Skip to content

Argyle 8888413815Recalls & reported problems

Mozarc Medical US LLC · Peritoneal Dialysis Catheter, Swan Neck Curl Cath, 2 Cuffs Right · Class 2 · Implant

All models of Argyle (36)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521004627), received between Apr 23, 2018 and Jan 18, 2024 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Migration or Expulsion of Device2

Patient problems

  • Foreign Body In Patient2
  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Hematoma1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 18, 2024Break; Break

    ACCORDING TO THE REPORTER, THE CATHETER WAS DEFECTIVE. IT WAS STATED THAT IT WAS LEAKING FROM A TINY CRACK IN THE EXTERNAL PORTION (EXIT SITE) OF THE DRAIN. THE CATHETER WAS NOT REPAIRED AND WAS REPLACED LATER IN THE SAME EVENING. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18536745 ↗
  • InjuryJan 5, 2024Break; Break

    ACCORDING TO THE REPORTER, THE PERITONEAL DIALYSIS CATHETER BROKE IN THE PATIENT'S ABDOMEN.

    Patient problems: Foreign Body In Patient

    MAUDE report 18455743 ↗
  • MalfunctionJun 4, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    ACCORDING TO THE REPORTER, THE PERITONEAL DIALYSIS CATHETER WAS NOT WORKING AND WAS NOT ADEQUATELY DRAINING THE DIALYSATE. THE CATHETER¿S COIL MOVED. IT WAS STATED THAT CATHETER WAS REPOSITIONED. IT WAS ALSO STATED THAT THE CATHETER DID NOT BREAK. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 11942355 ↗
  • InjuryApr 20, 2021Break; Migration or Expulsion of Device; Break; Migration or Expulsion of Device

    ACCORDING TO THE REPORTER, THE CATHETER WAS INSERTED ON (B)(6) 2020 AND REPOSITIONED ON (B)(6) 2020 AS THERE WAS PROBLEM WITH THE CATHETER NOT ADEQUATELY DRAINING THE DIALYSATE (THE PD WAS NOT WORKING, THE COIL MOVED BUT THE CATHETER DID NOT BREAK, IT WAS A FUNCTIONAL PROBLEM). THE PATIENT HAD A SMALL ABDOMINAL TRAUMA AT HOME IN (B)(6) 2020 (WITH HEMOPERITONEUM), THE PATIENT BANGED ON A CORNER OF A TABLE AND THERE WAS BLEEDING AROUND THE CATHETER AND THE PATIENT HAD 2 ABDOMEN SCANS. THE FIRST SCAN WAS ON (B)(6) 2020 FOLLOWING THE TRAUMA THAT WAS SHOWING A COLLECTION AROUND THE CATHETER (IN THE ABDOMINAL WALL). THE CATHETER WAS BROKEN DURING THIS SCAN BUT IT WAS NOT NOTICED WHEN READING THE SCAN. THE PATIENT HAD ANOTHER CONTROL SCAN ON (B)(6) 2020 (MOST PROBABLY BEFORE RESTARTING THE PERITONEAL DIALYSIS). THE SCAN THEN SHOWED THAT THE COLLECTION/HEMATOMA AROUND THE CATHETER (IN THE ABDOMINAL WALL) ALMOST DISAPPEARED. THE BROKEN PART OF THE CATHETER IS AT THE SAME LOCATION AS IN THE LAST SCAN, BUT WAS STILL NOT DETECTED W HEN READING THE SCAN. IN RETROSPECT, THEY REALIZED THAT THE CATHETER WAS ALREADY BROKEN AT THAT TIME (7 CM OF THE CATHETER LODGED ITSELF IN THE LEFT SUBDIAPHRAGMATI

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body In Patient; Hematoma; Hemorrhage/Blood Loss/Bleeding; Unintended Radiation Exposure

    MAUDE report 11698061 ↗
  • MalfunctionJul 25, 2019Break; Break

    ACCORDING TO THE REPORTER, DURING PERITONEAL DIALYSIS, THE CATHETER WAS SECURED DURING EXTERNAL TRACTION AND GETTING WITHOUT DISTAL FOAM. THE BREAK OF THE CATHETER WAS FOUND AT THE DISTAL END(NEAR THE SKIN) WHICH MAKES THE CATHETER TO BURST AND MAKING ITS UNFEASIBLE. THE PROCEDURE WAS COMPLETED BUT THEY HAD TO EXTEND THE SURGICAL TIME OF 20 MINUTES DUE TO THE CATHETER REPLACEMENT. THE CATHETER WAS NOT REPAIRED AND THERE WAS NO BLOOD LEAK. TEGO WAS NOT UTILIZED AND THERE WAS NO LUER ADAPTER ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8829176 ↗
  • MalfunctionApr 23, 2018Break; Break

    ACCORDING TO THE REPORTER, POST LAPAROSCOPIC INSTALLATION OF CATHETER IN A FILTRATION, WHEN THE PATIENT CHANGED POSITION THEY NOTICED THAT THE DRESSING BECAME WET AND THERE WAS A LEAK. THEY STILL CONTINUED WITH THE THERAPY FOR SEVERAL DAYS BECAUSE OF FILTRATION REASONS. THEY TESTED THE CATHETER AND THEY CERTIFIED THAT THERE WAS A FISSURE APPROXIMATELY 1 CM AFTER THE SECOND CUFF NEAR THE RADIOPAQUE LINE. THEY SCHEDULED FOR ANOTHER LAPAROSCOPIC PROCEDURE TO REMOVE THE CATHETER, AND INSTALL A NEW ONE TO RESOLVE THE CASE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7452490 ↗

Other CATHETER, PERITONEAL, LONG-TERM INDWELLING devices

All CATHETER, PERITONEAL, LONG-TERM INDWELLING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page