MAHURKAR 8888135194Recalls & reported problems
Mozarc Medical US LLC · Acute Dual Lumen Catheter Tray,High Flow Straight Extensions with IC* Components · Class 2
All models of MAHURKAR (81)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521006508), received between Feb 28, 2019 and Sep 16, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
9 other recalls cover devices cleared under the same FDA submission (K030209, K192302); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Dec 7, 2022Class IOpen, Classified
There is a potential leaking condition within the hub of affected devices, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Z-0877-2023 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
- Device Markings/Labelling Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2021Break; Break
ACCORDING TO THE REPORTER, ON AUG 26, IT WAS SAID THAT THE PATIENT HAD A CATHETER LINE PLACEMENT. ON AUG 29 AT 20:00, BLOOD LEAKAGE WAS NOTICED FROM THE BIFURCATE OF THE DUAL LUMEN CATHETER. THE DOCTOR INSPECTED THE CATHETER AND NOTICED THAT THERE WERE CIRCULAR CRACK AROUND THE CATHETER - BIFURCATE (WHITE EXTERNAL PORTION), HOWEVER, THE DOCTOR DID NOT SAVE THE CATHETER. THE CATHETER WAS NOT REPAIRED. THERE WAS NO LUER ADAPTER ISSUE. ON AUG 30, THE CATHETER WAS REMOVED AND REPLACED AS INTERVENTION DUE TO THE EVENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12481054 ↗InjuryFeb 28, 2019Device Markings/Labelling Problem; Device Markings/Labelling Problem
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, CLINICIAN STATED THAT THERE WAS A DIFFERENCE OF .6ML BETWEEN THE PRINTED PRIMING VOLUME AND THE ACTUAL LUMEN CAPACITY, THE PATIENT WAS GIVEN WITH 5000 UNITS OF HEPARIN PER ML, THIS GAVE THE PATIENT WHO HAD EMBOLIC STROKE 3000 UNITS THAT WENT SYSTEMIC THUS INCREASING THE RISK OF A HEMORRHAGIC STROKE, THE PATIENT SPENT A COUPLE OF DAYS IN THE INTENSIVE CARE UNIT(ICU) (INTENSIVE CARE UNIT) AS HEPARIN BOLUS WAS A MAJOR FACTOR FOR THE EXTENDED STAY, PATIENTS' ICU STAY WAS INITIALLY FOR EMBOLIC STROKE INTERVENTION. IT WAS STATED THAT PRIMING VOLUME PRINTED ON THE CATHETER AND IN THE IFU IS 1.6 ML ON THE ARTERIAL (RED) LUMEN AND 1.7 ML ON THE VENOUS (BLUE) LUMEN WAS NOT ACCURATE. IT WAS ALSO STATED THAT THE "ACTUAL" VOLUME THAT THE CATHETER LUMENS HOLD IS LESS THAN PRINTED IN THE IFU (INSTRUCTIONS FOR USE) AND ON THE CATHETER, TESTED PRIMING VOLUMES WERE OVER BY ABOUT .4-.6ML. IT WAS REPLACED WITH A COMPETITORS PRODUCT. THERE WAS NO REPORTED PATIENT HARM AND THE PATIENT WAS DISCHARGED AFTER 1-2 DAYS OF ADDITIONAL OBSERVATION FROM THE ICU.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Code Available; No Consequences Or Impact To Patient
MAUDE report 8380042 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K030209, K192302. One submission often covers many models, so check the product and lot details in each record.
- Z-0422-2026 · Oct 17, 2025 · Open, ClassifiedMAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 16 cm;
- Z-0423-2026 · Oct 17, 2025 · Open, ClassifiedMAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20 cm;
- Z-0425-2026 · Oct 17, 2025 · Open, ClassifiedMAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 24 cm;
- Z-0421-2026 · Oct 17, 2025 · Open, ClassifiedMAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 13 cm;
- Z-0426-2026 · Oct 17, 2025 · Open, ClassifiedMAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Safety Components: 12 Fr/Ch (4.0 mm) x 24 cm;
- Z-0424-2026 · Oct 17, 2025 · Open, ClassifiedMAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20 cm;
- Z-1955-2019 · Apr 29, 2019 · TerminatedMAHURKAR and Argyle acute hemodialysis catheters, Material numbers 8813793009 8813793013 8813794005 8813794009 8813816005 8813816009 8813817005 8813817009 8815668007 88156
- Z-0233-04 · Dec 2, 2003 · TerminatedMahurkar QPlus Dual Lumen Catheter Kit with Curved Extension, 13.5 Fr/Ch x 19.5cm For hemodialysis, apheresis, and infusion. Reference: 8888-135192
- Z-0234-04 · Dec 2, 2003 · TerminatedMahurkar QPlus Dual Lumen Catheter Insertion Tray with Curved Extension, 13.5 Fr/Ch x 19.5cm For hemodialysis, apheresis, and infusion Reference: 8888-135195
Other CATHETER, HEMODIALYSIS, NON-IMPLANTED devices
- MAHURKAR 8888135191Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, NON-IMPLANTED1 recall175 reports
- MAHURKAR 8888135241Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, NON-IMPLANTED1 recall1 death100 reports
- ARROW IPN036723Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls2 reports
- ARROW IPN036412Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls1 report
- DuoGlide catheter 5625240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls1 report
- ARROW IPN036428Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls0 reports
- ARROW IPN036419Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls0 reports
- ARROW IPN927343Teleflex Incorporated · Catheter, hemodialysis, non-implanted2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page