Skip to content

Biosyn GM-291Recalls & reported problems

Covidien LP · Monofilament Absorbable Suture · Class 2 · Implant

All models of Biosyn (94)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521032651), received between Jan 29, 2020 and Feb 2, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

28 other recalls cover devices cleared under the same FDA submission (K000037); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, PRIOR TO A PROCEDURE, WHEN THE SUTURE WAS REMOVED FROM THE PACK, THE NEEDLE DETACHED FROM THREAD. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18632509 ↗
  • MalfunctionDec 30, 2021Break; Break

    ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEMICOLECTOMY, THE THREAD OF THE SUTURE BROKE WHILE SEWING. ANOTHER ONE WAS USED, BUT THE SAME ISSUE OCCURRED. ANOTHER DEVICE FROM ANOTHER PACKAGE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13119121 ↗
  • MalfunctionAug 19, 2021Break; Break

    ACCORDING TO THE REPORTER, DURING A CHOLEDOCHOJEJUNOSTOMY, IT WAS REPORTED THAT THE SUTURE BROKE EASILY. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE AND COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12345983 ↗
  • MalfunctionJan 29, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, ON A PROXIMAL PANCREATICODUODENECTOMY, DURING USE, THE NEEDLE DISENGAGED FROM THE THREAD. THE DEVICE OCCURRED CONSECUTIVELY. ANOTHER DEVICE WAS FROM ANOTHER LOT USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9641340 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K000037. One submission often covers many models, so check the product and lot details in each record.

  • Z-1280-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2 36" VIOLET GS-24, Item Code CL911 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not i
  • Z-1287-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 0 VIOLET 30" V-20, Item Code GL124 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
  • Z-1294-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture UD 3/0 30 P-14, Item Code SL5679 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in c
  • Z-1277-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 3/0 30 VIOLET GS-11, Item Code CL815 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1285-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2-0 UNDYED 36" GS-21, Item Code CL953 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1289-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2-0 UNDYED 30" V-20, Item Code GL323 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1284-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 1 36 VIOLET GS-21, Item Code CL925 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
  • Z-1293-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 6/0 18" UNDYED P-10, Item Code SL1625G - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but no
  • Z-1291-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 6/0 12"" VL SS-24 D/A, Item Code L1742K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but n
  • Z-1276-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2/0 30 VIOLET GS-21, Item Code CL811 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1292-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 6-0 18" UNDYED SS-24 D/A, Item Code L2752K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, bu
  • Z-1281-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 0 36 VIOLET GS-24, Item Code CL914 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
  • Z-1282-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 1 36 VIOLET GS-24, Item Code CL915 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
  • Z-1286-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2-0 VIOLET 30" V-20, Item Code GL123 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1275-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 0 30" UNDYED HOS-11, Item Code CL537 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not

…and 13 more.

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page