Diamondback Peripheral DBP-EX-150CLA145Recalls & reported problems

Cardiovascular Systems, Inc. · Diamondback Peripheral, Exchangeable Series, 1.50 Classic, 145 cm shaft, 140 krpm High Speed, 90 krpm Medium Speed, 60 krpm Low Speed · Class 2

All models of Diamondback Peripheral (33)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10852528005909), received between Apr 26, 2021 and Jan 8, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Entrapment of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Entrapment of Device2

Patient problems

  • Device Embedded In Tissue or Plaque2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 8, 2024Entrapment of Device; Entrapment of Device

    THE DIAMONDBACK 360 EXCHANGEABLE PERIPHERAL ORBITAL ATHERECTOMY DEVICE (OAD) WAS ADVANCED TO A 3MM. 95% STENOSED AND HEAVILY CALCIFIED MID DISTAL POSTERIOR TIBIAL (PT) AND SPUN ON LOW SPEED AND MEDIUM SPEED FOR TREATMENT. THE OAD DRIVESHAFT WAS ADVANCED THE TOTAL LENGTH AND DURING THE PULL BACK BECAME STUCK. TROUBLESHOOTING STEPS WERE MADE TO FREE THE OAD BUT WERE UNSUCCESSFUL. A GUIDE CATHETER WAS ADVANCED TO ATTEMPT TO FREE THE OAD BUT WAS UNSUCCESSFUL. SURGERY WAS PERFORMED TO REMOVE THE OAD. THE PATIENT WAS STABLE AND DISCHARGED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 18463455 ↗
  • InjuryApr 26, 2021Entrapment of Device; Entrapment of Device

    OCCLUDED LESIONS IN THE SUPERFICIAL FEMORAL (SFA), POSTERIOR TIBIAL (PT) AND ANTERIOR TIBIAL (AT) ARTERIES WERE CROSSED FROM AN ANTEGRADE APPROACH. THE AT WAS ACCESSED VIA THE PEDAL ARCH AND A DIAMONDBACK PERIPHERAL EXCHANGEABLE ORBITAL ATHERECTOMY DEVICE (OAD) WAS USED TO TREAT THE PT. GLIDEASSIST WAS USED TO NAVIGATE THE OAD AROUND THE PEDAL ARCH AND TREATMENT WAS PERFORMED IN THE AT. WHEN TREATMENT WITH THE OAD WAS COMPLETED, GLIDEASSIST WAS ACTIVATED TO REMOVE THE DEVICE, HOWEVER IT WAS STUCK IN THE VESSEL. MULTIPLE ATTEMPTS WERE MADE TO REMOVE THE DEVICE, BUT THEY WERE UNSUCCESSFUL. CUT DOWN SURGERY WAS PERFORMED TO REMOVE THE DEVICE AND THE PATIENT WAS STABLE FOLLOWING THE PROCEDURE.

    Patient problems: Device Embedded In Tissue or Plaque