Skip to content

CerebroFlo EVD Catheter Set 37.550.501Recalls & reported problems

Arkis Biosciences Inc. · The CerebroFlo EVD Catheter is a polyurethane catheter for diverting cerebrospinal fluid (CSF) from the ventricles of the brain through the drainage holes near the catheter's tip. The catheter is barium-sulfate impregnated to provide radiopacity. The CerebroFlo EVD Catheter is provided with a Stylette, Trocar, Barbed Leur Connector, Male Leur Cap, and a suture clip. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10866614000105), received between Mar 17, 2022 and Mar 17, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

1 other recall cover devices cleared under the same FDA submission (K170599); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 17, 2022Leak/Splash; Leak/Splash

    CLINICAL STUDY: PATIENT HAD ON (B)(6) 2021 A CEREBROFLO EVD CATHETER PLACED TO TREAT SECONDARY INTRAVENTRICULAR HEMORRHAGE (IVH) DUE TO RUPTURED ANEURYSM. ON (B)(6) 2021 IT WAS REPORTED THE CEREBROFLO EVD CATHETER WAS LEAKING AT EXIST SITE. A CT SCAN ON (B)(6) REVEALS SOME AIR IN THE VENTRICLES AND IT WAS CONSIDERED REPLACING HER EVD, BUT WAS DEFERRED AS IT WAS FELT THE AIR WAS ENTRAINED FROM AROUND THE EVD RATHER THAN DUE TO MALFUNCTION OF EVD. AN ADDITIONAL SUTURE WAS PLACED AT THE EXIT SITE AND THE OUTCOME WAS RECOVERED/RESOLVED AND THE CATHETER WAS REMOVED ON (B)(6) 2022.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13790695 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170599. One submission often covers many models, so check the product and lot details in each record.

  • Z-2173-2026 · Apr 10, 2026 · Open, Classified
    Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD Catheter is a 3.3 mm diameter (10 Fr.), 35 cm long polyurethane

Other Shunt, Central Nervous System And Components devices

All Shunt, Central Nervous System And Components devices →