Basic Series Hydrophobic Dressing Kit, Small 8463540 0583Recalls & reported problems

Pensar Medical, LLC · Basic Series Hydrophobic Dressing Kit, Small, Case of 10 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10851978007228), received between Jun 19, 2026 and Jun 19, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Gas/Air Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2026Material Separation; Gas/Air Leak

    CUSTOMER MEDCO - LOUISIANA REPORTED QUALITY ISSUES WITH THE BASIC HYDROPHOBIC SMALL WOUND DRESSING KIT (P/N 0583, LOT 1596) AND BASIC HYDROPHOBIC MEDIUM KIT (P/N 0584, LOT 1614); TUBING WAS OBSERVED TO PULL APART OR BREAK AT THE STINGRAY CONNECTION. TWO ADDITIONAL UNITS OPENED BY THE CUSTOMER DID NOT EXHIBIT THE SAME FAILURE. DEVICES WERE NOT RETURNED. NO PATIENT WAS REPORTED TO HAVE BEEN INVOLVED. REPLACEMENT PRODUCT PROVIDED. CONSERVATIVE DETERMINATION: MDR REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25566391 ↗

Other Negative Pressure Wound Therapy Powered Suction Pump devices

All Negative Pressure Wound Therapy Powered Suction Pump devices →

Sources and updates