETACHED. A NEW OAD WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16973892 ↗InjuryMay 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED FOR TREATMENT IN THE SUPERFICIAL FEMORAL ARTERY (SFA), WHICH HAD SOFT CALCIUM. THE OAD WAS SPUN ON LOW SPEED FOR THREE TREATMENTS AND THEN SPUN ON MEDIUM SPEED. THE PATIENT EXPRESSED PAIN AND THE OAD PRODUCED AN UNUSUAL NOISE. AT THIS TIME, THE PHYSICIAN FELT THE OAD "BUCKLED" AND THEREFORE REMOVED THE OAD. INTRAVASCULAR ULTRASOUND (IVUS) WAS REVIEWED, AND NO DISSECTION WAS OBSERVED. THE OAD WAS REMOVED, AND TISSUE WAS OBSERVED ON THE DRIVESHAFT. BASED ON THIS, THE PHYSICIAN BELIEVED A DISSECTION MAY HAVE OCCURRED AT THE OSTIUM OF THE VESSEL. ANGIOPLASTY AND STENT PLACEMENT WERE PERFORMED TO COMPLETE THE PROCEDURE. THE PATIENT WAS STABLE.
Patient problems: Intraoperative Pain; Vascular Dissection
MAUDE report 16964874 ↗InjuryApr 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED TO TREAT A MID-SUPERFICIAL FEMORAL ARTERY (SFA) OF A 5 CM MIXED MORPHOLOGY LESION. THERE WAS A THREE-VESSEL RUNOFF, ALL OF WHICH HAD EMBOLISMS, AND TWO VESSELS WERE OCCLUDED. TISSUE PLASMINOGEN ACTIVATOR (TPA), NITROGLYCERIN, AND MANUAL ASPIRATION WERE USED TO RESOLVE THE EVENT.
Patient problems: Embolism/Embolus
MAUDE report 16707619 ↗InjuryFeb 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A STEALTH 360 ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED TO TREAT A LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). THE OAD WAS ADVANCED OVER THE VIPERWIRE ADVANCE GUIDE WIRE TO CROSS THE LESION. A DISTAL SLOW FLOW AND DISSECTION WERE OBSERVED AND RESOLVED AFTER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) WAS PERFORMED ON THE DISTAL VESSEL. NITROGLYCERIN AND ADENOSINE WERE ADMINISTERED. THE PHYSICIAN STATED THEY SHOULD HAVE REPOSITIONED THE WIRE BEFORE PLACING THE OAD CROWN PROXIMAL TO THE LESION. IN THE PHYSICIAN'S OPINION, THE CAUSE OF THE DISSECTION WAS DUE TO THE CROWN BEING ADVANCED THROUGH THE LESION WHILE NOT BEING TURNED ON. THE PATIENT WAS IN STABLE CONDITION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Vascular Dissection
MAUDE report 16404512 ↗InjuryDec 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INTRAVASCULAR ULTRASOUND (IVUS) OF THE VESSELS REVEALED MIXED MORPHOLOGY PLAQUE. ATHERECTOMY WAS PERFORMED IN THE LEFT COMMON FEMORAL ARTERY (CFA), SUPERFICIAL FEMORAL ARTERY (SFA) AND POPLITEAL ARTERY USING A DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD). ACCESS WAS GAINED IN THE CFA WITH A 6FR CATHETER. TWO 20 SECOND TREATMENTS WERE PERFORMED ON LOW SPEED, FOLLOWED BY TWO 20 SECOND TREATMENTS ON MEDIUM SPEED IN THE POPLITEAL ARTERY WITH 30 SECOND BREAKS BETWEEN TREATMENTS. VASODILATORS WERE ADMINISTERED FOLLOWING EACH TREATMENT. IVUS WAS PERFORMED AND REVEALED A SPIRAL, FLOW LIMITING DISSECTION FROM THE CFA TO THE MID SFA. BALLOON ANGIOPLASTY WAS PERFORMED. A STENT WAS DEPLOYED AND COVERED THE DISSECTION. IVUS WAS PERFORMED AND REVEALED THE DISSECTION TO HAVE RESOLVED. FOLLOWING THE PROCEDURE, THE PATIENT WAS IN STABLE CONDITION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Vascular Dissection
MAUDE report 15931680 ↗InjurySep 2, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE STEALTH 360 PERIPHERAL ORBITAL ATHERECTOMY DEVICE WAS USED FOR TREATMENT TWENTY TIMES AT TWENTY SECONDS. THE PROCEDURE WENT WELL WITH NO COMPLICATIONS. POST OPERATION, THE PATIENT EXPERIENCED DARK URINE DUE TO MECHANICAL DAMAGE OF RED BLOOD CELLS DURING THE PROCEDURE. THE HEMOGLOBINURIA WAS SUSPECTED TO BE CAUSED BY THE DEVICE. THE PATIENT WAS GIVEN EXTRA FLUIDS TO RESOLVE THE HEMOGLOBINURIA. THE PATIENT WAS STABLE.
Patient problems: Unspecified Kidney or Urinary Problem
MAUDE report 15347987 ↗
Other Catheter, Peripheral, Atherectomy devices
- HawkOne™ H1-MCovidien LP · Catheter, peripheral, atherectomy1 recall367 reports
- JETSTREAM™ XC 112266-001Boston Scientific Corporation · Catheter, peripheral, atherectomy288 reports
- JETSTREAM™ XC 112264-001Boston Scientific Corporation · Catheter, peripheral, atherectomy262 reports
- SET Rotarex®S 6F 135cm (Straub Medical®) Version 1.0Straub Medical AG · Catheter, Peripheral, Atherectomy2 deaths206 reports
- SET Rotarex™S 6 F x 135 cm 80237Straub Medical AG · Catheter, Peripheral, Atherectomy1 recall79 reports
- HawkOne™ H1-SCovidien LP · Catheter, peripheral, atherectomy1 recall70 reports
- Turbo-Elite Laser Atherectomy Catheter 420-006Spectranetics Corporation · Catheter, Peripheral, Atherectomy2 recalls19 reports
- Turbo-Elite Laser Atherectomy Catheter 417-152Spectranetics Corporation · Catheter, Peripheral, Atherectomy2 recalls16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page