Skip to content

Biosyn GM-323Recalls & reported problems

Covidien LP · Monofilament Absorbable Suture · Class 2 · Implant

All models of Biosyn (94)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521032705), received between Oct 28, 2019 and Oct 28, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

28 other recalls cover devices cleared under the same FDA submission (K000037); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2019Break; Break

    ACCORDING TO THE REPORTER, PRIOR TO PROSTATECTOMY OPEN PROCEDURE, THE THREAD BROKE AFTER TAKING OUT OF THE PACKAGE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 9247947 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K000037. One submission often covers many models, so check the product and lot details in each record.

  • Z-1291-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 6/0 12"" VL SS-24 D/A, Item Code L1742K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but n
  • Z-1275-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 0 30" UNDYED HOS-11, Item Code CL537 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1293-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 6/0 18" UNDYED P-10, Item Code SL1625G - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but no
  • Z-1296-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 4/0 30 VIO SC-2 124K, Item Code SL654 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1287-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 0 VIOLET 30" V-20, Item Code GL124 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
  • Z-1294-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture UD 3/0 30 P-14, Item Code SL5679 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in c
  • Z-1290-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 3/0 30" UNDYED V-20, Item Code GLS322 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1289-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2-0 UNDYED 30" V-20, Item Code GL323 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1288-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2/0 UNDYED 30" V-30 (75CM), Item Code GL228 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, b
  • Z-1276-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2/0 30 VIOLET GS-21, Item Code CL811 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1283-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2-0 VIOLET 36" GS-21, Item Code CL923 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1300-2020 · Dec 18, 2019 · Terminated
    Biosyn Monofilament Absorbable Suture 6/0 30" VIOLET CV-23, Item Code UM201
  • Z-1277-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 3/0 30 VIOLET GS-11, Item Code CL815 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
  • Z-1280-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 2 36" VIOLET GS-24, Item Code CL911 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not i
  • Z-1284-2020 · Dec 18, 2019 · Terminated
    Polysorb Braided Absorbable Suture 1 36 VIOLET GS-21, Item Code CL925 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in

…and 13 more.

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page