Skip to content

Medline Industries ENFIT70552Recalls & reported problems

Medline Industries, Inc. · SET,PUMP,1000ML BAG,ENFIT,MEDLINE · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10884389190333), received between May 14, 2019 and Sep 30, 2026 — classified as 3 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Crack3
  • Leak/Splash3
  • Obstruction of Flow2
  • Detachment of Device or Device Component1
  • Entrapment of Device1
  • Fluid/Blood Leak1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Feeding Problem1
  • Foreign Body In Patient1
  • Hypoglycemia1
  • Insufficient Information1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Break

    It was reported that "THE ISSUE IS WITH THE ENFIT PORTION, ENFIT TWIST CONNECTOR" and "THE ENFIT PIECE HAS BEEN SHATTERING CAUSING PIECES TO BE STUCK IN THE END OF THE EXTENSION TUBING OR JUST LEAKING."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946757 ↗
  • MalfunctionSep 24, 2026Break; No Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884867 ↗
  • MalfunctionSep 18, 2026Crack; Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820319 ↗
  • MalfunctionJun 5, 2026Break

    IT WAS REPORTED THAT THE CONNECTING PIECE BETWEEN THE BAG AND THE CATHETER IS BREAKING INSIDE OF THE CATHETER.

    Patient problems: Insufficient Information

    MAUDE report 25410048 ↗
  • InjurySep 10, 2025Material Fragmentation; Leak/Splash; Melted; Material Fragmentation; Leak/Splash; Melted

    ACCORDING TO THE FACILITY, ON (B)(6) 2025, "TUBE FEED BAG WAS CONNECTED TO G -TUBE. IT WAS LEAKING NOT LONG AFTER CONNECTION AND WHEN DISCONNECTED IT WAS NOTICED TO HAVE CRACKED AND MELTED INTO THE G - TUBE, CAUSING A PATIENT TO HAVE TO RECEIVE INTERVENTION AND HIS G -TUBE EMERGENTLY REPLACED. PER THE FACILITY, "BITS OF PLASTIC THAT CRACKED OFF FROM TUBE FEED CONNECTOR" GOT INSIDE THE PATIENT. "PATIENT IS DOING WELL."

    Patient problems: Foreign Body In Patient

    MAUDE report 23015853 ↗
  • InjuryJun 12, 2024Leak/Splash; Leak/Splash

    ACCORDING TO THE FACILITY ON 04/08/24 THERE WAS AN ISSUE "AT THE CONNECTION SITE WHERE IT SCREWS INTO EXTENSION TUBING" CAUSING THE FEEDING TO LEAK.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19519754 ↗
  • MalfunctionApr 12, 2023Obstruction of Flow

    MEDLINE, ENTRAFLOW DOES NOT FLOW AFTER PRIMING. CHAMBER FILLS AND WILL NOT DRIP. THE STAFF TRIED TROUBLE-SHOOTING BY MANUALLY FILLING AND CLEARING THE SYSTEM. IT CONTINUES TO BEEP AND DISPLAYS "OCCLUSION". MANUFACTURER RESPONSE FOR TUBES, GASTROINTESTINAL (AND ACCESSORIES), MEDLINE INDUSTRIES (PER SITE REPORTER) WILL OBTAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16723954 ↗
  • MalfunctionAug 3, 2021Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT A PIECE OF THE FEEDING PUMP CONNECTOR'S TIP BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12269663 ↗
  • MalfunctionJul 28, 2021Break

    IT WAS REPORTED THAT A PIECE OF THE TIP OF THE ENFIT PUMP SET CONNECTOR WAS FOUND TO HAVE BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12238795 ↗
  • MalfunctionJul 28, 2021Break

    IT WAS REPORTED THAT A PIECE OF THE TIP OF THE ENFIT PUMP SET CONNECTOR WAS FOUND TO HAVE BROKEN OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12238668 ↗
  • MalfunctionApr 7, 2021Break; Crack; Entrapment of Device

    IT WAS REPORTED, "THE TIP ON THE PURPLE ADAPTERS ARE CRACKING AND BREAKING WHILE IN THE CONNECTOR." ONE TIP PIECE CRACKED AND MADE ITS WAY INTO THE G-TUBE. G-TUBE REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11628263 ↗
  • MalfunctionJun 29, 2020Crack; Obstruction of Flow

    IT WAS REPORTED THE FEEDING BAG CRACK AND PIECES GOT STUCK IN THE GTT, WHICH STOPPED MEDICATION AND FEEDINGS.

    Patient problems: Feeding Problem

    MAUDE report 10205647 ↗
  • InjuryMay 14, 2019Leak/Splash

    IT WAS REPORTED THAT THE FEEDING SET WAS LEAKING AND PATIENT REQUIRED TREATMENT FOR HYPOGLYCEMIA.

    Patient problems: Hypoglycemia

    MAUDE report 8607683 ↗

Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices

All TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page