Skip to content

Dover 67000Recalls & reported problems

Cardinal Health, Inc. · Bulb Syringe,Irrigation with Protective Cap · Class 1

All models of Dover (90)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521007994), received between Dec 9, 2022 and Dec 9, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Low Blood Pressure/ Hypotension1
  • Muscle Weakness/Atrophy1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT THE CLEAR PLASTIC PART OF THE BALL SUCTION CAME OFF THE BALL AND FELL ONTO THE PATIENT'S CORD. ADDITIONAL INFORMATION RECEIVED ON 28NOV2022 STATED THAT THE PATIENT UNDERWENT A CERVICAL LAMINECTOMY, SO THE SPINAL CORD (COVERED BY DURA) WAS EXPOSED AND UNPROTECTED BY THE BONE. THE ENTIRE CLEAR PIECE (SYRINGE) FILLED WITH NORMAL SALINE SEPARATED FROM THE GREEN RUBBER BALL AND FELL ONTO THE EXPOSED DURA/SPINAL CORD. IT WAS IMMEDIATELY REMOVED FROM THE SURGICAL FIELD. THERE WAS A DECLINE IN THE NEUROMONITORING DURING THE CASE AND THEY HAD TO ELEVATE THE BLOOD PRESSURE TO MAPS > 85. THE PATIENT DID HAVE SOME POSTOPERATIVE WEAKNESS WHICH HAS NOW RESOLVED.

    Patient problems: Low Blood Pressure/ Hypotension; Muscle Weakness/Atrophy

    MAUDE report 15953316 ↗

FDA clearance and classification

The FDA regulates product code KYZ (syringe, irrigating (non dental)) under 21 CFR 880.6960 as a Class I device. 21 CFR 880.6960 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Dover 67000

Has the Dover 67000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521007994), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Dover 67000?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Dec 9, 2022. That report lists device problem “Detachment of Device or Device Component” (1); patient problems “Low Blood Pressure/ Hypotension” (1) and “Muscle Weakness/Atrophy” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dover 67000?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Dover 67000 FDA approved?

Devices of this type (product code KYZ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cardinal Health, Inc.

Who makes the Dover 67000, and is it still sold?

The FDA's device database lists Cardinal Health, Inc. as the company responsible for it (brand Dover, model 67000). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SYRINGE, IRRIGATING (NON DENTAL) devices

All SYRINGE, IRRIGATING (NON DENTAL) devices →

Sources and updates