CerebroFlo EVD Catheter Set 37.550.501Recalls & reported problems
Arkis Biosciences Inc. · The CerebroFlo EVD Catheter is a polyurethane catheter for diverting cerebrospinal fluid (CSF) from the ventricles of the brain through the drainage holes near the catheter's tip. The catheter is barium-sulfate impregnated to provide radiopacity. The CerebroFlo EVD Catheter is provided with a Stylette, Trocar, Barbed Leur Connector, Male Leur Cap, and a suture clip. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10866614000105), received between Mar 17, 2022 and Mar 17, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Leak/Splash” (1 report).
1 other recall cover devices cleared under the same FDA submission (K170599); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 17, 2022Leak/Splash; Leak/Splash
CLINICAL STUDY: PATIENT HAD ON (B)(6) 2021 A CEREBROFLO EVD CATHETER PLACED TO TREAT SECONDARY INTRAVENTRICULAR HEMORRHAGE (IVH) DUE TO RUPTURED ANEURYSM. ON (B)(6) 2021 IT WAS REPORTED THE CEREBROFLO EVD CATHETER WAS LEAKING AT EXIST SITE. A CT SCAN ON (B)(6) REVEALS SOME AIR IN THE VENTRICLES AND IT WAS CONSIDERED REPLACING HER EVD, BUT WAS DEFERRED AS IT WAS FELT THE AIR WAS ENTRAINED FROM AROUND THE EVD RATHER THAN DUE TO MALFUNCTION OF EVD. AN ADDITIONAL SUTURE WAS PLACED AT THE EXIT SITE AND THE OUTCOME WAS RECOVERED/RESOLVED AND THE CATHETER WAS REMOVED ON (B)(6) 2022.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13790695 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170599. One submission often covers many models, so check the product and lot details in each record.
- Z-2173-2026 · Apr 10, 2026 · Open, ClassifiedCodman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD Catheter is a 3.3 mm diameter (10 Fr.), 35 cm long polyurethane
Other Shunt, Central Nervous System And Components devices
- NATUS EDS3 DRAINAGE SYSTEM NT821731CNatus Medical Incorporated · Shunt, Central Nervous System And Components1 recall208 reports
- Exacta™ 46700Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall2 deaths163 reports
- Strata® 42866Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 death160 reports
- proGAV 2.0® FX410TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components119 reports
- Becker® 46128Medtronic Ps Medical, Inc. · Shunt, central nervous system and components1 recall50 reports
- PROGAV 2.0 FX643TAesculap AG · Shunt, central nervous system and components100 reports
- M.BLUE PLUS FX804TAesculap AG · Shunt, central nervous system and components82 reports
- proSA® FV701TChristoph Miethke GmbH & Co. KG · Shunt, central nervous system and components82 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page